Influence of Emission Spectrum and Irradiance on Light Curing of Resin-Based Composites

被引:49
|
作者
Shimokawa, C. A. K. [1 ]
Sullivan, B. [2 ]
Turbino, M. L. [1 ]
Soares, C. J. [3 ]
Price, R. B. [2 ]
机构
[1] Univ Sao Paulo, Dept Restorat Dent, Sch Dent, Sao Paulo, Brazil
[2] Dalhousie Univ, Dept Dent Clin Sci, Fac Dent, Halifax, NS, Canada
[3] Univ Fed Uberlandia, Dept Operat Dent & Dent Mat, Sch Dent, Uberlandia, MG, Brazil
关键词
EXPOSURE RECIPROCITY LAW; MONOWAVE LED UNITS; EMITTING DIODE; MICRO-HARDNESS; CONVERSION; MICROHARDNESS; CURE; PHOTOPOLYMERIZATION; POLYWAVE; DISTANCE;
D O I
10.2341/16-349-L
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study examined the influence of different emission spectra ( single-peak and broad-spectrum) light-curing units (LCUs)delivering the same radiant exposures at irradiance values of 1200 or 3600 mW/cm(2) on the polymerization and light transmission of four resin-based composites (RBCs). Methods and Materials: Two prototype LCUs that used the same light tip, but were either a single-peak blue or a broad-spectrum LED, were used to deliver the same radiant exposures to the top surfaces of the RBCs using either standard (1200 mW/cm(2)) or high irradiance (3600 mW/cm(2)) settings. The emission spectrum and radiant power from the LCUs were measured with a laboratory-grade integrating sphere coupled to a spectrometer, and the light beam was assessed with a beam profiler camera. Four RBCs (Filtek Supreme Ultra A2, Tetric EvoCeram A2, Tetric EvoCeram T, and TPH Spectra High Viscosity A2) were photoactivated using four different light conditions: single-peak blue/standard irradiance, single-peak blue/high irradiance, broadspectrum/standard irradiance, and broadspectrum/high irradiance. The degree of conversion (N=5) and microhardness at the top and bottom of 2.3-mm-diameter by 2.5-mmthick specimens (N=5) were analyzed with analysis of variance and Tukey tests. The real-time light transmission through the RBCs was also measured. Results: For all light conditions, the 2.3-mmdiameter specimens received a homogeneous irradiance and spectral distribution. Although similar radiant exposures were delivered to the top surfaces of the RBCs, the amount of light energy emitted from the bottom surfaces was different among the four RBCs, and was also greater for the single-peak lights. Very little violet light ( wavelengths below 420 nm) reached the bottom of the 2.5-mm-thick specimens. The degree of conversion and microhardness results varied according to the RBC ( p < 0.05). The RBCs that included alternative photoinitiators had greater microhardness values at the top when cured with broad-spectrum lights, while at the bottom, where little violet light was observed, the results were equal or higher when they were photoactivated with single-peak blue lights. With the exception of the microhardness at the top of TPH, equivalent or higher microhardness and degree-of conversion values were achieved at the bottom surface when the standard ( 1200 mW/cm(2)) irradiance levels were used compared to when high irradiance levels were used. Conclusions: Considering the different behaviors of the tested RBCs, the emission spectrum and irradiance level influenced the polymerization of some RBCs. The RBCs that included alternative photoinitiators produced greater values at the top when cured with broadspectrum lights, while at the bottom, results were equal or higher for the RBCs photoactivated with single-peak blue lights.
引用
收藏
页码:537 / 547
页数:11
相关论文
共 50 条
  • [41] Influence of LED irradiance on flexural properties and Vickers hardness of resin-based composite materials
    Bhamra, G. S.
    Fleming, G. J. P.
    Darvell, B. W.
    DENTAL MATERIALS, 2010, 26 (02) : 148 - 155
  • [42] Subsurface degradation of resin-based composites
    Bagheri, Rafat
    Tyas, Martin J.
    Burrow, Michael F.
    DENTAL MATERIALS, 2007, 23 (08) : 944 - 951
  • [43] Bond strength of resin-based restorations polymerized with different light-curing sources
    D'Alpino, Paulo H. P.
    Wang, Linda
    Rueggeberg, Frederick A.
    Svizero, Nadia R.
    Pereira, Jose C.
    Pashley, David H.
    Carvalho, Ricardo M.
    JOURNAL OF ADHESIVE DENTISTRY, 2006, 8 (05): : 293 - 298
  • [44] Repair of dental resin-based composites
    Veronika Baur
    Nicoleta Ilie
    Clinical Oral Investigations, 2013, 17 : 601 - 608
  • [45] Statement of posterior resin-based composites
    JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 1998, 129 (11): : 1627 - 1628
  • [46] A cross-sectional study on dentists' learning preferences for learning about light-curing units and resin-based composites
    Al-Zain, Afnan O.
    Baghlaf, Khlood
    Abdulwassi, Omar
    Almukairin, Reem
    Alanazi, Maram
    Alshomrani, Elaf
    Alftaikhah, Sultan
    Price, Richard B.
    SAUDI DENTAL JOURNAL, 2024, 36 (12) : 1673 - 1680
  • [47] Repair of dental resin-based composites
    Baur, Veronika
    Ilie, Nicoleta
    CLINICAL ORAL INVESTIGATIONS, 2013, 17 (02) : 601 - 608
  • [48] Influence of composition on rate of polymerization contraction of light-curing resin composites
    Asmussen, E
    Peutzfeldt, A
    ACTA ODONTOLOGICA SCANDINAVICA, 2002, 60 (03) : 146 - 150
  • [49] Light-emitting diode curing: Influence on selected properties of resin composites
    Asmussen, E
    Peutzfeldt, A
    QUINTESSENCE INTERNATIONAL, 2003, 34 (01): : 71 - 75
  • [50] The influence of different light-curing modes on microleakage of posterior resin composites
    Marghalani, Hanadi Yousif
    JOURNAL OF ADHESION SCIENCE AND TECHNOLOGY, 2014, 28 (02) : 136 - 150