Evaluation of the Tibiofemoral Contact Characteristics of a Customized Surface-Guided Knee Implant

被引:3
|
作者
Pejhan, Shabnam [1 ,4 ]
Khosravipour, Ida [1 ]
Gascoyne, Trevor [2 ]
Bohm, Eric [2 ,3 ]
Brandt, Jan-Mels [1 ]
Luo, Yunhua [1 ]
Wyss, Urs [1 ,2 ]
机构
[1] Univ Manitoba, Dept Mech Engn, Winnipeg, MB R3T 5V6, Canada
[2] Orthopaed Innovat Ctr, Winnipeg, MB R2L 4L5, Canada
[3] Univ Manitoba, Dept Surg, Sect Orthoped Surg, Winnipeg, MB R3A 1R9, Canada
[4] Univ Manitoba, Dept Mech Engn, Engn & Informat Technol Complex, Winnipeg, MB, Canada
关键词
Total knee replacement; Pressure-sensitive film; Finite element; Surface-guided; HIGH-FLEXION; REPLACEMENT; STRESSES; JOINT; DAMAGE; AREAS; GAIT; FILM;
D O I
10.1007/s40846-018-0399-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A surface-guided total knee replacement (TKR) aims to achieve close to normal kinematics through specially shaped tibiofemoral articulating surfaces. The resulting geometry of the TKR components affects the tibiofemoral contact mechanics. Therefore, evaluation of the contact characteristics of the implant is critical, as the tibiofemoral contact stress influences the short-term and long-term performance and durability of the implants. In this study, the contact behavior at the tibiofemoral articulating interface for a customized surface-guided TKR was studied. The contact area and mean contact stress at specific flexion angles were measured mechanically by pressure sensitive films to validate the outcomes from finite element analysis (FEA) under similar load conditions. Two activities of daily living were studied; lunging and squatting. As expected, FEA predicted larger contact areas at various flexion angles due to the limited sensitivity range of the pressure sensitive films. During lunging and squatting, the contact area increased as the flexion angle reached 90 degrees, and decreased as the knee was flexed to higher angles. The mean contact pressure was less than 10MPa at 120 degrees of flexion under squatting load. The outcomes revealed that a high range of motion can be achieved, while the mean contact pressure remains below the material limits of the tibial polyethylene insert.
引用
收藏
页码:205 / 212
页数:8
相关论文
共 49 条
  • [1] Evaluation of the Tibiofemoral Contact Characteristics of a Customized Surface-Guided Knee Implant
    Shabnam Pejhan
    Ida Khosravipour
    Trevor Gascoyne
    Eric Bohm
    Jan-Mels Brandt
    Yunhua Luo
    Urs Wyss
    Journal of Medical and Biological Engineering, 2019, 39 : 205 - 212
  • [2] Customized surface-guided knee implant: Contact analysis and experimental test
    Khosravipour, Ida
    Pejhan, Shabnam
    Luo, Yunhua
    Wyss, Urs P.
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2018, 232 (01) : 90 - 100
  • [3] Design and virtual evaluation of a customized surface-guided knee implant
    Pejhan, Shabnam
    Bohm, Eric
    Brandt, Jan-Mels
    Wyss, Urs
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (10) : 949 - 961
  • [4] Kinematic behavior of a customized surface-guided knee implant during simulated knee-bending
    Pejhan, Shabnam
    Bohm, Eric
    Brandt, Jan-Mels
    Gascoyne, Trevor
    Wyss, Urs
    MEDICAL ENGINEERING & PHYSICS, 2017, 48 : 168 - 175
  • [5] Evaluation of dynamic accuracy and latency of a surface-guided radiotherapy system
    Yamauchi, Ryohei
    Tomita, Fumihiro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2025, 18 (01) : 136 - 146
  • [6] The influence of geometric design variables on the kinematic performance of a surface-guided total knee replacement
    Pejhan, Shabnam
    Bohm, Eric
    Brandt, Jan-Mels
    Wyss, Urs
    JOURNAL OF ORTHOPAEDIC SURGERY, 2017, 25 (03)
  • [7] Evaluation and implementation of open face masks with surface-guided radiotherapy (SGRT)
    Nachtigall, Rayk
    Seeto, Nicholas
    Exner, Sebastian
    Bislich, Oliver
    Behrens, Felix
    Peters, Bjoern
    Fehlauer, Fabian
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S1399 - S1400
  • [8] Evaluation and implementation of open face masks with surface-guided radiotherapy (SGRT)
    Nachtigall, Rayk
    Exner, Sebastian
    Seeto, Nicholas
    Fehlauer, Fabian
    RADIOTHERAPY AND ONCOLOGY, 2024, 192 : S237 - S238
  • [9] TIBIOFEMORAL CONTACT STRESS AND STRESS-DISTRIBUTION EVALUATION OF TOTAL KNEE ARTHROPLASTIES
    SZIVEK, JA
    CUTIGNOLA, L
    VOLZ, RG
    JOURNAL OF ARTHROPLASTY, 1995, 10 (04): : 480 - 491
  • [10] Evaluation of a method to map tibiofemoral contact points in the normal knee using MRI
    Scarvell, JM
    Smith, PN
    Refshauge, KM
    Galloway, HR
    Woods, KR
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (04) : 788 - 793