Evaluation of a new fiber-optic pressure recording system for cardiovascular measurements in mice

被引:21
|
作者
Woldbaek, PR [1 ]
Stromme, TA
Sande, JB
Christensen, G
Tonnessen, T
Ilebekk, A
机构
[1] Univ Oslo, Ulleval Hosp, Dept Cardiothorac Surg, Expt Med Res Inst, N-0407 Oslo, Norway
[2] Univ Oslo, Expt Med Res Inst, Oslo, Norway
来源
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY | 2003年 / 285卷 / 05期
关键词
catheter size; fiber optics; aortic pressure;
D O I
10.1152/ajpheart.01123.2002
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We have tested a new fiber-optic pressure recording system, Samba, with a thin fiber [outer diameter (OD) = 0.25 mm] and a pressure sensor (length and OD = 0.42 mm) attached to the end. The accuracy of the system tested in vitro was good, with a coefficient of variation of 2.54% at 100 mm.Hg. The drift was <0.45 mmHg/h, and the temperature sensitivity was -0.07 mmHg/1degreesC between 22 and 37degreesC. The frequency response characteristics were similar to a 1.4-Fr Millar catheter (0 - 200 Hz). Introduction of the Samba sensor from the right carotid artery into the left ventricle in six mice caused no drop in mean aortic pressure, whereas introduction of a 1.4-Fr Millar catheter (OD = 0.47 mm; n = 6) caused a pressure drop from 91.6 +/- 9.2 to 65.1 +/- 6.2 mmHg; P < 0.05. Thus the Samba sensor system may represent a new alternative to assess hemodynamic variables in the murine cardiovascular system.
引用
收藏
页码:H2233 / H2239
页数:7
相关论文
共 50 条
  • [1] Evaluation of a fiber-optic technique for recording intramuscular pressure in the human leg
    Nilsson, Andreas
    Zhang, Qiuxia
    Styf, Jorma
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2016, 30 (05) : 699 - 705
  • [2] Evaluation of a fiber-optic technique for recording intramuscular pressure in the human leg
    Andreas Nilsson
    Qiuxia Zhang
    Jorma Styf
    Journal of Clinical Monitoring and Computing, 2016, 30 : 699 - 705
  • [3] Hydroacoustic Pressure Gradient Recording by a System of Two Fiber-Optic Accelerometers
    O. T. Kamenev
    Yu. S. Petrov
    A. A. Podlesnykh
    V. A. Kolchinskiy
    I. N. Zavestovskaya
    Yu. N. Kulchin
    R. V. Romashko
    Bulletin of the Lebedev Physics Institute, 2021, 48 : 135 - 138
  • [4] Hydroacoustic Pressure Gradient Recording by a System of Two Fiber-Optic Accelerometers
    Kamenev, O. T.
    Petrov, Yu S.
    Podlesnykh, A. A.
    Kolchinskiy, V. A.
    Zavestovskaya, I. N.
    Kulchin, Yu N.
    Romashko, R., V
    BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2021, 48 (05) : 135 - 138
  • [5] Feasibility study of a fiber-optic system for invasive blood pressure measurements
    Reesink, KD
    van der Nagel, T
    Bovelander, J
    Jansen, JRC
    van der Veen, FH
    Schreuder, JJ
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2002, 57 (02) : 272 - 276
  • [6] Fiber-optic system for the remote recording of explosive gases
    Mirumyants, S.O.
    Maksimyuk, V.S.
    Journal of Optical Technology (A Translation of Opticheskii Zhurnal), 1997, 64 (02): : 157 - 160
  • [7] Fiber-optic system for the remote recording of explosive gases
    Mirumyants, SO
    Maksimyuk, VS
    JOURNAL OF OPTICAL TECHNOLOGY, 1997, 64 (02) : 157 - 160
  • [8] A fiber-optic system for recording skin conductance in the MRI scanner
    Lagopoulos, J
    Malhi, GS
    Shnier, RC
    BEHAVIOR RESEARCH METHODS, 2005, 37 (04) : 657 - 664
  • [9] A fiber-optic system for recording skin conductance in the MRI scanner
    Jim Lagopoulos
    Gin S. Malhi
    Ronald C. Shnier
    Behavior Research Methods, 2005, 37 : 657 - 664
  • [10] Highly sensitive fiber-optic sensor for dynamic pressure measurements
    Bock, WJ
    Nawrocka, MS
    Urbanczyk, W
    Rostkowski, J
    IMTC/2000: PROCEEDINGS OF THE 17TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE: SMART CONNECTIVITY: INTEGRATING MEASUREMENT AND CONTROL, 2000, : 591 - 594