An immunosensor: Immobilization of anti-HBs antibody on glow-discharge treated piezoelectric quartz crystal for HBs-AG detection

被引:16
|
作者
Mutlu, S
Saber, R
Koçum, C
Piskin, E [1 ]
机构
[1] Hacettepe Univ, Dept Chem Engn, TR-06532 Ankara, Turkey
[2] Hacettepe Univ, Bioengn Div, TR-06532 Ankara, Turkey
关键词
immunosensor; piezoelectric quartz crystal; glow-discharge treatment; anti-HBs; HBs-Ag;
D O I
10.1080/00032719908542823
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, an immunosensor with a piezoelectric quartz crystal was developed. An oscillator system, in order to measure the frequency shifts caused by the increase of mass on the piezoelectric crystal, was used. The surfaces of the quartz crystals were treated by a glow-discharge technique utilizing ethylene diamine monomer to substitute amino groups on the surface of the crystal. These groups were activated by a bifunctional cross-linker, glutaraldehyde, to immobilize an antibody (so called anti-HBs). Optimum conditions for each of the modification steps for application on the quartz crystal surfaces were established and the results were represented for optimum time, pH and concentration. The sensor was calibrated at optimum conditions for HBs-Ag detection.