Physical parameters and correction factors for ionization chambers for absolute measurement of air kerma in γ-ray fields

被引:2
|
作者
Takata, N. [1 ]
Kurosawa, T.
Begum, A.
Sugita, T.
机构
[1] AIST, NMIJ, Ionizing Radiat Sect, Tsukuba, Ibaraki 3058568, Japan
[2] BUET, Dept Phys, Dhaka 1000, Bangladesh
[3] Syst Sci Lab, Ibaraki 30917, Japan
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2007年 / 580卷 / 01期
关键词
air kerma; ionization chambers; mean mass stopping power ratio; wall correction factor; Cs-137; gamma-rays; Co-60;
D O I
10.1016/j.nima.2007.05.175
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Values of physical parameters and correction factors essential for the absolute measurement of air kerma in Cs-137 and Co-60 gamma-ray fields were obtained using an EGS5 program for spherical, cylindrical and pancake ionization chambers. The mean mass collision stopping power ratio for graphite and air, was found to vary depending on the cutoff energy of electrons employed in calculation. The ratio between the energies deposited in cavity air due to Compton electrons emitted from the air and those from the graphite wall increases as the chamber size is increased. It also increases as the gamma-ray energy is reduced and is equal to 0.09 for Cs-137 gamma-rays in a spherical ionization chamber of cavity diameter 12 cm. Correction factors for gamma-ray attenuation in chamber walls and those for the contribution of scattered gamma-rays to chamber responses were obtained separately. The wall correction factor, which is equal to the product of these two factors, is close to unity for pancake chambers. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:346 / 349
页数:4
相关论文
共 50 条
  • [42] Diaphragm correction factors for free-air chamber standards for air kerma in x-rays
    Burns, D. T.
    Kessler, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09): : 2737 - 2745
  • [43] CORRECTION FACTORS FOR CALIBRATION OF PLANE-PARALLEL IONIZATION CHAMBERS WITH A CO-60 GAMMA-RAY BEAM
    LAITANO, RF
    GUERRA, AS
    PIMPINELLA, M
    NYSTROM, H
    KARLSSON, M
    SVENSSON, H
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (01): : 39 - 54
  • [44] Recombination correction for ionization chambers calibrated by PTB in Co-60 radiation fields
    Derikum, K
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S198 - S198
  • [45] Air kerma rate measurements from a miniature X-ray source using free-air ionization
    Davis, S.
    Micka, J.
    DeWerd, L.
    Rusch, T.
    MEDICAL PHYSICS, 2006, 33 (06) : 1983 - 1984
  • [46] MEASUREMENT OF FIELD DISTORTION IN FREE-AIR IONIZATION CHAMBERS BY ANALOG METHOD
    MILLER, W
    KENNEDY, RJ
    JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS, 1955, 55 (05): : 291 - 297
  • [47] Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel
    Casar, Bozidar
    Gershkevitsh, Eduard
    Mendez, Ignasi
    Jurkovic, Slaven
    Huq, M. Saiful
    MEDICAL PHYSICS, 2020, 47 (01) : 242 - 259
  • [48] Monte Carlo calculation of magnetic field correction factors for two ionization chambers
    Alissa, M.
    Zink, K.
    Schoenfeld, A. A.
    Czarneck, D.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1310 - S1311
  • [49] Calculated beam quality correction factors for ionization chambers in MV photon beams
    Tikkanen, J.
    Zink, K.
    Pimpinella, M.
    Teles, P.
    Borbinha, J.
    Ojala, J.
    Siiskonen, T.
    Goma, C.
    Pinto, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (07):
  • [50] ELIMINATION OF AIR BUOYANCY CORRECTION IN ABSOLUTE MEASUREMENT OF SMALL MASSES
    PROKIC, D
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1974, 7 (01): : 22 - 23