Use of an Infrared Spectroscopic Method for Isotopic Analysis of Gaseous Uranium Hexafluoride

被引:1
|
作者
Fessler, K. Alicia Strange [1 ]
O'Rourke, Patrick E. [1 ]
DeRoller, Nicholas F. [1 ]
Simmons, Darrell [2 ]
Serkiz, Steven M. [3 ,4 ]
机构
[1] Savannah River Natl Lab, 301 Gateway Dr, Aiken, SC 29803 USA
[2] Oak Ridge Natl Lab, 1 Bethel Valley Rd, Oak Ridge, TN 37830 USA
[3] Bechtel Natl Corp, 3000 Post Oak Blvd, Houston, TX 77056 USA
[4] Clemson Univ, Calhoun Dr, Clemson, SC 29634 USA
来源
METAL-MATRIX COMPOSITES: ADVANCES IN ANALYSIS, MEASUREMENT, AND OBSERVATIONS | 2021年
关键词
Uranium hexafluoride; Quantum cascade laser; Infrared spectroscopy; Isotope analysis; Uranium isotopes; High performance infrared; SPECTRUM;
D O I
10.1007/978-3-030-65249-4_12
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A high performance infrared (HPIR) system was developed and demonstrated for the infrared absorption analysis of uranium-235 and uranium-238 isotopes in uranium hexafluoride gas samples. The method takes advantage of the recent commercial availability of quantum cascade lasers (QCL) and has required the use of advanced statistical data analysis, as well as materials selection due to the chemical reactivity of uranium hexafluoride gas. Sweeping the QCL over the spectral range and sampling via a high-rate analog-to-digital converter provided 0.0005 cm(-1) spectral resolution, allowing for high-precision measurements of the isotopic peak shift. A data analysis method was developed using principal component analysis to predict the isotope weight percent content of uranium-235. The HPIR precision, accuracy, and error were evaluated for a wide range of isotopic ratio samples (0.287-93.7 weight percent uranium-235), and the results were compared to the International Target Values (ITVs) set forth by the International Atomic Energy Agency (IAEA) for nondestructive and destructive analytical techniques used in safeguards verification under the Treaty on the Non-proliferation of Nuclear Weapons. The method meets or surpasses the IAEA ITVs for nondestructive analysis of samples with isotopic content of depleted to highly enriched. The results also demonstrated the capability of the HPIR system to correctly predict the uranium-235 weight percent content of a mislabeled sample whose isotopic distribution was validated by mass spectroscopic measurements. The HPIR measurement is nondestructive and, thus, allows for confirmatory analyses of the exact sample at a designated IAEA laboratory if higher resolution or a certified analysis is needed.
引用
收藏
页码:183 / 199
页数:17
相关论文
共 50 条
  • [31] Restoration of the isotopic composition of reprocessed uranium hexafluoride using cascade with additional product
    Palkin, Valerii
    Maslyukov, Eugenii
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2020, 52 (12) : 2867 - 2873
  • [32] PRELIMINARY HAZARD ANALYSIS OF URANIUM HEXAFLUORIDE ACCIDENT
    Chen Lei
    Zhang JianGang
    Li GuoQiang
    Sun ShuTang
    Meng DongYuan
    Wang Ning
    PROCEEDINGS OF THE 26TH INTERNATIONAL CONFERENCE ON NUCLEAR ENGINEERING, 2018, VOL 4, 2018,
  • [33] CATALYSIS OF REACTION BETWEEN URANIUM TETRAFLUORIDE AND OXYGEN - NOVEL APPROACH TO PREPARATION OF URANIUM HEXAFLUORIDE BY A METHOD NOT REQUIRING USE OF FLUORINE
    EKSTROM, A
    BATLEY, GE
    INORGANIC & NUCLEAR CHEMISTRY LETTERS, 1973, 9 (11): : 1157 - 1161
  • [34] FRENCH PROGRAM FOR URANIUM ISOTOPIC ENRICHMENT BY GASEOUS DIFFUSION - INTRODUCTION
    PECQUEUR, M
    BULLETIN D INFORMATIONS SCIENTIFIQUES ET TECHNIQUES DU COMMISSARIAT A L ENERGIE ATOMIQUE, 1975, (206): : 3 - 3
  • [35] ISOTOPIC EXCHANGE BETWEEN GASEOUS OXYGEN + SOME URANIUM COMPOUNDS
    SPITSYN, I
    LAVUT, EG
    DOKLADY AKADEMII NAUK SSSR, 1964, 159 (03): : 626 - &
  • [36] CORRELATING OF OPERATING AND STATIC PRESSURES - URANIUM HEXAFLUORIDE GASEOUS DIFFUSION CASCADE INVENTORY CONTROL
    RUTLEDGE, GP
    WERNECKE, AH
    INDUSTRIAL AND ENGINEERING CHEMISTRY, 1959, 51 (02): : 203 - 204
  • [37] INVESTIGATING MOLECULAR-INTERACTIONS IN GASEOUS URANIUM HEXAFLUORIDE BY NUCLEAR MAGNETIC-RESONANCE
    DEMCO, DE
    BOGDAN, M
    FITORI, P
    BALIBANU, F
    DARABONT, A
    REVUE ROUMAINE DE PHYSIQUE, 1988, 33 (4-6): : 509 - 514
  • [38] DESIGN AND ANALYSIS OF A URANIUM HEXAFLUORIDE ACTINIDE TRANSMUTATION REACTOR
    WAN, PT
    CLEMENT, JD
    ANNALS OF NUCLEAR ENERGY, 1983, 10 (11) : 607 - 616
  • [39] ANALYSIS OF ABSORPTION SPECTRUM OF CESIUM URANIUM(V) HEXAFLUORIDE
    REISFELD, MJ
    CROSBY, GA
    INORGANIC CHEMISTRY, 1965, 4 (01) : 65 - &
  • [40] Thermal impacts on nondestructive analysis measurements of uranium hexafluoride
    Allison T. Greaney
    Kathryn Peruski
    Jason M. Richards
    Susan K. Smith
    Ramkumar Venkataraman
    Glenn A. Fugate