A measurement of the fast-neutron sensitivity of a Geiger-Muller detector in the pulsed neutron beam from a superconducting cyclotron

被引:9
|
作者
Maughan, RL [1 ]
Yudelev, M [1 ]
Kota, C [1 ]
机构
[1] WAYNE STATE UNIV, DETROIT, MI 48230 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1996年 / 41卷 / 08期
关键词
D O I
10.1088/0031-9155/41/8/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The k(U) value of a commercially available miniature energy compensated Geiger-Muller (GM) detector has been determined using the modified lead attenuation method of Hough. The measurements were made in a d(48.5)-Be neutron beam produced by the superconducting cyclotron based neutron therapy facility at Harper Hospital. The unique problems associated with making measurements in a 2 ms duration pulsed beam with a 20% duty cycle are discussed. The beam monitoring system, which allows the beam pulse shape at low beam intensities to be measured, is described. By gating the GM output with a discriminator pulse derived from the beam pulse shape, the gamma-ray count rates and dead-time corrections within the 2 ms pulse and between pulses can be measured separately. The k(U) value of (0.0245 +/- 0.0015) determined for this GM detector is consistent with the values measured by other workers with identical and similar detectors in neutron beams with comparable, but not identical, neutron spectra.
引用
收藏
页码:1341 / 1351
页数:11
相关论文
共 48 条
  • [41] SENSITIVITY OF CR-39 SSNTDS FOR MEASUREMENTS OF DEPTH-DOSE AND MICRODOSIMETRIC SPECTRA OF A P(43) BE(22) FAST-NEUTRON BEAM
    AROUMOUGAME, R
    BLUE, TE
    BLUE, JW
    ROBERTS, WK
    MEDICAL PHYSICS, 1986, 13 (04) : 578 - 578
  • [42] PRE-THERAPEUTIC EXPERIMENTS WITH THE FAST NEUTRON BEAM FROM THE MEDICAL RESEARCH COUNCIL CYCLOTRON - GENERAL REVIEW
    FOWLER, JF
    MORGAN, RL
    BRITISH JOURNAL OF RADIOLOGY, 1963, 36 (422): : 115 - 121
  • [43] EVALUATION OF FAST-NEUTRON BEAM TELETHERAPY OF METASTATIC CERVICAL ADENOPATHY FROM SQUAMOUS-CELL CARCINOMAS OF THE HEAD AND NECK REGION
    GRIFFIN, TW
    LARAMORE, GE
    PARKER, RG
    GERDES, AJ
    HEBARD, DW
    BLASKO, JC
    GROUDINE, M
    CANCER, 1978, 42 (06) : 2517 - 2520
  • [44] Measurement and calculation of the fast-neutron and photon spectra from the core boundary to the biological shielding in the WWER-1000 reactor model
    Osmera, B
    Cvachovec, F
    Kyncl, J
    Smutny, V
    RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) : 92 - 94
  • [45] A FAST-NEUTRON POLARIMETER AND THE POLARIZATION MEASUREMENT OF NEUTRONS FROM THE H-2(D,N)HE-3 REACTION
    ZHU, SJ
    XU, SD
    CHEN, ZM
    ZHOU, YL
    XU, XY
    ARABIAN JOURNAL FOR SCIENCE AND ENGINEERING, 1983, 8 (04): : 367 - 376
  • [46] MEASUREMENT OF RELATIVE BIOLOGICAL EFFECTIVENESS (RBE) FOR THE RADIATION BEAM FROM NEUTRON SOURCE REACTOR YAYOI - COMPARISONS WITH CYCLOTRON NEUTRON AND CO-60 GAMMA-RAY
    WAKABAYASHI, H
    SUZUKI, S
    ITO, A
    JOURNAL OF RADIATION RESEARCH, 1983, 24 (04) : 305 - 316
  • [47] Measurement of fast neutron spectra from the interaction of 20 MeV protons with thick Be and C targets using CR-39 detector
    Tripathy, S. P.
    Paul, S.
    Sahoo, G. S.
    Suman, V.
    Sunil, C.
    Joshi, D. S.
    Bandyopadhyay, T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2014, 318 : 237 - 240
  • [48] Direct measurement of 59Ni(n, p) 59Co and 59Ni(n, α) 56Fe at fast-neutron energies from 500 keV to 10 MeV
    Kuvin, S. A.
    Lee, H. Y.
    DiGiovine, B.
    Eiroa-Lledo, C.
    Georgiadou, A.
    Herman, M.
    Kawano, T.
    Mocko, V
    Mosby, S.
    Vermeulen, C.
    Votaw, D.
    White, M.
    Zavorka, L.
    Perdikakis, G.
    Tsintari, P.
    Kim, H., I
    PHYSICAL REVIEW C, 2022, 105 (04)