Frequency Synchronization Analysis in Digital lock-in Methods for Bio-impedance Determination

被引:3
|
作者
Brajkovic, Robert [1 ]
Zagar, Tomaz [2 ]
Krizaj, Dejan [1 ]
机构
[1] Univ Ljubljana, Fac Elect Engn, Ljubljana 1000, Slovenia
[2] Geodet Inst Slovenia, Ljubljana 1000, Slovenia
来源
MEASUREMENT SCIENCE REVIEW | 2014年 / 14卷 / 06期
关键词
Impedance; bioimpedance measurements; frequency synchronization; lock-in method; PHASE-SENSITIVE DETECTION;
D O I
10.2478/msr-2014-0047
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The lock-in method is one of the most frequently used methods for reconstruction of measured signals and as such frequently applied in the (bio) impedance method to determine the modulus and phase of the (bio) impedance. In implementation of the method in a (bio) impedance measurement device one has to consider possible non synchronized frequencies of the reference and the analyzed signals as well as potential sources of noise. In this work we analyzed these errors theoretically and experimentally. We show that both amplitude and phase errors depend on the relative difference of the frequencies of the reference and investigated signal as well as the number of integration periods. Theoretically, these errors vanish during the determination of the (bio) impedance modulus and phase. In practical implementation the inaccuracies appear at points of very low determined signal amplitudes due to the limited accuracy of analog to digital converters and are distributed around these points due to other sources of noise inherent in implementation of the measurement device.
引用
收藏
页码:343 / 349
页数:7
相关论文
共 50 条
  • [31] Determination of Freshness of Freshwater Fish Based on BP-ANN and Bio-impedance Characteristics
    Zhang Jun
    Li XiaoYu
    Wang Wei
    Zhou Zhu
    PROCEEDINGS OF THE 2009 WRI GLOBAL CONGRESS ON INTELLIGENT SYSTEMS, VOL IV, 2009, : 68 - +
  • [32] Mixing frequency bio-impedance measurement technology based on DFT and virtual reference vector
    Wang Chao
    Huang Chunyan
    Zhang Xiaoli
    Wang Huaxiang
    BMEI 2008: PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOL 2, 2008, : 455 - +
  • [33] Fully-digital low-frequency lock-in amplifier for photoluminescence measurements
    Pollastrone, Fabio
    Piccinini, Massimo
    Pizzoferrato, Roberto
    Palucci, Antonio
    Montereali, Rosa Maria
    ANALOG INTEGRATED CIRCUITS AND SIGNAL PROCESSING, 2023, 115 (01) : 67 - 76
  • [34] Gastric emptying assessment in frequency and time domain using bio-impedance:: Preliminary results
    Huerta-Franco, R.
    Vargas-Luna, M.
    Hernandez, E.
    Cordova, T.
    Sosa, M.
    Gutierrez, G.
    Reyes, P.
    Mendiola, C.
    MEDICAL PHYSICS, 2006, 854 : 209 - +
  • [35] Fully-digital low-frequency lock-in amplifier for photoluminescence measurements
    Fabio Pollastrone
    Massimo Piccinini
    Roberto Pizzoferrato
    Antonio Palucci
    Rosa Maria Montereali
    Analog Integrated Circuits and Signal Processing, 2023, 115 : 67 - 76
  • [36] Frequency lock-in and phase synchronization of vortex shedding behind circular cylinder due to surface waves
    Gunnoo, Hans
    Abcha, Nizar
    Ezersky, Alexander
    PHYSICS LETTERS A, 2016, 380 (7-8) : 863 - 868
  • [37] Measurement of mutual inductance from the frequency dependence of impedance of AC coupled circuits using a digital dual-phase lock-in amplifier
    Schauber, Michael J.
    Newman, Seth A.
    Goodman, Lindsey R.
    Suzuki, Itsuko S.
    Suzuki, Masatsugu
    AMERICAN JOURNAL OF PHYSICS, 2008, 76 (02) : 129 - 132
  • [38] Hardware Design and Implementation of Digital Lock-in Amplifier for Multi-channel Parallel Impedance Measurement
    Xu, Jianze
    Cui, Zigiang
    Sun, Can
    Wang, Huaxiang
    2024 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, I2MTC 2024, 2024,
  • [39] Wrist-worn Heartbeat Monitoring System Based on Bio-Impedance Analysis
    Xu, Jia
    Gao, Xiaomeng
    Lee, Alexander
    Yamada, Shuhei
    Yavari, Ehsan
    Lubecke, Victor
    Boric-Lubecke, Olga
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 6294 - 6297
  • [40] Low-cost bio-impedance analysis system for the evaluation of fruit ripeness
    Ibba, Pietro
    Falco, Aniello
    Rivadeneyra, Almudena
    Lugli, Paolo
    2018 IEEE SENSORS, 2018, : 1468 - 1471