Xenograft Tumor Volume Measurement in Nude Mice: Estimation of 3D Ultrasound Volume Measurements Based on Manual Caliper Measurements

被引:11
|
作者
Baris, Mustafa Mahmut [1 ]
Serinan, Efe [2 ]
Calisir, Meryem [3 ]
Simsek, Kursat [1 ]
Aktas, Safiye [2 ]
Yilmaz, Osman [3 ]
Ozdemir, Sevgi Kilic [4 ]
Secil, Mustafa [1 ]
机构
[1] Dokuz Eylul Univ, Sch Med, Radiol, Izmir, Turkey
[2] Dokuz Eylul Univ, Inst Oncol, Basic Oncol, Izmir, Turkey
[3] Dokuz Eylul Univ, Lab Anim Sci, Izmir, Turkey
[4] Izmir Inst Technol, Chem Engn, Izmir, Turkey
来源
关键词
Tumor volume; nude mice; ultrasonography; 3-D imaging; laboratory animal science; TESTICULAR VOLUME; BONE-MARROW; FORMULAS;
D O I
10.30621/jbachs.2020.902
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Objectives: Volume measurement of subcutaneous xenograft tumors in nude mice models is an important metric to assess tumor growth or response to therapy. Manual calipers are widely used for this purpose. But the measurements with manual calipers may be inaccurate. Contrarily, three-dimensional (3D) ultrasonographic measurements give reliable and accurate tumor volume calculation. We aim to; evaluate the accuracy of common four formulas given in the literature to estimate xenograft tumor volumes based on manual caliper measurements and offer a new coefficient for a better estimation of the tumor volumes. Patients and Methods: Detailed manual diameter measurements of xenograft tumors were in 14 nude mice performed using Vernier caliper. Tumor volumes were calculated using the suggested formulas in the literature based on manual measurements. 3D ultrasound volume measurements were performed on same xenograft tumors using high resolution Vevo 2100 imaging system. To propose a new coefficient; means of ratio between manual and ultrasound volume measurement values were used. Also, data set was divided into two subgroups as tumor volume under 800 mm(3) and over 800 mm(3). New coefficients for each subgroup were defined. Results: Only with prolate ellipsoid formula there was no statistically significant difference between volume measurements with two methods (p=0,24). Our proposed formula (0,45 L*W*H) could estimate tumor volumes as good as prolate ellipsoid formula. Coefficient 0,35 and 0,81 in the same formula were found efficient in the selected subgroups. Conclusion: Using one common coefficient/formula for tumor volume estimation in any tumor size can be inaccurate. Appropriate coefficient should be chosen according to the dataset worked with.
引用
收藏
页码:90 / 95
页数:6
相关论文
共 50 条
  • [21] Does 3D ultrasound measurement improve the assessment of ovarian volume?
    Brett, S.
    Kelsey, T. W.
    Wallace, W. H. B.
    Rajkhowa, M.
    HUMAN REPRODUCTION, 2007, 22 : I56 - I57
  • [22] An Innovative Way of 3D Placental Volume Measurements in Early Pregnancy
    Reus, Averil D.
    El-Harbachi, Hakima
    Rousian, Melek
    Exalto, Niek
    Steegers-Theunissen, Regine P. M.
    Steegers, Eric P.
    REPRODUCTIVE SCIENCES, 2011, 18 (03) : 203A - 203A
  • [23] Three-dimensional (3D) area and volume measurements for rhinoplasty
    Celikoyar, M. Mazhar
    Topsakal, Oguzhan
    Sawyer, Philip
    JOURNAL OF PLASTIC RECONSTRUCTIVE AND AESTHETIC SURGERY, 2023, 83 : 189 - 197
  • [24] Comparing hoof dimensional measurements in cows based on 3D image creation and manual measurement
    Murakami, Takashi
    Ohtake, Ayaka
    Ishikawa, Takaaki
    Sato, Ayano
    VETERINARY JOURNAL, 2024, 306
  • [25] 3D Breast Volume Estimation
    Gouveia, Pedro F.
    Oliveira, Helder P.
    Monteiro, Joao P.
    Teixeira, Joao F.
    Silva, Nuno L.
    Pinto, David
    Mavioso, Carlos
    Anacleto, Joao
    Martinho, Marta
    Duarte, Ines
    Cardoso, Jaime S.
    Cardoso, Fatima
    Cardoso, Maria Joao
    EUROPEAN SURGICAL RESEARCH, 2022, 63 (01) : 3 - 8
  • [26] Estimation of 3D Sphericity by Volume Measurement - Application to Coarse Aggregates
    Bhattacharya, Sumana
    Subedi, Sudip
    Lee, Seung Jae
    Pradhananga, Nipesh
    TRANSPORTATION GEOTECHNICS, 2020, 23
  • [27] Evaluation of superficial xenograft volume estimation by ultrasound and caliper against MRI in a longitudinal pre-clinical radiotherapeutic setting
    Roth, Daniel
    Safi, Marcella
    Timmermand, Oskar Vilhelmsson
    Sereti, Evangelia
    Molendowska, Malwina
    Gottschalk, Michael
    Bjartell, Anders
    Ceberg, Crister
    Szczepankiewicz, Filip
    Strand, Joanna
    PLOS ONE, 2024, 19 (07):
  • [28] 3D volume ultrasound in reproductive medicine
    Weichert, J.
    Schroeer, A.
    GYNAKOLOGE, 2012, 45 (10): : 761 - 769
  • [29] Renal volume assessment with 3D ultrasound
    Brancaforte, A.
    Serantoni, S.
    Barbosa, F. Silva
    Di Leo, G.
    Sardanelli, F.
    Cornalba, G. P.
    RADIOLOGIA MEDICA, 2011, 116 (07): : 1095 - 1104
  • [30] Evaluation of 3D Laser Scanning for Estimation of Heating-Induced Volume Shrinkage and Prediction of Cooking Loss of Pork Cuboids Compared to Manual Measurements
    Vaskoska, Rozita
    Ha Minh
    Tran Ha Thi Thu
    Khoshelham, Kourosh
    White, Jason D.
    Warner, Robyn D.
    FOOD AND BIOPROCESS TECHNOLOGY, 2020, 13 (06) : 938 - 947