A novel hypothermic machine perfusion system using a LifePort Kidney Transporter for the preservation of rat liver

被引:4
|
作者
Zeng, Cheng [1 ,2 ,3 ]
Hu, Xiaoyan [1 ,2 ,3 ]
Wang, Yanfeng [1 ,2 ,3 ]
Zeng, Xianpeng [1 ,2 ,3 ]
Xiong, Yan [1 ,2 ,3 ]
Li, Ling [1 ,2 ,3 ]
Ye, Qifa [1 ,2 ,3 ,4 ]
机构
[1] Wuhan Univ, Zhongnan Hosp, Transplant Ctr, 169 Donghu Rd, Wuhan 430071, Hubei, Peoples R China
[2] Wuhan Univ, Inst Hepatobiliary Dis, Wuhan 430071, Hubei, Peoples R China
[3] Wuhan Univ, Zhongnan Hosp, Hubei Key Lab Med Technol Transplantat, Wuhan 430071, Hubei, Peoples R China
[4] Cent S Univ, Xiangya Hosp 3, Natl Hlth Minist Transplantat Med Engn & Technol, Res Ctr, Changsha 410013, Hunan, Peoples R China
基金
中国国家自然科学基金;
关键词
hypothermic machine perfusion; preservation; liver transplantation; COLD PRESERVATION; FLUORESCENT DYES; CELL VIABILITY; TRANSPLANTATION; EXPERIENCE; DONORS; OXYGEN; DONATION; IMPACT; DEATH;
D O I
10.3892/etm.2017.5587
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The protective mechanisms for liver preservation associated with hypothermic machine perfusion (HMP) remain unclear. However, the lack of a common and portable HMP system for rat livers limits the study of HMP. The present study aimed to develop a novel, modified HMP system using a LifePort Kidney Transporter for preserving rat livers. A simple Y' shunt combined with a pressoreceptor for flow and pressure regulation was adapted to perfuse rat livers via the portal vein continuously using a LifePort Kidney Transporter under its prime mode' setting. An electronic scale was installed under the liver container to calculate the portal inflow according to the association with weight, density and volume of the perfusate. A total of 10 rat livers underwent 6 h of HMP using histidine-tryptophan-ketoglutarate solution enriched with acridine orange (AO) and propidium iodide (PI). The perfusion status of HMP was assessed by comparison of AO+PI-positive cell count in core region (CR) and peripheral region (PR) of rat liver under fluorescence microscopy. The dynamics (inflow, pressure and intrahepatic resistance of perfusion) were assessed to identify whether this system met the demands for HMP of rat livers. Biochemical [alanine transaminase (ALT), lactate dehydrogenase (LDH) and endothelin levels] and histological parameters (sinusoidal dilatation, endothelial cell detachment and vacuolization) were measured to determine cellular damage associated with HMP. No significant difference was observed between the CR and PR according to the comparison of the AO+PI-positive cell count, which indicated that complete perfusion was achieved. Intrahepatic resistance significantly decreased during the initial 3 h of HMP (P<0.01), but remained stable during the final 3 h. ALT and LDH levels significantly increased over the 6 h HMP duration: ALT (0 h, 42.67 +/- 5.81 U/l; 3 h, 90.67 +/- 6.74 U/l; 6 h, 164.33 +/- 7.31 U/l; P<0.01) and LDH (0 h, 492.90 +/- 90.20 U/l; 3 h, 973.53 +/- 97.4; 6 h, 1,843.40 +/- 85.78 U/l; P<0.01) However, the levels of endothelin and oxygen consumption were constant throughout HMP. Furthermore, histological analysis indicated sinusoidal dilation was significantly increased in the post-HMP group compared with the pre-HMP group (P<0.01); however, no other significant differences were observed. Combined with the results of ATP test (640.64 +/- 29.46 nmol/l) and bile production (4.88 +/- 0.69 mu l/h/g of liver) at the end of HMP, the present results demonstrated minimal cellular injury associated with HMP while retaining the dependability and portability of the LifePort Kidney Transporter, which suggests the modified HMP system met the demands required and may be suitable for rat liver preservation.
引用
收藏
页码:1410 / 1416
页数:7
相关论文
共 50 条
  • [1] A NOVEL HYPOTHERMIC HYPERBARIC MACHINE PERFUSION REDUCES RAT LIVER PRESERVATION INJURY
    Giannone, Ferdinando
    Caraceni, Paolo
    Casalena, Gabriella
    Domenicali, Marco
    Grattagliano, Ignazio
    Baracca, Alessandro
    Solaini, Giancarlo
    Trere, Davide
    Bernardi, Mauro
    HEPATOLOGY, 2008, 48 (04) : 638A - 638A
  • [2] Hypothermic machine perfusion for liver graft preservation
    Amin, Arpit
    Panayotova, Guergana
    Guarrera, James V.
    CURRENT OPINION IN ORGAN TRANSPLANTATION, 2022, 27 (02) : 98 - 105
  • [3] Improved kidney graft function after preservation using a novel hypothermic machine perfusion device
    Maathuis, Mark-Hugo J.
    Manekeller, Steffen
    van der Plaats, Arjan
    Leuvenink, Henri G. D.
    't Hart, Nils A.
    Lier, A. Bastiaan
    Rakhorst, Gerhard
    Ploeg, Rutger J.
    Minor, Thomas
    ANNALS OF SURGERY, 2007, 246 (06) : 982 - 991
  • [4] Hypothermic Machine Perfusion Preservation of the DCD Kidney: Machine Effects
    Lindell, Susanne L.
    Muir, Heather
    Brassil, John
    Mangino, Martin J.
    JOURNAL OF TRANSPLANTATION, 2013, 2013
  • [5] Improved rat liver preservation by hypothermic continuous machine perfusion using Polysol, a new, enriched preservation solution
    Bessems, M
    Doorschodt, BM
    van Vliet, AK
    van Gulik, TM
    LIVER TRANSPLANTATION, 2005, 11 (05) : 539 - 546
  • [6] Porcine kidney transplantation after machine preservation using the Groningen hypothermic organ perfusion system
    Maathuis, MHJ
    Manekeller, S
    van der Plaats, A
    t'Hart, NA
    Leuvenink, HGD
    Rakhorst, G
    Ploeg, RJ
    Minor, T
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2006, 29 (05): : 550 - 550
  • [7] Improved perfusion in hypothermic machine preservation of the liver.
    Hart, NAT
    van der Plaats, A
    Leuvenink, HGD
    van Goor, H
    Wiersema-Buist, J
    Verkerke, BJ
    Rakhorst, G
    Ploeg, RJ
    AMERICAN JOURNAL OF TRANSPLANTATION, 2004, 4 : 397 - 397
  • [8] A Novel Oxygenated Machine Perfusion System for Preservation of the Liver
    Dirkes, Marcel C.
    Post, Ivo C. J. H.
    Heger, Michal
    van Gulik, Thomas M.
    ARTIFICIAL ORGANS, 2013, 37 (08) : 719 - 724
  • [9] Short term hypothermic machine perfusion in a model of kidney preservation
    Hosgood, S. A.
    Mohamed, I. H.
    Nicholson, M. L.
    BRITISH JOURNAL OF SURGERY, 2010, 97 : 78 - 78
  • [10] The Effect of Hypothermic Machine Perfusion and Cold Storage for Kidney Preservation
    Chusri, Yanee
    Gesprasert, Goragoch
    Khienwad, Thananya
    Naiyanate, Phornphop
    2017 10TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON), 2017,