Defibrillation failure in patients undergoing replacement of subcutaneous defibrillator pulse generator

被引:26
|
作者
Rudic, Boris [1 ,2 ]
Tueluemen, Erol [1 ,2 ]
Fastenrath, Fabian [1 ,2 ]
Akin, Ibrahim [1 ,2 ]
Borggrefe, Martin [1 ,2 ]
Kuschyk, Juergen [1 ,2 ]
机构
[1] Univ Med Ctr Mannheim, Dept Med Cardiol 1, Theodor Kutzer Ufer 1-3, D-68167 Mannheim, Germany
[2] German Ctr Cardiovasc Res DZHK, Partner Site Heidelberg Mannheim, Mannheim, Germany
关键词
Defibrillation threshold test; Generator replacement; Subcutaneous defibrillator; PRAETORIAN score; Ventricular fibrillation; IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; THRESHOLD; OUTCOMES; TIME;
D O I
10.1016/j.hrthm.2019.10.024
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Defibrillation threshold (DFT) testing is commonly performed in patients undergoing subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation. Growing evidence indicates that successful DFT testing correlates with the technique used for implantation of the defibrillation lead and pulse generator. However, evidence on whether DFT testing should be performed in patients undergoing S-ICD pulse generator replacement is lacking. OBJECTIVE The purpose of this study was to determine the outcome of DFT testing in patients undergoing replacement of an S-ICD pulse generator. METHODS A total of 357 S-ICD implantations were performed between November 2010 and July 2019. Twenty-five consecutive patients underwent S-ICD replacement between 2015 and 2019. Clinical data, perioperative medication, technical measurements, and PRAETORIAN score were assessed and grouped according to the outcome of DFT testing. RESULTS In 5 of 25 patients (20%), induced ventricular fibrillation was not successfully terminated after the first or second 65-J shock after pulse generator replacement with need for external defibrillation. Repositioning of the pulse generator and/or capsulectomy at the pocket site were necessary to achieve effective DFT with 65 J in all cases. Shock impedance increased in all patients at the time of pulse generator replacement compared to first implantation and was significantly higher in patients with ineffective DFT (119 +/- 17 Omega vs 93 +/- 26 Omega; P = .03). Otherwise, no differences in clinical characteristics, comorbidities, body mass index, intraoperative medication, or PRAETORIAN score were predictive of defibrillation failure. CONCLUSION The high proportion of patients with DFT failure after S-ICD pulse generator replacement indicates that DFT testing is mandatory to ensure safe function of the S-ICD.
引用
收藏
页码:455 / 459
页数:5
相关论文
共 50 条
  • [1] Successful defibrillation testing in patients undergoing elective subcutaneous implantable cardioverter-defibrillator generator replacement
    de Veld, Jolien A.
    Pepplinkhuizen, Shari
    van der Stuijt, Willeke
    Quast, Anne-Floor B. E.
    Nordkamp, Louise R. A. Olde
    Kooiman, Kirsten M.
    Wilde, Arthur A. M.
    Smeding, Lonneke
    Knops, Reinoud E.
    EUROPACE, 2023, 25 (07):
  • [2] Defibrillation testing of the subcutaneous implantable cardioverter-defibrillator at the time of generator replacement
    Schaller, Robert D.
    Hyman, Matthew
    Supple, Gregory E.
    Santangeli, Pasquale
    Riley, Michael P.
    Nazarian, Saman
    Arkles, Jeffrey
    Garcia, Fermin
    Lin, David
    Guandalini, Gustavo
    Kumareswaran, Ramanan
    Deo, Rajat
    Bode, Weeranun
    Markman, Timothy
    Epstein, Andrew
    Callans, David J.
    Dixit, Sanjay
    Brozoski, Joseph
    Marchlinski, Francis E.
    Frankel, David S.
    HEART RHYTHM, 2024, 21 (01) : 117 - 118
  • [3] CHRONIC DEFIBRILLATION THRESHOLD TESTING IN PATIENTS UNDERGOING SUBCUTANEOUS IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR GENERATOR CHANGE
    Patel, Raj
    Shahab, Ahmed
    Broman, Michael
    Aziz, Zaid
    Beaser, Andrew D.
    Upadhyay, Gaurav A.
    Kannan, Arun
    Jameria, Zenith
    Chen, Andrew
    Ozcan, Cevher
    Tung, Roderick
    Nayak, Hemal M.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (11) : 538 - 538
  • [4] Defibrillation threshold in elective subcutaneous implantable defibrillator generator replacements: Time to reduce the size of the pulse generator?
    Stuijt, W. van der
    Pepplinkhuizen, S.
    de Veld, J. A.
    Quast, A. B. E.
    van Halm, V. P.
    Bijsterveld, N. R.
    Nordkamp, L. R. A. Olde
    Wilde, A. A. M.
    Smeding, L.
    Knops, R. E.
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2024, 398
  • [5] Complications related to elective generator replacement of the subcutaneous implantable defibrillator
    van der Stuijt, Willeke
    Quast, Anne-Floor B. E.
    Baalman, Sarah W. E.
    de Wilde, Koen C.
    Brouwer, Tom F.
    Wilde, Arthur A. M.
    Knops, Reinoud E.
    EUROPACE, 2021, 23 (03): : 395 - 399
  • [6] Defibrillation Testing in Patients Undergoing Replacement of the S-ICD Generator: Is There Still a Need?
    Biller, Benedikt
    Reinke, Florian
    Biller, Katharina
    Koebe, Julia
    Rath, Benjamin
    Eckardt, Lars
    Frommeyer, Gerrit
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2025, 48 (01): : 3 - 8
  • [7] Clinical Profile and Incidence of Ventricular Arrhythmia in Patients Undergoing Defibrillator Generator Replacement in Spain
    Fontenla, Adolfo
    Lopez Gil, Maria
    Martinez Ferrer, Jose
    Alzueta, Javier
    Fernandez Lozano, Ignacio
    Vinolas, Xavier
    Rodriguez, Anibal
    Fernandez de la Concha, Joaquin
    Anguera, Ignasi
    Arribas, Fernando
    REVISTA ESPANOLA DE CARDIOLOGIA, 2014, 67 (12): : 986 - 992
  • [8] One-Year Mortality in Patients Undergoing an Implantable Cardioverter Defibrillator or Cardiac Resynchronization Therapy Pulse Generator Replacement: Identifying Patients at Risk
    Feijen, Michelle
    Egorova, Anastasia D.
    Kuijken, Teresa
    Bootsma, Marianne
    Schalij, Martin J.
    van Erven, Lieselot
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (17)
  • [9] Effects of defibrillation shock in patients implanted with a subcutaneous defibrillator: a biomarker study
    D'Onofrio, Antonio
    Russo, Vincenzo
    Bianchi, Vatter
    Cavallaro, Ciro
    Leonardi, Silvia
    De Vivo, Stefano
    Vecchione, Filippo
    Rago, Anna
    Ammendola, Ernesto
    Tavoletta, Vincenzo
    Atripaldi, Luigi
    Mocavero, Paola Elvira
    Nigro, Gerardo
    EUROPACE, 2018, 20 : F233 - F239
  • [10] PREDICTORS OF DEFIBRILLATION EFFICACY IN PATIENTS UNDERGOING EPICARDIAL DEFIBRILLATOR IMPLANTATION
    LEITCH, JW
    YEE, R
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (07) : 1632 - 1637