Reliability of corneal thickness and endothelial cell density measures

被引:0
|
作者
Nichols, JJ [1 ]
Kosunick, GM [1 ]
Bullimore, MA [1 ]
机构
[1] Ohio State Univ, Coll Optometry, Columbus, OH 43210 USA
关键词
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: We evaluated the reliability and agreement between the Orbscan, an ultrasonic pachymeter (Humphrey 855), and the Konan SP 9000-LC in terms of central corneal thickness. The Konan was also used to study the reliability and agreement between endothelial cell density measures. METHODS: Twenty-five normal subjects were examined on two occasions (mean separation = 9 +/- 5 days) by a single examiner using all three instruments for central corneal pachymetry. The Konan Center Method and a manual counting method were performed by two examiners to determine endothelial cell density. Reliability and agreement were assessed by calculating the 95% limits of agreement (LoA) and intraclass correlation coefficients (ICC). RESULTS: For corneal pachymetry test-retest reliability, the 95% limits of agreement were -20 to +17 mum for the ultrasound, -27 to +22 mum for the Konan, and -13 to +13 mum for the Orbscan. There was fair-to-good agreement between the pachymeters (intraclass correlation coefficients range = 0.85 to 0.92). For endothelial cell density test-retest reliability, the 95% limits of agreement for the Konan Center Method was -498 to +530, and -482 to +333 cells/mm(2) for examiners 1 and 2, respectively. The test-retest 95% limits of agreement for the manual overlaid grid method was -355 to +355, and -535 to +670 cells/mm(2) for examiners 1 and 2, respectively. CONCLUSIONS: The reliability and agreement of the Orbscan and Konan corneal pachymeters was good, although the reliability of the Konan for estimating endothelial cell density was fair, at best.
引用
收藏
页码:344 / 352
页数:9
相关论文
共 50 条
  • [31] Corneal thickness and endothelial density before and after cataract surgery
    Ventura, ACS
    Wälti, R
    Böhnke, M
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (01) : 18 - 20
  • [32] Influence of dorzolamide on corneal thickness, endothelial cell count and corneal sensibility
    Kaminski, S
    Hommer, A
    Koyuncu, D
    Biowski, R
    Barisani, T
    Baumgartner, I
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 1998, 76 (01): : 78 - 79
  • [33] Corneal endothelial cell counts and corneal thickness in normal horses.
    Andrew, SE
    Ramsey, DT
    Brooks, DE
    Hauptman, JG
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S445 - S445
  • [34] EFFECTS OF PHACOEMULSIFICATION AND EXTRACAPSULAR LENS REMOVAL ON CORNEAL THICKNESS AND ENDOTHELIAL CELL-DENSITY IN THE DOG
    GWIN, RM
    WARREN, JK
    SAMUELSON, DA
    GUM, GG
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1983, 24 (02) : 227 - 236
  • [35] Corneal Endothelial Cell Density of Cytomegalovirus Corneal Endotheiitis Cases
    Ikeda, Toshihide
    Fukuoka, Hideki
    Kitazawa, Koji
    Inatomi, Tsutomu
    Koizumi, Noriko
    Sotozono, Chie
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2024, 65 (07)
  • [36] Corneal Endothelial Cell Proliferation: A Function of Cell Density
    Patel, Sangita P.
    Bourne, William M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (06) : 2742 - 2746
  • [37] Corneal thickness and endothelial cell count after phacoemulsification
    Kohlhaas, M
    Stahlhut, O
    Tholuck, J
    Richard, G
    OPHTHALMOLOGE, 1997, 94 (07): : 515 - 518
  • [38] CORNEAL ENDOTHELIAL CELL-DENSITY IN TWINS
    PRICE, NC
    BARBOUR, DJ
    BRITISH JOURNAL OF OPHTHALMOLOGY, 1981, 65 (12) : 812 - 814
  • [39] EFFECT OF PHACOEMULSIFICATION ON CORNEAL ENDOTHELIAL CELL DENSITY
    SUGAR, J
    MITCHELSON, J
    KRAFF, M
    ARCHIVES OF OPHTHALMOLOGY, 1978, 96 (03) : 446 - 448
  • [40] INDIRECT EVALUATION OF CORNEAL ENDOTHELIAL CELL DENSITY
    SPERLING, S
    ACTA OPHTHALMOLOGICA, 1978, 56 (03): : 445 - 454