High-performance liquid chromatographic determination of cefepime in human plasma and in urine and dialysis fluid using a column-switching technique

被引:34
|
作者
Cherti, N
Kinowski, JM
Lefrant, JY
Bressolle, F
机构
[1] Univ Montpellier 1, Clin Pharmacokinet Lab, Montpellier, France
[2] Caremeau Hosp, Serv Pharm, Pharmacokinet Lab, Nimes, France
[3] Fac Pharm Montpellier, Federat Anesthesiol Intens Care & Emergency, F-34060 Montpellier 2, France
来源
JOURNAL OF CHROMATOGRAPHY B | 2001年 / 754卷 / 02期
关键词
cefepime;
D O I
10.1016/S0378-4347(00)00630-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A high-performance liquid chromatographic method with UV absorbance was developed for the analysis of cefepime in human plasma and urine, and in dialysis fluid. Detection was performed at 280 nm. The assay procedure for cefepime in plasma involves the addition of an internal standard (cefpirome) followed by treatment of the samples with trichloracetic acid, acetonitrile and dichloromethane. To quantify cefepime in diluted urine (1:20) and in the dialysis fluid, samples spiked with the internal standard (cefpirome) were analysed using a column-switching technique. The HPLC column, Nucleosil C-18, was equilibrated with an eluent mixture composed of acetonitrile-ammonium acetate (pH 4). Linear detector responses were observed for the calibration curve standards in the range 0.5 to 100 mug/ml, which spans what is currently thought to be the clinically relevant range for cefepime concentrations in body fluids. The limit of quantification was 0.5 mug/ml in the three matrices. Extraction recoveries proved to be more than 84%. Precision, expressed as %RSD, was in the range 1.5 to 9%. Accuracy ranged from 93 to 105%. This method was used to follow the time course of the concentration of cefepime in plasma, urine and dialysate outlet samples after a 10-min infusion period of 2 g of this drug in patients with acute renal failure undergoing hemodiafiltration. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:377 / 386
页数:10
相关论文
共 50 条
  • [31] AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC AND COLUMN-SWITCHING TECHNIQUE FOR ONLINE CLEANUP AND ANALYSIS OF DILTIAZEM IN HUMAN-PLASMA
    ASCALONE, V
    DALBO, L
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 423 : 239 - 249
  • [32] Direct analysis of indomethacin in rat plasma using a column-switching high-performance liquid chromatographic system
    Liu, SC
    Kamijo, M
    Takayasu, T
    Takayama, S
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 767 (01): : 53 - 60
  • [33] High-performance liquid chromatographic determination of finasteride in human plasma using direct injection with column switching
    Takano, T
    Hata, S
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 676 (01): : 141 - 146
  • [34] AUTOMATED-DETERMINATION OF MIDAZOLAM IN HUMAN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING COLUMN-SWITCHING
    LAUBER, R
    MOSIMANN, M
    BUHRER, M
    ZBINDEN, AM
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 654 (01): : 69 - 75
  • [35] Determination of angiotensin metabolites in human plasma by fluorimetric high-performance liquid chromatography using a heart-cut column-switching technique
    Matsui, T
    Tamaya, K
    Kawasaki, T
    Osajima, Y
    JOURNAL OF CHROMATOGRAPHY B, 1999, 729 (1-2): : 89 - 95
  • [36] Automated high-performance liquid chromatographic determination of amphetamine in biological fluids using column-switching and on-column derivatization
    R. Herráez-Hernández
    P. Campíns-Falcó
    S. Díaz-Oltra
    Chromatographia, 1999, 49 : 188 - 196
  • [37] Automated high-performance liquid chromatographic determination of amphetamine in biological fluids using column-switching and on-column derivatization
    Herráez-Hernández, R
    Campíns-Falcó, P
    Díaz-Oltra, S
    CHROMATOGRAPHIA, 1999, 49 (3-4) : 188 - 196
  • [38] AUTOMATED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF SPIRAMYCIN BY DIRECT INJECTION OF PLASMA, USING COLUMN-SWITCHING FOR SAMPLE CLEANUP
    DOW, J
    LEMAR, M
    FRYDMAN, A
    GAILLOT, J
    JOURNAL OF CHROMATOGRAPHY, 1985, 344 (NOV): : 275 - 283
  • [39] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF SPIRONOLACTONE AND ITS MAJOR METABOLITE CANRENONE IN URINE USING ULTRAVIOLET DETECTION AND COLUMN-SWITCHING
    HERRAEZHERNANDEZ, R
    SORIANOVEGA, E
    CAMPINSFALCO, P
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1994, 658 (02): : 303 - 310
  • [40] High-performance liquid chromatographic determination of cefepime and cefazolin in human plasma and dialysate
    Farthing, Christine
    Farthing, Don
    Brophy, Donald F.
    Larus, Terri
    Maynor, Lena
    Fakhry, Itaf
    Gehr, Todd W. B.
    CHROMATOGRAPHIA, 2008, 67 (5-6) : 365 - 368