Effectiveness of visual biofeedback-guided respiratory-correlated 4D-MRI for radiotherapy guidance on the MR-linac

被引:2
|
作者
Keijnemans, Katrinus [1 ,2 ]
Borman, Pim [1 ]
Raaymakers, Bas [1 ]
Fast, Martin [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiotherapy, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
基金
荷兰研究理事会;
关键词
4D-MRI; abdominothoracic cancer; MR-linac; radiotherapy; visual biofeedback; INTERNAL LUNG MOTION; AUDIOVISUAL BIOFEEDBACK; GATED RADIOTHERAPY; TUMOR MOTION; IMPACT; CT; MANAGEMENT;
D O I
10.1002/mrm.29857
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose:<bold> </bold>Respiratory-correlated 4D-MRI may provide motion characteristics for radiotherapy but is susceptible to irregular breathing. This study investigated the effectiveness of visual biofeedback (VBF) guidance for breathing regularization during 4D-MRI acquisitions on an MR-linac. Methods:<bold> </bold>A simultaneous multislice-accelerated 4D-MRI sequence was interleaved with a one-dimensional respiratory navigator (1D-RNAV) in 10 healthy volunteers on a 1.5T Unity MR-linac (Elekta AB, Stockholm, Sweden). Volunteer-specific breathing amplitudes and periods were derived from the 1D-RNAV signal obtained during unguided 4D-MRI acquisitions. These were used for the guidance waveform, while the 1D-RNAV positions were overlayed as VBF. VBF effectiveness was quantified by calculating the change in coefficient of variation (CVdiff) for the breathing amplitude and period, the position SD of end-exhale, end-inhale and midposition locations, and the agreement between the 1D-RNAV signals and guidance waveforms. The 4D-MRI quality was assessed by quantifying amounts of missing data. Results:<bold> </bold>VBF had an average latency of 520 +/- 2 ms. VBF reduced median breathing variations by 18% to 35% (amplitude) and 29% to 57% (period). Median position SD reductions ranged from -3% to 35% (end-exhale), 29% to 38% (end-inhale), and 25% to 37% (midposition). Average differences between guidance waveforms and 1D-RNAV signals were 0.0 s (period) and +1.7 mm (amplitude). VBF also decreased the median amount of missing data by 11% and 29%. Conclusion:<bold> </bold>A VBF system was successfully implemented, and all volunteers were able to adapt to the guidance waveform. VBF during 4D-MRI acquisitions drastically reduced breathing variability but had limited effect on missing data in respiratory-correlated 4D-MRI.
引用
收藏
页码:297 / 311
页数:15
相关论文
共 50 条
  • [1] Impact of Visual Biofeedback On Breathing Regularity During 4D-MRI Acquisition On the MR-Linac
    Keijnemans, K.
    Borman, P.
    Raaymakers, B.
    Fast, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E248 - E248
  • [2] Evaluation of 4D-MRI Guided Liver SBRT on the MR-Linac
    van de Lindt, T. N.
    Nowee, M. E.
    Janssen, T.
    van Pelt, V.
    Jansen, E. P. M.
    Schneider, C.
    Remeijer, P.
    Sonke, J. J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : S66 - S66
  • [3] Geometric validation of a 4D-MRI guided correction strategy on the MR-Linac
    Van de Lindt, T.
    Koopman, R.
    Van de Schoot, A.
    Torres-Xirau, I.
    Van der Heide, U.
    Sonke, J. J.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S468 - S469
  • [4] The noise navigator: a surrogate for respiratory-correlated 4D-MRI for motion characterization in radiotherapy
    Navest, R. J. M.
    Mandija, S.
    Bruijnen, T.
    Stemkens, B.
    Tijssen, R. H. N.
    Andreychenko, A.
    Lagendijk, J. J. W.
    van den Berg, C. A. T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (01):
  • [5] Registration accuracy of 4D-MRI in lung acquired on the MR-linac
    Rossi, M.
    Fast, M.
    Van de Lndt, T.
    Nowee, M.
    Belderbos, J.
    Lalezari, F.
    Sonke, J.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S547 - S547
  • [6] Technical feasibility and clinical evaluation of 4D-MRI guided liver SBRT on the MR-linac
    van de Lindt, T. N.
    Nowee, M. E.
    Janssen, T.
    Schneider, C.
    Remeijer, P.
    van Pelt, V. W. J.
    Betgen, A.
    Jansen, E. P. M.
    Sonke, J. J.
    RADIOTHERAPY AND ONCOLOGY, 2022, 167 : 285 - 291
  • [7] Validation of a 4D-MRI guided liver stereotactic body radiation therapy strategy for implementation on the MR-linac
    van de Lindt, Tessa N.
    Fast, Martin F.
    van den Wollenberg, Wouter
    Kaas, Jochem
    Betgen, Anja
    Nowee, Marlies E.
    Jansen, Edwin Pm
    Schneider, Christoph
    van der Heide, Uulke A.
    Sonke, Jan-Jakob
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (10):
  • [8] Self-Navigated, Respiratory-Correlated 4D-MRI with Switchable Contrast Modes
    Paulson, E.
    Mickevicius, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 3313 - 3313
  • [9] Simultaneous Multi-Slice Accelerated 4D-MRI for the Unity MR-Linac
    Keijnemans, K.
    Borman, P. T. S.
    Raaymakers, B. W.
    Fast, M. F.
    MEDICAL PHYSICS, 2020, 47 (06) : E440 - E440
  • [10] A Self-Sorting Coronal 4D-MRI Method for Daily Image Guidance of Liver Lesions on an MR-LINAC
    van de Lindt, Tessa
    Sonke, Jan-Jakob
    Nowee, Marlies
    Jansen, Edwin
    van Pelt, Vivian
    van der Heide, Uulke
    Fast, Martin
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (04): : 875 - 884