Use of an Infrared Spectroscopic Method for Isotopic Analysis of Gaseous Uranium Hexafluoride

被引:1
|
作者
Fessler, K. Alicia Strange [1 ]
O'Rourke, Patrick E. [1 ]
DeRoller, Nicholas F. [1 ]
Simmons, Darrell [2 ]
Serkiz, Steven M. [3 ,4 ]
机构
[1] Savannah River Natl Lab, 301 Gateway Dr, Aiken, SC 29803 USA
[2] Oak Ridge Natl Lab, 1 Bethel Valley Rd, Oak Ridge, TN 37830 USA
[3] Bechtel Natl Corp, 3000 Post Oak Blvd, Houston, TX 77056 USA
[4] Clemson Univ, Calhoun Dr, Clemson, SC 29634 USA
来源
METAL-MATRIX COMPOSITES: ADVANCES IN ANALYSIS, MEASUREMENT, AND OBSERVATIONS | 2021年
关键词
Uranium hexafluoride; Quantum cascade laser; Infrared spectroscopy; Isotope analysis; Uranium isotopes; High performance infrared; SPECTRUM;
D O I
10.1007/978-3-030-65249-4_12
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A high performance infrared (HPIR) system was developed and demonstrated for the infrared absorption analysis of uranium-235 and uranium-238 isotopes in uranium hexafluoride gas samples. The method takes advantage of the recent commercial availability of quantum cascade lasers (QCL) and has required the use of advanced statistical data analysis, as well as materials selection due to the chemical reactivity of uranium hexafluoride gas. Sweeping the QCL over the spectral range and sampling via a high-rate analog-to-digital converter provided 0.0005 cm(-1) spectral resolution, allowing for high-precision measurements of the isotopic peak shift. A data analysis method was developed using principal component analysis to predict the isotope weight percent content of uranium-235. The HPIR precision, accuracy, and error were evaluated for a wide range of isotopic ratio samples (0.287-93.7 weight percent uranium-235), and the results were compared to the International Target Values (ITVs) set forth by the International Atomic Energy Agency (IAEA) for nondestructive and destructive analytical techniques used in safeguards verification under the Treaty on the Non-proliferation of Nuclear Weapons. The method meets or surpasses the IAEA ITVs for nondestructive analysis of samples with isotopic content of depleted to highly enriched. The results also demonstrated the capability of the HPIR system to correctly predict the uranium-235 weight percent content of a mislabeled sample whose isotopic distribution was validated by mass spectroscopic measurements. The HPIR measurement is nondestructive and, thus, allows for confirmatory analyses of the exact sample at a designated IAEA laboratory if higher resolution or a certified analysis is needed.
引用
收藏
页码:183 / 199
页数:17
相关论文
共 50 条
  • [1] Infrared spectroscopic method for uranium isotopic analysis
    Fessler, K. Alicia Strange
    O'Rourke, Patrick E.
    DeRoller, Nicholas F.
    Simmons, Darrell
    Serkiz, Steven M.
    CHEMICAL, BIOLOGICAL, RADIOLOGICAL, NUCLEAR, AND EXPLOSIVES (CBRNE) SENSING XXI, 2020, 11416
  • [2] URANIUM HEXAFLUORIDE ISOTOPIC ANALYSIS
    BONIFORTI, R
    BRACCIOCURTI, G
    ANNALI DI CHIMICA, 1974, 64 (3-4) : 247 - 255
  • [3] Analytical characterization of laser induced plasmas towards uranium isotopic analysis in gaseous uranium hexafluoride
    Chan, George C-Y
    Martin, Leigh R.
    Trowbridge, Lee D.
    Zhu, Zhenli
    Mao, Xianglei
    Russo, Richard E.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2021, 176 (176)
  • [4] THE VISCOSITY OF GASEOUS URANIUM HEXAFLUORIDE
    MYERSON, AL
    EICHER, JH
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1952, 74 (11) : 2758 - 2761
  • [5] HIGH-PRECISION ISOTOPIC ANALYSIS OF URANIUM HEXAFLUORIDE
    BISHOP, J
    DAVIDSON, DF
    EVANS, PBF
    ROBBINS, EJ
    MCKNIGHT, JA
    HAMER, AN
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1961, 38 (04): : 109 - &
  • [6] ISOTOPIC EXCHANGE BETWEEN URANIUM HEXAFLUORIDE AND PENTAFLUORIDE
    GRIGOREV, GY
    DOROFEEV, SB
    ZAMETALOV, VA
    KOLESNIKOV, ON
    TERENTEV, AA
    KHIMICHESKAYA FIZIKA, 1984, 3 (10): : 1438 - 1441
  • [7] STUDY OF IN-LINE PLUTONIUM ISOTOPIC ANALYSIS FOR GASEOUS PLUTONIUM HEXAFLUORIDE.
    Li, T.K.
    Nuclear materials management, 1987, 16 : 359 - 363
  • [8] SORPTION OF GASEOUS URANIUM HEXAFLUORIDE ON SOLID BARIUM FLUORIDE
    KORINEK, K
    PEKA, I
    COLLECTION OF CZECHOSLOVAK CHEMICAL COMMUNICATIONS, 1968, 33 (10) : 3113 - +
  • [9] Certification of uranium hexafluoride reference materials for isotopic composition
    S. Mialle
    S. Richter
    C. Hennessy
    J. Truyens
    U. Jacobsson
    Y. Aregbe
    Journal of Radioanalytical and Nuclear Chemistry, 2015, 305 : 255 - 266
  • [10] Certification of uranium hexafluoride reference materials for isotopic composition
    Mialle, S.
    Richter, S.
    Hennessy, C.
    Truyens, J.
    Jacobsson, U.
    Aregbe, Y.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2015, 305 (01) : 255 - 266