X-ray spectroscopy applied to radiation shielding calculation in mammography

被引:4
|
作者
Kuenzel, Roseli
Levenhagen, Ronaldo Savarino [1 ]
Herdade, Silvio Bruni [2 ]
Terini, Ricardo Andrade [3 ]
Costa, Paulo Roberto [4 ]
机构
[1] Ctr Univ Franciscano, Area Ciencias Nat & Tecnol, BR-97010032 Santa Maria, RS, Brazil
[2] Univ Sao Paulo, Inst Fis, BR-05508900 Sao Paulo, Brazil
[3] Pontificia Univ Catolica Sao Paulo, Dept Fis, BR-01303050 Sao Paulo, Brazil
[4] Univ Sao Paulo, Inst Eletrotecn & Energia, Secao Tecn Desenvolvimento Tecnol Saude, BR-05508010 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
x-ray spectroscopy; scattered radiation; air kerma; shielding design; scatter fraction;
D O I
10.1118/1.2953566
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The protective shielding design of a mammography facility requires the knowledge of the scattered radiation by the patient and image receptor components. The shape and intensity of secondary x-ray beams depend on the kVp applied to the x-ray tube, target/filter combination, primary x-ray field size, and scattering angle. Currently, shielding calculations for mammography facilities are performed based on scatter fraction data for Mo/Mo target/filter, even though modern mammography equipment is designed with different anode/filter combinations. In this work we present scatter fraction data evaluated based on the x-ray spectra produced by a Mo/Mo, Mo/Rh and W/Rh target/filter, for 25, 30 and 35 kV tube voltages and scattering angles between 30 and 165 degrees. Three mammography phantoms were irradiated and the scattered radiation was measured with a CdZnTe detector. The primary x-ray spectra were computed with a semiempirical model based on the air kerma and HVL measured with an ionization chamber. The results point out that the scatter fraction values are higher for W/Rh than for Mo/Mo and Mo/Rh, although the primary and scattered air kerma are lower for W/Rh than for Mo/Mo and Mo/Rh target/filter combinations. The scatter fractions computed in this work were applied in a shielding design calculation in order to evaluate shielding requirements for each of these target/filter combinations. Besides, shielding requirements have been evaluated converting the scattered air kerma from mGy/week to mSv/week adopting initially a conversion coefficient from air kerma to effective dose as 1 Sv/Gy and then a mean conversion coefficient specific for the x-ray beam considered. Results show that the thickest barrier should be provided for Mo/Mo target/filter combination. They also point out that the use of the conversion coefficient from air kerma to effective dose as 1 Sv/Gy is conservatively high in the mammography energy range and overestimate the barrier thickness. (c) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:3539 / 3545
页数:7
相关论文
共 50 条
  • [21] Radiation detectors for X-ray and gamma-ray spectroscopy
    Knoll, GF
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2000, 243 (01) : 125 - 131
  • [22] Characterization of the radiation shielding properties of fired lead sample for X-ray shielding applications
    Umeh, Chibuike Dominic
    Agwu, Kenneth Kalu
    Okoye, Chukwuemeka Madubuko Ijeoma
    Ahia, Christian Chinedu
    Ikegbu, Godswill Ogbu
    PROGRESS IN NUCLEAR ENERGY, 2021, 137
  • [23] Energy dependence of the response of X-ray multimeter for radiation qualities in mammography
    Elisabeth Salomon
    Peter Homolka
    Istvan Csete
    Paula Toroi
    Scientific Reports, 15 (1)
  • [24] A Mathematical Model of Photon Scattering in Matter for Problems of Calculation and Optimization of Radiation Shielding in X-Ray Inspection Systems
    Petrunin, V. I.
    Ogorodnikov, S. A.
    Arlychev, M. A.
    Shevelev, I. E.
    MOSCOW UNIVERSITY PHYSICS BULLETIN, 2015, 70 (02) : 118 - 124
  • [25] A mathematical model of photon scattering in matter for problems of calculation and optimization of radiation shielding in X-ray inspection systems
    V. I. Petrunin
    S. A. Ogorodnikov
    M. A. Arlychev
    I. E. Shevelev
    Moscow University Physics Bulletin, 2015, 70 : 118 - 124
  • [26] X-RAY SPECTROSCOPY OF MONOCRYSTALS WITH UTILIZATION OF POLARIZED RADIATION
    DREGER, G
    BRYUMMER, O
    IZVESTIYA AKADEMII NAUK SSSR SERIYA FIZICHESKAYA, 1976, 40 (11): : 2437 - 2440
  • [27] Bismuth and thorium fluorides as efficient X-ray radiation shielding materials
    Bukhvalova, Svetlana Yu
    Asmolova, Nina F.
    Lopatina, Tatiana I.
    Bochkarev, Mikhail N.
    Radiation Physics and Chemistry, 2021, 182
  • [28] Bismuth and thorium fluorides as efficient X-ray radiation shielding materials
    Bukhvalova, Svetlana Yu
    Asmolova, Nina F.
    Lopatina, Tatiana, I
    Bochkarev, Mikhail N.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 182
  • [29] Advancing radiation shielding: A review the role of Bismuth in X-ray protection
    Bakri, Fahrul
    Gareso, Paulus Lobo
    Tahir, Dahlang
    RADIATION PHYSICS AND CHEMISTRY, 2024, 217
  • [30] RADIATION SHIELDING OF MULTIPLE X-RAY SOURCES IN DIAGNOSTIC-RADIOLOGY
    SIMPKIN, DJ
    HEALTH PHYSICS, 1986, 50 (01): : 117 - 122