Defibrillation failure in patients undergoing replacement of subcutaneous defibrillator pulse generator

被引:26
|
作者
Rudic, Boris [1 ,2 ]
Tueluemen, Erol [1 ,2 ]
Fastenrath, Fabian [1 ,2 ]
Akin, Ibrahim [1 ,2 ]
Borggrefe, Martin [1 ,2 ]
Kuschyk, Juergen [1 ,2 ]
机构
[1] Univ Med Ctr Mannheim, Dept Med Cardiol 1, Theodor Kutzer Ufer 1-3, D-68167 Mannheim, Germany
[2] German Ctr Cardiovasc Res DZHK, Partner Site Heidelberg Mannheim, Mannheim, Germany
关键词
Defibrillation threshold test; Generator replacement; Subcutaneous defibrillator; PRAETORIAN score; Ventricular fibrillation; IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; THRESHOLD; OUTCOMES; TIME;
D O I
10.1016/j.hrthm.2019.10.024
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Defibrillation threshold (DFT) testing is commonly performed in patients undergoing subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation. Growing evidence indicates that successful DFT testing correlates with the technique used for implantation of the defibrillation lead and pulse generator. However, evidence on whether DFT testing should be performed in patients undergoing S-ICD pulse generator replacement is lacking. OBJECTIVE The purpose of this study was to determine the outcome of DFT testing in patients undergoing replacement of an S-ICD pulse generator. METHODS A total of 357 S-ICD implantations were performed between November 2010 and July 2019. Twenty-five consecutive patients underwent S-ICD replacement between 2015 and 2019. Clinical data, perioperative medication, technical measurements, and PRAETORIAN score were assessed and grouped according to the outcome of DFT testing. RESULTS In 5 of 25 patients (20%), induced ventricular fibrillation was not successfully terminated after the first or second 65-J shock after pulse generator replacement with need for external defibrillation. Repositioning of the pulse generator and/or capsulectomy at the pocket site were necessary to achieve effective DFT with 65 J in all cases. Shock impedance increased in all patients at the time of pulse generator replacement compared to first implantation and was significantly higher in patients with ineffective DFT (119 +/- 17 Omega vs 93 +/- 26 Omega; P = .03). Otherwise, no differences in clinical characteristics, comorbidities, body mass index, intraoperative medication, or PRAETORIAN score were predictive of defibrillation failure. CONCLUSION The high proportion of patients with DFT failure after S-ICD pulse generator replacement indicates that DFT testing is mandatory to ensure safe function of the S-ICD.
引用
收藏
页码:455 / 459
页数:5
相关论文
共 50 条
  • [31] RARE CASE OF TWIDDLER'S SYNDROME IN A SUBCUTANEOUS IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR PRESENTING AS FAILURE OF DEFIBRILLATION THRESHOLD TEST
    Bhattarai, Mukul
    Salih, Mohsin
    Al-Akchar, Mohammad
    Labedi, Mohamed Rafa
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2022, 79 (09) : 2498 - 2498
  • [32] Predictors of appropriate shock after generator replacement in patients with an implantable cardioverter defibrillator
    Arcinas, Liane A.
    Chew, Derek S.
    Seifer, Colette M.
    Baranchuk, Adrian
    Supel, Izabella
    Exner, Derek V.
    Boles, Usama
    McIntyre, William F.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2021, 44 (05): : 911 - 918
  • [33] Predictors of the improvement of defibrillation thresholds with an active pulse generator
    Gold, MR
    Khalighi, K
    Shorofsky, SR
    CIRCULATION, 1996, 94 (08) : 2567 - 2567
  • [34] Reduction in defibrillation threshold by modifications to the subcutaneous implantable-cardioverter defibrillator coil
    Heist, E. Kevin
    Knops, Reinoud E.
    Yap, Sing-Chien
    Boersma, Lucas V. A.
    Friedman, Daniel J.
    Poole, Jeanne E.
    Stahl, Wyatt
    Belalcazar, Andres
    HEART RHYTHM, 2023, 20 (05) : 783 - 785
  • [35] Subcutaneous defibrillator lead management in a patient undergoing a sternotomy
    Gabriels, James
    Donnelly, Joseph
    Willner, Jonathan
    Beldner, Stuart
    Patel, Apoor
    Lima, Brian
    Epstein, Laurence M.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (12): : 1681 - 1683
  • [36] Intermuscular Subcutaneous Implantable Cardioverter-Defibrillator Implantation Without Defibrillation Testing
    Mistry, Amar
    Pooranachandran, Vivetha
    Dhutia, Harshil
    Chelliah, Rajesh
    Pathmanathan, Ravi K.
    JACC-CLINICAL ELECTROPHYSIOLOGY, 2021, 7 (01) : 124 - 126
  • [37] Predictive value of a positive pocket swab for the occurrence of device infection in implantable cardioverter defibrillator patients undergoing generator replacement or lead revision
    Kleemann, T.
    Mousavi, S. B.
    Kouraki, K.
    Strauss, M.
    Mohammad, O.
    Wenz, A.
    Weisse, U.
    Werling, C.
    Sack, F.
    Zahn, R.
    EUROPEAN HEART JOURNAL, 2024, 45
  • [38] Incidence of Ventricular Arrhythmias and 1-Year Predictors of Mortality in Patients Treated With Implantable Cardioverter-Defibrillator Undergoing Generator Replacement
    Demarchi, Andrea
    Cornara, Stefano
    Sanzo, Antonio
    Savastano, Simone
    Petracci, Barbara
    Vicentini, Alessandro
    Pontillo, Lorenzo
    Baldi, Enrico
    Frigerio, Laura
    Astuti, Matteo
    Leonardi, Sergio
    Ghio, Stefano
    Visconti, Luigi Oltrona
    Rordorf, Roberto
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2021, 10 (04): : 1 - 9
  • [39] Defibrillation testing is mandatory in patients with subcutaneous implantable cardioverter-defibrillator to confirm appropriate ventricular fibrillation detection
    de Waroux, Jean-Benoit le Polain
    Ploux, Sylvain
    Mondoly, Pierre
    Eschalier, Romain
    Strik, Marc
    Houard, Laura
    Pierre, Bertrand
    Buliard, Samuel
    Klotz, Nicolas
    Ritter, Philippe
    Haissaguerre, Michel
    Mahfouz, Karim
    Bordachar, Pierre
    HEART RHYTHM, 2018, 15 (05) : 642 - 650
  • [40] Subcutaneous transthoracic defibrillation in implantable cardioverter defibrillator patients using a pectoral can to posterior coil electrode configuration
    Lieberman, RA
    Shorofsky, SR
    Rashba, E
    Havel, WJ
    Christensen, JL
    De Groot, PJ
    CIRCULATION, 2005, 112 (17) : U429 - U429