A Self-Sorting Coronal 4D-MRI Method for Daily Image Guidance of Liver Lesions on an MR-LINAC

被引:32
|
作者
van de Lindt, Tessa [1 ]
Sonke, Jan-Jakob [1 ]
Nowee, Marlies [1 ]
Jansen, Edwin [1 ]
van Pelt, Vivian [1 ]
van der Heide, Uulke [1 ]
Fast, Martin [1 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Plesmanlaan 1, NL-1066 CX Amsterdam, Netherlands
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2018年 / 102卷 / 04期
关键词
GUIDED RADIOTHERAPY TREATMENT; IMPLANTED FIDUCIAL MARKERS; BODY RADIATION-THERAPY; RESPIRATORY MOTION; PHASE; RECONSTRUCTION; METASTASES; PRECISION; NAVIGATOR; OUTCOMES;
D O I
10.1016/j.ijrobp.2018.05.029
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Novel hybrid MR-LINAC devices provide MRI's superior soft-tissue contrast in the treatment room and thus have the potential to increase accuracy of liver stereotactic body radiation therapy (SBRT). Requirements for daily position verification using 4-dimensional MRI include tumor visibility and short acquisition-reconstruction time (preferably <5 min). The proposed method provides fast acquisition-reconstruction time and the flexibility to vary T1- and T2-weighting, using standard imaging sequences for straightforward implementation on an MR-LINAC. Methods and Materials: Images were acquired using a coronal 2-dimensional, multislice, single-shot turbo spin-echo (TSE) and turbo field-echo (TFE) sequence, which were repeated 30 times. An image-based self-sorting signal (ImS) was extracted from the data, and rigid registration of the diaphragm per slice position was performed and corrected for amplitude variation in the anteroposterior direction. Data were sorted into 10 bins according to amplitude and phase. ImS was validated in 4 healthy volunteers against a navigator signal. Positional variations within bins, missing data, and smoothness of the liver dome were compared between amplitude and phase binning in 10 volunteers. Tumor contrast and registration were investigated in 3 patients. Results: Each ImS was found to be in excellent agreement with the navigator signal with a correlation coefficient of >0.95 and binning differences of <1 bin. Better liver dome smoothness per bin in case of amplitude binning compared with that in phase binning (2.0-2.6 mm vs 2.4-3.7 mm, respectively) is a tradeoff for more missing data (3.5%-17.5% vs 3.5%-4.7%, respectively). Liver lesions were visible in almost all coronal TSE and TFE images, but the lesion boundary was better defined in the TSE images. Rigid registrations could be performed on the tumor area. Conclusions: An efficient self-sorted 4-dimensional MRI method was developed and validated using standard sequences and fast reconstruction on a LINAC-integrated MRI scanner providing good tumor visibility for daily image-guided liver stereotactic body radiation therapy. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:875 / 884
页数:10
相关论文
共 37 条
  • [1] A self-gated coronal 4D-MRI method for daily imaging of liver lesions on an MR-Linac
    Van de Lindt, T.
    Fast, M.
    Nowee, M.
    Jansen, E.
    Pelt, V.
    Van der Heide, U.
    Sonke, J. J.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S523 - S524
  • [2] Evaluation of 4D-MRI Guided Liver SBRT on the MR-Linac
    van de Lindt, T. N.
    Nowee, M. E.
    Janssen, T.
    van Pelt, V.
    Jansen, E. P. M.
    Schneider, C.
    Remeijer, P.
    Sonke, J. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : S66 - S66
  • [3] Registration accuracy of 4D-MRI in lung acquired on the MR-linac
    Rossi, M.
    Fast, M.
    Van de Lndt, T.
    Nowee, M.
    Belderbos, J.
    Lalezari, F.
    Sonke, J.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S547 - S547
  • [4] Technical feasibility and clinical evaluation of 4D-MRI guided liver SBRT on the MR-linac
    van de Lindt, T. N.
    Nowee, M. E.
    Janssen, T.
    Schneider, C.
    Remeijer, P.
    van Pelt, V. W. J.
    Betgen, A.
    Jansen, E. P. M.
    Sonke, J. J.
    RADIOTHERAPY AND ONCOLOGY, 2022, 167 : 285 - 291
  • [5] Correcting geometric image distortions in slice-based 4D-MRI on the MR-linac
    Keesman, Rick
    van de Lindt, Tessa N.
    Juan-Cruz, Celia
    van den Wollenberg, Wouter
    van der Bijl, Erik
    Nowee, Marlies E.
    Sonke, Jan-Jakob
    van der Heide, Uulke A.
    Fast, Martin F.
    MEDICAL PHYSICS, 2019, 46 (07) : 3044 - 3054
  • [6] The dosimetric impact of geometric image distortions in slice-based 4D-MRI on the MR-linac
    Fast, M.
    Keesman, R.
    Van den Wollenberg, W.
    Juan-Cruz, C.
    Van de Lindt, T.
    Van der Heide, U.
    Sonke, J.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S537 - S538
  • [7] Geometric validation of a 4D-MRI guided correction strategy on the MR-Linac
    Van de Lindt, T.
    Koopman, R.
    Van de Schoot, A.
    Torres-Xirau, I.
    Van der Heide, U.
    Sonke, J. J.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S468 - S469
  • [8] Better motion-compensated daily imaging using 4D-MRI for MR-linac gating workflows
    Keijnemans, Katrinus
    Schakel, Tim
    Borman, Pim T. S.
    Raaymakers, Bas W.
    Paulson, Eric S.
    Fast, Martin F.
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S4237 - S4240
  • [9] Effectiveness of visual biofeedback-guided respiratory-correlated 4D-MRI for radiotherapy guidance on the MR-linac
    Keijnemans, Katrinus
    Borman, Pim
    Raaymakers, Bas
    Fast, Martin
    MAGNETIC RESONANCE IN MEDICINE, 2024, 91 (01) : 297 - 311
  • [10] Validation of a 4D-MRI guided liver stereotactic body radiation therapy strategy for implementation on the MR-linac
    van de Lindt, Tessa N.
    Fast, Martin F.
    van den Wollenberg, Wouter
    Kaas, Jochem
    Betgen, Anja
    Nowee, Marlies E.
    Jansen, Edwin Pm
    Schneider, Christoph
    van der Heide, Uulke A.
    Sonke, Jan-Jakob
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (10):