A preliminary study of the novel application of normoxic polymer gel dosimeters for the measurement of CTDI on diagnostic x-ray CT scanners

被引:46
|
作者
Hill, B
Venning, AJ
Baldock, C
机构
[1] Canberra Hosp, Biomed Engn Serv, Med Phys Sect, Woden, ACT 2606, Australia
[2] Univ Sydney, Inst Med Phys, Woden, ACT 2606, Australia
[3] Queensland Univ Technol, Sch Phys & Chem Sci, Brisbane, Qld 4001, Australia
[4] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
关键词
computer tomography; normoxic polymer gel dosimetry; dosimeter; CT; MRI; doseprofile;
D O I
10.1118/1.1925181
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Computer tomography dose index (CTDI) is a measurement undertaken during acceptance testing and subsequent quality assurance measurements of diagnostic x-ray CT scanners for the determination of patient dose. Normoxic polymer gel dosimeters have been used for the first time to measure dose and subsequently CTDI during acceptance testing of a CT scanner and compared with the conventional ionization chamber measurement for a range of imaging protocols. The normoxic polymer gel dosimeter was additionally used to simultaneously determine slice-width dose profiles and CTDI in the transaxial plane, the measurements of which are usually determined with thermoluminescent dosimetry or film. The resulting CTDI for all slice widths calculated from the normoxic polymer gel dosimeter were within corresponding ionization chamber CTDI values. Slice-width dose-profiles full-width half-maximum values from the normoxic polymer gel dosimeter were compared to the slice sensitivity profiles and were within the tolerances of the manufacturer. Normoxic polymer gel dosimeters have been shown to be a useful device for determining CTDI and dose distributions for CT equipment, and provide additional information not possible with just the use of an ionization chamber. (c) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:1589 / 1597
页数:9
相关论文
共 50 条
  • [41] Polymer gel dosimetry using x-ray computed tomography: a feasibility study
    Hilts, M
    Audet, C
    Duzenli, C
    Jirasek, A
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09): : 2559 - 2571
  • [42] Assessment of the effects of CT dose in averaged x-ray CT images of a dose-sensitive polymer gel
    Kairn, T.
    Kakakhel, M. B.
    Johnston, H.
    Jirasek, A.
    Trapp, J. V.
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [43] PRELIMINARY-STUDY ON APPLICATION OF X-RAY CT SCANNER TO MEASUREMENT OF VOID FRACTIONS IN STEADY-STATE 2-PHASE FLOWS
    IKEDA, T
    KOTANI, K
    MAEDA, Y
    KOHNO, H
    JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY, 1983, 20 (01) : 1 - 12
  • [44] An x-ray CT polymer gel dosimetry prototype: I. Remnant artefact removal
    Jirasek, A.
    Carrick, J.
    Hilts, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (10): : 3137 - 3153
  • [45] Influence of difference in cross-sectional dose profile in a CTDI phantom on X-ray CT dose estimation: A Monte Carlo study
    Haba T.
    Koyama S.
    Ida Y.
    Radiological Physics and Technology, 2014, 7 (1) : 133 - 140
  • [46] Preliminary Study of Temporal Performances of a Novel CVD Diamond X-Ray Detector
    Hou Lifei
    Du Huabing
    Che Xingsen
    Wei, Minxi
    Yu Bo
    Yang Guohong
    Liu Shenye
    ACTA OPTICA SINICA, 2018, 38 (07)
  • [47] Pin-photodiode array for the measurement of fan-beam energy and air kerma distributions of X-ray CT scanners
    Haba, Tomonobu
    Koyama, Shuji
    Aoyama, Takahiko
    Kinomura, Yutaka
    Ida, Yoshihiro
    Kobayashi, Masanao
    Kameyama, Hiroshi
    Tsutsumi, Yoshinori
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (07): : 905 - 913
  • [48] The Application of X-ray Micro-CT in the Study of HTS Tape Coils
    Minasyan, Vitaly B.
    Ivanov, Nikolay S.
    Malykh, Elizaveta A.
    Zanegin, Yuri A.
    Douine, Bruno
    INVENTIONS, 2022, 7 (03)
  • [49] X-ray interferometry without analyzer for breast CT application: a simulation study
    Xu, Jingzhu
    Ham, Kyungmin
    Dey, Joyoni
    JOURNAL OF MEDICAL IMAGING, 2020, 7 (02)
  • [50] APPLICATION OF X-RAY CT SCANNER TO MEASUREMENT OF VOID FRACTIONS IN 2-PHASE FLOWS
    IKEDA, T
    KOTANI, K
    MAEDA, Y
    MASUHARA, Y
    SUMIDA, I
    TRANSACTIONS OF THE AMERICAN NUCLEAR SOCIETY, 1982, 41 : 684 - 685