A new laser ablation system for quantitative analysis of solid samples with ICP-MS

被引:4
|
作者
Ishida, Tomoharu [1 ]
Akiyoshi, Takanori [2 ]
Sakashita, Akiko [3 ]
Kinoshiro, Satoshi [4 ]
Fujimoto, Kyoko [1 ]
Chino, Atsushi [4 ]
机构
[1] JFE Steel Corp, Steel Res Lab, Anal & Characterizat Res Dept, Chuo Ku, Chiba 2600835, Japan
[2] Japan Iron & Steel Federat, Standardizat Ctr, Chuo Ku, Tokyo 1030025, Japan
[3] JFE Techno Res Corp, Anal & Characterizat Div, Kawasaki, Kanagawa 2100855, Japan
[4] JFE Steel Corp, Steel Res Lab, Anal & Characterizat Res Dept, Kawasaki, Kanagawa 2100855, Japan
关键词
D O I
10.2116/analsci.24.563
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The laser ablation (LA) method is an effective technique for quantitative analysis. In the present work, a new LA system was developed for the high-sensitivity analysis of metal materials using inductively coupled plasma mass spectrometry (ICP-MS). This system consists of a high-frequency Q-switched laser and 2 scanning mirrors for scanning the ablation spot in an adequately large area of the specimen without vacant spaces. The influence of elemental fractionation (non-stoichiometric generation of vapor species) can be eliminated by repetitive irradiation of this pattern on the same area. Particles generated with an average laser power of 0.6 W with the developed LA system gave intensity and stability substantially similar to that of a 500 mu g/ml solution steel sample in solution ICP-MS. The analytical performance of the developed LA-ICP-MS was compared with that of a solution ICP-MS using NIST steel SRMs. The performance of the newly-developed system is comparable to that of conventional solution ICP-MS in both accuracy and precision. The correlation coefficients between the contents and the intensity ratios to Fe were over 0.99 for most elements. The relative standard deviation (RSD) obtained by LA-ICP-MS revealed that this system can analyze iron samples with good precision. The results of ultra trace level analysis of high-purity iron showed that developed LA-ICP-MS is capable of analyzing ppm concentration levels with a 20 - 30 ppb level standard deviation. The detection limit was on the order of 10 ppb for most elements.
引用
收藏
页码:563 / 569
页数:7
相关论文
共 50 条
  • [31] Major element analysis of natural silicates by laser ablation ICP-MS
    Humayun, Munir
    Davis, Fred A.
    Hirschmann, Marc M.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2010, 25 (07) : 998 - 1005
  • [32] Evaluation of infrared femtosecond laser ablation for the analysis of geomaterials by ICP-MS
    Freydier, R.
    Candaudap, F.
    Poitrasson, F.
    Arbouet, A.
    Chatel, B.
    Dupre, B.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (05) : 702 - 710
  • [33] APPLICATION OF LASER-ABLATION ICP-MS TO ELEMENTAL ANALYSIS OF GLASSES
    IMBERT, JL
    TELOUK, P
    MIKROCHIMICA ACTA, 1993, 110 (4-6) : 151 - 160
  • [34] ICP-MS laser ablation oak wood metals wood standard semi-quantitative analysis
    Alteyrac, J
    Augagneur, S
    Medina, B
    Vivas, N
    Glories, Y
    ANALUSIS, 1995, 23 (10) : 523 - 526
  • [35] A CRITICAL-LOOK AT QUANTITATIVE LASER-ABLATION ICP-MS ANALYSIS OF NATURAL AND SYNTHETIC GLASSES
    STIX, J
    GAUTHIER, G
    LUDDEN, JN
    CANADIAN MINERALOGIST, 1995, 33 : 435 - 444
  • [36] Quantitative analysis of total and insoluble elements and inclusion composition in metal by laser ablation ICP-MS method
    Karasev, AV
    Inoue, R
    Suito, H
    ISIJ INTERNATIONAL, 2001, 41 (07) : 757 - 765
  • [37] LASER ABLATION ICP-MS: APPLICATION IN BIOMEDICAL RESEARCH
    Sussulini, Alessandra
    Becker, J. Susanne
    Becker, J. Sabine
    MASS SPECTROMETRY REVIEWS, 2017, 36 (01) : 47 - 57
  • [38] Identification of pharmaceutical glasses by laser ablation ICP-MS
    Schmidt, T
    Surmann, JP
    Stephanowitz, H
    Hoffmann, E
    PHARMAZIE, 2001, 56 (11): : 852 - 856
  • [39] Laser ablation ICP-MS developments and trends for 2003
    Sylvester, PJ
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2005, 29 (01) : 41 - 52
  • [40] Influence of wavelength on fractionation in laser ablation ICP-MS
    Russo, RE
    Mao, XL
    Borisov, OV
    Liu, HC
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (09) : 1115 - 1120