Recharging Difficulty With Pulse Generator After Deep Brain Stimulation: A Case Series of Five Patients

被引:3
|
作者
Li, Hongyang [1 ,2 ]
Su, Daoqing [3 ]
Lai, Yijie [1 ,2 ]
Xu, Xinmeng [1 ,2 ]
Zhang, Chencheng [1 ,2 ]
Sun, Bomin [1 ,2 ]
Li, Dianyou [1 ,2 ]
Pan, Yixin [1 ,2 ]
机构
[1] Shanghai Jiao Tong Univ, Ruijin Hosp, Dept Neurosurg, Sch Med, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Ruijin Hosp, Ctr Funct Neurosurg, Sch Med, Shanghai, Peoples R China
[3] Shandong First Med Univ, Liaocheng Peoples Hosp, Dept Neurosurg, Liaocheng Clin Sch, Liaocheng, Shandong, Peoples R China
关键词
hardware complication; deep brain stimulation; implanted pulse generator; rechargeable battery; Parkinson's disease; TWIDDLERS SYNDROME;
D O I
10.3389/fnins.2021.705483
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Background: Deep brain stimulation (DBS) is a well-established treatment for a variety of movement disorders. Rechargeable cell technology was introduced to pulse generator more than 10 years ago and brought great benefits to patients. However, with the widespread use of rechargeable implanted pulse generators (r-IPGs), a new hardware complication, when charging the r-IPG has been difficult, was encountered. Objective: The aims of this study were to report five cases confronted with r-IPG charging difficulty postoperatively and to explore the predisposing factors and treatment strategies for this rare complication. Methods: We retrospectively reviewed our DBS patient database for those who were implanted with r-IPGs. From 2012, we identified a total of 1,226 patients, with five of them experiencing charging difficulties after surgery. Detailed patient profiles and clinical procedures were scrutinized and reviewed. Results: All the charging problems were resolved by reoperation. Cases 1 and 2 required their r-IPGs to be anchored to the muscle and fascia. Cases 3 and 4 had their r-IPGs inserted in the wrong orientation at the initial surgery, which was resolved by turning around the r-IPGs at the revision surgery. Case 5, in which we propose that the thick subcutaneous fat layer blocked the connection between the r-IPG and the recharger, required a second operation to reposition the r-IPG in a shallow layer underneath the skin. For all cases, the charging problems were resolved without reoccurrences to date. Conclusion: Our case series indicates a novel hardware complication of DBS surgery, which had been rarely reported before. In this preliminary study, we describe several underlying causes of this complication and treatment methods.
引用
收藏
页数:5
相关论文
共 50 条
  • [1] Alternative Method for the Recharging System of the New Implantable Pulse Generator of Deep Brain Stimulation in Chinese Patients
    Lau, Claire K. Y.
    Wong, Rosanna K. M.
    Chan, Danny T. M.
    Zhu, X. L.
    Poon, W. S.
    JOURNAL OF NEUROSCIENCE NURSING, 2012, 44 (03) : E1 - E3
  • [2] Reuse of Internal Pulse Generator in Cases of Infection after Deep Brain Stimulation Surgery
    Gocmen, Selcuk
    Celiker, Ozkan
    Topcu, Abdullah
    Panteli, Aikaterini
    Acar, Goksemin
    Acar, Feridun
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2014, 92 (03) : 140 - 144
  • [3] Intraoperative Test Stimulation with a Modified Implantable Pulse Generator in Deep Brain Stimulation
    J. K. Krauss
    T. Pohle
    Acta Neurochirurgica, 2000, 142 : 587 - 589
  • [4] Intraoperative test stimulation with a modified implantable pulse generator in deep brain stimulation
    Krauss, JK
    Pohle, T
    ACTA NEUROCHIRURGICA, 2000, 142 (05) : 587 - 589
  • [5] Telemedicine in patients with obsessive-compulsive disorder after deep brain stimulation: a case series
    Wan, Xiaonan
    Lin, Zhengyu
    Zeng, Zhitong
    Zhang, Yingying
    Duan, Chengcheng
    Zhang, Chencheng
    Li, Dianyou
    FRONTIERS IN HUMAN NEUROSCIENCE, 2024, 18
  • [6] A Rare Case of Green Gelatinous Mass Formation on a Deep Brain Stimulation Implantable Pulse Generator
    Poortvliet, Peter C.
    Mellick, George
    Coyne, Terry
    Silburn, Peter
    JOURNAL OF MOVEMENT DISORDERS, 2021, 14 (01) : 81 - 83
  • [7] Transaxillary Subpectoral Implantation of Implantable Pulse Generator for Deep Brain Stimulation
    Son, Byung-chul
    Choi, Yoon-suk
    Lee, Sang-Won
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2013, 91 : 193 - 193
  • [8] Transaxillary Subpectoral Implantation of Implantable Pulse Generator for Deep Brain Stimulation
    Son, Byung-chul
    Han, Soo-Hyung
    Choi, Yoon-suk
    Kim, Hyung-suk
    Kim, Min-cheol
    Yang, Seung-ho
    Lee, Sang-won
    NEUROMODULATION, 2012, 15 (03): : 260 - 266
  • [9] Replacement of a deep brain stimulation implantable pulse generator with a rechargeable device
    Chan, Danny Tat-Ming
    Zhu, Xian-Lun
    Lau, Claire Ka-Yee
    Wong, Rosanna Kit-Man
    Poon, Wai-Sang
    SURGICAL PRACTICE, 2014, 18 (03) : 140 - 142