Recharging Difficulty With Pulse Generator After Deep Brain Stimulation: A Case Series of Five Patients

被引:3
|
作者
Li, Hongyang [1 ,2 ]
Su, Daoqing [3 ]
Lai, Yijie [1 ,2 ]
Xu, Xinmeng [1 ,2 ]
Zhang, Chencheng [1 ,2 ]
Sun, Bomin [1 ,2 ]
Li, Dianyou [1 ,2 ]
Pan, Yixin [1 ,2 ]
机构
[1] Shanghai Jiao Tong Univ, Ruijin Hosp, Dept Neurosurg, Sch Med, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Ruijin Hosp, Ctr Funct Neurosurg, Sch Med, Shanghai, Peoples R China
[3] Shandong First Med Univ, Liaocheng Peoples Hosp, Dept Neurosurg, Liaocheng Clin Sch, Liaocheng, Shandong, Peoples R China
关键词
hardware complication; deep brain stimulation; implanted pulse generator; rechargeable battery; Parkinson's disease; TWIDDLERS SYNDROME;
D O I
10.3389/fnins.2021.705483
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Background: Deep brain stimulation (DBS) is a well-established treatment for a variety of movement disorders. Rechargeable cell technology was introduced to pulse generator more than 10 years ago and brought great benefits to patients. However, with the widespread use of rechargeable implanted pulse generators (r-IPGs), a new hardware complication, when charging the r-IPG has been difficult, was encountered. Objective: The aims of this study were to report five cases confronted with r-IPG charging difficulty postoperatively and to explore the predisposing factors and treatment strategies for this rare complication. Methods: We retrospectively reviewed our DBS patient database for those who were implanted with r-IPGs. From 2012, we identified a total of 1,226 patients, with five of them experiencing charging difficulties after surgery. Detailed patient profiles and clinical procedures were scrutinized and reviewed. Results: All the charging problems were resolved by reoperation. Cases 1 and 2 required their r-IPGs to be anchored to the muscle and fascia. Cases 3 and 4 had their r-IPGs inserted in the wrong orientation at the initial surgery, which was resolved by turning around the r-IPGs at the revision surgery. Case 5, in which we propose that the thick subcutaneous fat layer blocked the connection between the r-IPG and the recharger, required a second operation to reposition the r-IPG in a shallow layer underneath the skin. For all cases, the charging problems were resolved without reoccurrences to date. Conclusion: Our case series indicates a novel hardware complication of DBS surgery, which had been rarely reported before. In this preliminary study, we describe several underlying causes of this complication and treatment methods.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Functional disorders in Parkinson's disease patients with deep brain stimulation - a case series
    Papic, E.
    Racki, V.
    Vuletic, V.
    EUROPEAN JOURNAL OF NEUROLOGY, 2023, 30 : 620 - 620
  • [22] The effect of deep brain stimulation surgery on repetitive behavior in Parkinson patients: A case series
    Wood, Mary F.
    Nguyen, Fredrick N.
    Okun, Michael S.
    Rodriguez, Ramone L.
    Foote, Kelly D.
    Fernandez, Hubert H.
    NEUROCASE, 2010, 16 (01) : 31 - 36
  • [23] Deep brain stimulation in five patients with severe disorders of consciousness
    Lemaire, Jean-Jacques
    Sontheimer, Anna
    Pereira, Bruno
    Coste, Jerome
    Rosenberg, Sarah
    Sarret, Catherine
    Coll, Guillaume
    Gabrillargues, Jean
    Jean, Betty
    Gillart, Thierry
    Coste, Aurelien
    Roche, Basile
    Kelly, Antony
    Pontier, Benedicte
    Feschet, Fabien
    ANNALS OF CLINICAL AND TRANSLATIONAL NEUROLOGY, 2018, 5 (11): : 1372 - 1384
  • [24] Preliminary Experience with a Four-Lead Implantable Pulse Generator for Deep Brain Stimulation
    Rusheen, Aaron E.
    Jensen, Michael A.
    Gregg, Nicholas M.
    Kaufmann, Timothy J.
    Van Gompel, Jamie J.
    Lee, Kendall H. H.
    Klassen, Bryan T.
    Miller, Kai J.
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2023, 101 (04) : 254 - 264
  • [25] Pulse generator battery life in deep brain stimulation: out with the old ... in with the less durable?
    de Schlichting, Emmanuel
    Coll, Guillaume
    Zaldivar-Jolissaint, Julien Francisco
    Coste, Jerome
    Marques, Ana Raquel
    Mulliez, Aurelien
    Durif, Franck
    Lemaire, Jean-Jacques
    ACTA NEUROCHIRURGICA, 2019, 161 (10) : 2043 - 2046
  • [26] Cosmetic considerations for placement of deep brain stimulation pulse generator: the submammary subfascial approach
    Jason Yuen
    Khaled Alameddine
    Eugene S. Bah
    Kendall H. Lee
    Basel A. Sharaf
    Acta Neurochirurgica, 2023, 165 : 735 - 739
  • [27] Pulse generator battery life in deep brain stimulation: out with the old… in with the less durable?
    Emmanuel de Schlichting
    Guillaume Coll
    Julien Francisco Zaldivar-Jolissaint
    Jérôme Coste
    Ana Raquel Marques
    Aurélien Mulliez
    Franck Durif
    Jean-Jacques Lemaire
    Acta Neurochirurgica, 2019, 161 : 2043 - 2046
  • [28] Thalamic local field potentials recorded using the deep brain stimulation pulse generator
    Buijink, A. W. G.
    Pina-Fuentes, D. A.
    Stam, M. J.
    Bot, M.
    Schuurman, P. R.
    van den Munckhof, P.
    van Rootselaar, A. F.
    de Bie, R. M. A.
    Beudel, M.
    CLINICAL NEUROPHYSIOLOGY PRACTICE, 2022, 7 : 103 - 106
  • [29] Investigation of Deep Brain Stimulation Mechanisms During Implantable Pulse Generator Replacement Surgery
    Swan, Brandon D.
    Grill, Warren M.
    Turner, Dennis A.
    NEUROMODULATION, 2014, 17 (05): : 419 - 424
  • [30] Life time of implanted programmable pulse generator for subthalamic deep-brain stimulation
    Anheim, M.
    Fraix, V.
    Chabardes, S.
    Krack, P.
    Benabid, A.
    Pollak, P.