Autonomous calibration method of the zero-difference without using a standard gauge for a straightness-measuring machine

被引:7
|
作者
Fujimoto, Ikumatsu [1 ]
Nishimura, Kunitoshi [2 ]
Pyoun, Young-Sik [3 ]
机构
[1] Okuwa Tech Res Ctr, Saitama, Japan
[2] Tokyo Metropolitan Univ, Dept Urban Liberal Arts Sci & Engn, Tokyo 158, Japan
[3] Sunmoon Univ, Dept Mech Engn, Chungnam, South Korea
关键词
Autonomous calibration method; Straightness measuring machine; Three point method; Cone disc; Zero difference; Motion error; Error analysis; PROFILE MEASUREMENT; SYSTEM; PROBE;
D O I
10.1016/j.precisioneng.2010.09.007
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
An autonomous method to calibrate the zero difference for the three-point method of measuring surface straightness is proposed A simple cone-disc which can be made by a CNC turning machine is used for the calibration The calibration method is comprised of two steps In the first step the cone-disc rotates around and parallel to three displacement sensors built-in to a holder In the second step the geometrical parameters between the sensors and the cone-disc are simultaneously acquired during each predetermined movement of the sensors and the zero-differences are computed by the proposed algorithm within a few revolutions The conceptual design and the algorithm to realize the proposed method are explained with theoretical analyses and numerical simulations The results of this study are as follows (1) The repeatable accuracy of the zero-difference determination using the proposed method is at the same level of quality as that of the reverse method using the three point method (2) When the zero difference is calibrated using a cone-disc gauge of 200 mm radius and 10 mu m peak height which can be manufactured by a CNC turning machine the shape en or of the objects surface (which is analyzed for its systematic error component and its repeatability error component) becomes less than 05 and 2 mu m at measured lengths of 100 and 200 trim respectively (C) 2010 Elsevier Inc All rights reserved
引用
收藏
页码:153 / 163
页数:11
相关论文
共 15 条
  • [11] Compensation of Sensor Gain Difference in Measuring Straightness Errors Using a Mixed Sequential Two-Probe Method
    Ji Hun Jeong
    Jeong Seok Oh
    Sung-Chong Chung
    International Journal of Precision Engineering and Manufacturing, 2019, 20 : 427 - 435
  • [12] Method of improving the accuracy of the three-coordinate measuring machine (2nd report, Calibration method using the three-dimensional standard)
    Oya, Makoto
    Hokari, Hisashi
    Tamura, Hisashi
    Nippon Kikai Gakkai Ronbunshu, C Hen/Transactions of the Japan Society of Mechanical Engineers, Part C, 1988, 54 (503): : 1445 - 1451
  • [13] A segmental calibration method for a miniature serial-link coordinate measuring machine using a compound calibration artefact
    Zhou, Awei
    Guo, Junjie
    Shao, Wei
    Li, Beizhan
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2013, 24 (06)
  • [14] Investigation of pressure measuring method using vacuum gauge controlled by pulse magnetion without detachment
    Liao, Congrong
    Zou, Jiyan
    Gaoya Dianqi/High Voltage Apparatus, 1997, 33 (01): : 20 - 23
  • [15] A Kinematic Calibration Method of the Articulated Arm Coordinate Measuring Machine Using Niching Chaos Optimization Algorithm
    Rim, Cholmin
    Rim, Chang-Hyon
    Chen, Gang
    Sin, Yongchol
    Kim, Kukchol
    JOURNAL OF TESTING AND EVALUATION, 2020, 48 (04) : 2547 - 2563