Evaluation of a new fiber-optic pressure recording system for cardiovascular measurements in mice

被引:21
|
作者
Woldbaek, PR [1 ]
Stromme, TA
Sande, JB
Christensen, G
Tonnessen, T
Ilebekk, A
机构
[1] Univ Oslo, Ulleval Hosp, Dept Cardiothorac Surg, Expt Med Res Inst, N-0407 Oslo, Norway
[2] Univ Oslo, Expt Med Res Inst, Oslo, Norway
来源
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY | 2003年 / 285卷 / 05期
关键词
catheter size; fiber optics; aortic pressure;
D O I
10.1152/ajpheart.01123.2002
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We have tested a new fiber-optic pressure recording system, Samba, with a thin fiber [outer diameter (OD) = 0.25 mm] and a pressure sensor (length and OD = 0.42 mm) attached to the end. The accuracy of the system tested in vitro was good, with a coefficient of variation of 2.54% at 100 mm.Hg. The drift was <0.45 mmHg/h, and the temperature sensitivity was -0.07 mmHg/1degreesC between 22 and 37degreesC. The frequency response characteristics were similar to a 1.4-Fr Millar catheter (0 - 200 Hz). Introduction of the Samba sensor from the right carotid artery into the left ventricle in six mice caused no drop in mean aortic pressure, whereas introduction of a 1.4-Fr Millar catheter (OD = 0.47 mm; n = 6) caused a pressure drop from 91.6 +/- 9.2 to 65.1 +/- 6.2 mmHg; P < 0.05. Thus the Samba sensor system may represent a new alternative to assess hemodynamic variables in the murine cardiovascular system.
引用
收藏
页码:H2233 / H2239
页数:7
相关论文
共 50 条
  • [21] Miniature fiber-optic pressure sensor
    Zhu, YZ
    Wang, AB
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2005, 17 (02) : 447 - 449
  • [22] New fiber-optic switch
    Zhang, H.Y.
    Feng, X.Y.
    Niu, D.F.
    Bandaoti Guangdian/Semiconductor Optoelectronics, 2001, 22 (01):
  • [23] Fiber-optic polarization pressure sensor
    Rzhavin, YI
    MEASUREMENT TECHNIQUES, 2002, 45 (07) : 738 - 741
  • [24] Fiber-Optic Polarization Pressure Sensor
    Yu. I. Rzhavin
    Measurement Techniques, 2002, 45 : 738 - 741
  • [25] Recording of Hydroacoustic Signals Using a Fiber-Optic Accelerometer
    O. T. Kamenev
    Yu. S. Petrov
    A. A. Podlesnykh
    V. A. Kolchinskii
    I. N. Zavestovskaya
    Yu. N. Kulchin
    R. V. Romashko
    Bulletin of the Lebedev Physics Institute, 2020, 47 : 146 - 148
  • [26] Recording of Hydroacoustic Signals Using a Fiber-Optic Accelerometer
    Kamenev, O. T.
    Petrov, Yu S.
    Podlesnykh, A. A.
    Kolchinskii, V. A.
    Zavestovskaya, I. N.
    Kulchin, Yu N.
    Romashko, R., V
    BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2020, 47 (05) : 146 - 148
  • [27] A new fiber optic modular sensing system for pressure, strain and temperature measurements
    Lesiak, P.
    Wolinski, T. R.
    Tul, J.
    Domanski, A. W.
    Wisniewski, R.
    OPTICAL MICRO- AND NANOMETROLOGY IN MICROSYSTEMS TECHNOLOGY, 2006, 6188
  • [28] Commissioning and Evaluation of a Fiber-Optic Sensor System for Bridge Monitoring
    Scott, Richard H.
    Banerji, Pradipta
    Chikermane, Sanjay
    Srinivasan, Sudarshan
    Basheer, P. A. Muhammed
    Surre, Frederic
    Sun, Tong
    Grattan, Kenneth T. V.
    IEEE SENSORS JOURNAL, 2013, 13 (07) : 2555 - 2562
  • [29] Dispersion measurements of anisotropic materials and a new fiber-optic sensor configuration
    Chlebus, R.
    Hlubina, P.
    Ciprian, D.
    15TH CZECH-POLISH-SLOVAK CONFERENCE ON WAVE AND QUANTUM ASPECTS OF CONTEMPORARY OPTICS, 2007, 6609
  • [30] Beyond the barostat: Minimally invasive assessment of gastric volume and pressure by combined magnetic resonance imaging and fiber-optic pressure recording system
    Fox, Mark R.
    Steingoetter, Andreas
    Kwiatek, Monika A.
    Treier, Reto
    Boesiger, Peter
    Fried, Michael
    Schwizer, Werner
    GASTROENTEROLOGY, 2008, 134 (04) : A578 - A578