Performance of a small animal PET scanner based on photosensitive wire chambers.

被引:0
|
作者
Bruyndonckx, P
Liu, X
Tavernier, S
Zhang, SP
机构
关键词
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A 3D small animal PET scanner, designed and built at the University of Brussels, has been operational since the beginning of 1996. The scanner has a transaxial field of view (FOV) of 110 mm and an axial FOV of 51 mm. The absolute sensitivity is around 35.000 coincidences/s/MBq for a point source at the center of the scanner and the time resolution is 29 ns FWHM. To measure the achievable spatial resolution, a thin Na-22 source was scanned at various distances from the scanner axis. The resolution in a reconstructed image for a source close to the center is 3.0 mm FWHM. Finally, to assess the overall image quality, scans were made of a cylindrical phantom with holes of varying diameter.
引用
收藏
页码:1335 / 1338
页数:4
相关论文
共 50 条
  • [41] Design and Calibration of a Small Animal PET Scanner Based on Continuous LYSO Crystals and PSPMTs
    Benlloch, J.
    Carrilero, V.
    Catret, J.
    Colom, R.
    Correcher, C.
    Gadea, R.
    Garcia Quiros, F.
    Gonzalez, A.
    Herrero, V.
    Lerche, Ch. W.
    Mora, F. J.
    Mora, C.
    Morera, C.
    Munar, A.
    Pavon, N.
    Ros, A.
    Sanchez, F.
    Sebastia, A.
    Vidal, L.
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2328 - 2332
  • [42] An MR compatible PET scanner based on RatCAP for small animal imaging at 9.4 T
    Maramraju, S. H.
    Junnarkar, S.
    Ravindranath, B.
    Southekal, S.
    Stoll, S.
    Smith, S. D.
    Lenz, W.
    Pratte, J. -F.
    Purschke, M.
    Rescia, S.
    Krishnamoorthy, S.
    Vaska, P.
    Woody, C.
    Schlyer, D.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 2954 - +
  • [43] Temperature dependence of the LabPET small-animal PET scanner
    Keereman, Vincent
    Van Holen, Roel
    Vanhove, Christian
    Vandenberghe, Stefaan
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [44] Measurements for the SiliPET Project: a Small Animal PET Scanner Based on Stacks of Silicon Detectors
    Auricchio, N.
    Di Domenico, G.
    Milano, L.
    Malaguti, R.
    Ambrosi, G.
    Ionica, M.
    Fiandrini, E.
    Zavattini, G.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3534 - +
  • [45] New PET scanner for small animal imaging based on innovative straw detector technology
    Lacy, Jeffrey
    Sun, Liang
    Athanasiades, Athanasios
    Martin, Chris
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [46] Development of a prototype of time-over-threshold based small animal PET scanner
    Shimazoe, K.
    Takahashi, H.
    Kamada, K.
    Yoshikawa, A.
    Kumagai, K.
    Kataoka, J.
    Itoh, S.
    Sato, H.
    Usuki, Y.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014, 753 : 84 - 90
  • [47] Attenuation correction for the NIH ATLAS small animal PET scanner
    Yao, RT
    Seidel, J
    Liow, JS
    Green, MV
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) : 664 - 668
  • [48] A High-Sensitivity Small-Animal PET Scanner: Development and Initial Performance Measurements
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2678 - 2688
  • [49] Analog processing board for the MiCES small animal PET scanner
    Park, B
    Lewellen, TK
    Miyaoka, RS
    Janes, M
    Kinahan, P
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3920 - 3923
  • [50] System integration of the MiCES small animal PET scanner.
    Lewellen, TK
    Janes, M
    Miyaoka, RS
    Gillespie, SB
    Park, B
    Lee, KS
    Kinahan, P
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3316 - 3320