Development and validation of an assay for the measurement of gentamicin concentrations in dried blood spots using UHPLC-MS/MS

被引:6
|
作者
Anibaletto dos Santos, Ana Laura [1 ,2 ]
Cezimbra da Silva, Anne Caroline [1 ,2 ]
Feltraco Lizot, Lilian de Lima [1 ,2 ]
Schneider, Anelise [2 ]
Meireles, Yasmin Fazenda [2 ]
Hahn, Roberta Zilles [1 ]
Pagnussat, Lidiane Riva [4 ]
Nonnenmacher, Julia Livia [5 ]
Hahn, Siomara Regina [3 ]
Linden, Rafael [1 ,2 ]
机构
[1] Univ Feevale, Analyt Toxicol Lab, Novo Hamburgo, RS, Brazil
[2] Univ Feevale, Grad Program Toxicol & Analyt Toxicol, Novo Hamburgo, Brazil
[3] Univ Passo Fundo, Coll Pharm, Passo Fundo, RS, Brazil
[4] Hosp Sao Vicente Paulo, Passo Fundo, RS, Brazil
[5] Univ Passo Fundo, Multidisciplinary Residency Program, Passo Fundo, RS, Brazil
关键词
Gentamicin; UHPLC-MS/MS; Dried blood spots; Therapeutic drug monitoring; Neonates; MASS-SPECTROMETRY; LC-MS/MS; HEMATOCRIT; TIME;
D O I
10.1016/j.jpba.2021.114448
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Gentamicin sulfate (GEN) is an aminoglycoside antibiotic with a narrow therapeutic range of plasma concentrations. The collection of venous blood represents a significant burden for patients, especially in neonatology. Dried blood spots (DBS) obtained from capillary blood can be an alternative for drug measurements in this particular population. This study aimed to develop and validate an assay for the quantification of GEN in DBS using UHPLC-MS/MS. Total GEN concentrations were obtained by adding the individual concentrations of the GEN forms C1, C1a, and C2. The assay used a DBS disk containing approximately 17 mu L of blood for GEN quantitation in the range of 0.1-40 mg L-1. Measurement accuracy for total GEN was in the range of 102.6-108.6%, inter-assay precision was 11.3-13.1% and intra-assay precision was 9.1-12.8.% GEN was stable for 21 days at - 20 and 8 degrees C, but only for 24 h at room temperature. Blood Hct affected the accuracy within acceptable limits (93.8-95% at Hct% of 30, 104.3-113% at Hct% of 50). Blood spotted volume did not affect GEN measurement accuracy. Concentrations of GEN in DBS obtained after heel pricks were correlated to plasma levels in a small cohort of neonatal patients. However, percentual differences between estimated plasma concentrations and actual plasma levels presented values between - 64-35.3% (average difference of - 1.9%). The use of DBS for the measurement of GEN concentrations can increase access to TDM of this antibiotic due to the ease of sample collection and the facilitated specimen transportation logistics when testing is not available onsite. (C) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Development and validation of the UHPLC-MS/MS method for the quantitative determination of 25 PFAS in dried blood spots
    Galletto, Martina
    Ververi, Christina
    Massano, Marta
    Alladio, Eugenio
    Vincenti, Marco
    Salomone, Alberto
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (26) : 5671 - 5687
  • [2] Development and validation of UHPLC-MS/MS methods for the quantification of colistin in plasma and dried plasma spots
    Cangemi, Giuliana
    Barco, Sebastiano
    Castagnola, Elio
    Tripodi, Gino
    Favata, Fabio
    D'Avolio, Antonio
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2016, 129 : 551 - 557
  • [3] Dried plasma spots for therapeutic monitoring of amikacin: Validation of an UHPLC-MS/MS assay and pharmacokinetic application
    Cezimbra da Silva, Anne Caroline
    Feltraco Lizot, Lilian de Lima
    Bastiani, Marcos Frank
    Antunes, Marina Venzon
    Brucker, Natalia
    Linden, Rafael
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2020, 184
  • [4] Ceramides biomarkers determination in quantitative dried blood spots by UHPLC-MS/MS
    Meikopoulos, Thomas
    Begou, Olga
    Theodoridis, Georgios
    Gika, Helen
    ANALYTICA CHIMICA ACTA, 2023, 1255
  • [5] Development and validation of a UHPLC-MS/MS assay for elvitegravir measurement in human plasma and cerebrospinal fluid
    Ocque, Andrew Jacob
    Hagler, Colleen E.
    Lapham, Jill
    Morse, Gene D.
    Letendre, Scott
    Ma, Qing
    SEPARATION SCIENCE PLUS, 2018, 1 (05) : 325 - 333
  • [6] A novel, rapid and simple UHPLC-MS/MS method for quantification of warfarin in dried blood spots
    Qu, Wenhao
    Li, Xianglian
    Tian, Guidi
    Liu, Linsheng
    Miao, Liyan
    ANALYTICAL BIOCHEMISTRY, 2022, 647
  • [7] Development and Validation of an UHPLC-MS/MS Assay for the Therapeutic Monitoring of Brivaracetam Plasma Concentrations in Patients with Epilepsy
    Mohamed, Susan
    Riva, Roberto
    Contin, Manuela
    THERAPEUTIC DRUG MONITORING, 2020, 42 (03) : 445 - 451
  • [8] Dried blood spot UHPLC-MS/MS analysis of oseltamivir and oseltamivircarboxylate—a validated assay for the clinic
    Gero P. Hooff
    Roland J. W. Meesters
    Jeroen J. A. van Kampen
    Nick A. van Huizen
    Birgit Koch
    Asmar F. Y. Al Hadithy
    Teun van Gelder
    Albert D. M. E. Osterhaus
    Rob A. Gruters
    Theo M. Luider
    Analytical and Bioanalytical Chemistry, 2011, 400 : 3473 - 3479
  • [9] Dried blood spot UHPLC-MS/MS analysis of oseltamivir and oseltamivircarboxylate-a validated assay for the clinic
    Hooff, Gero P.
    Meesters, Roland J. W.
    van Kampen, Jeroen J. A.
    van Huizen, Nick A.
    Koch, Birgit
    Al Hadithy, Asmar F. Y.
    van Gelder, Teun
    Osterhaus, Albert D. M. E.
    Gruters, Rob A.
    Luider, Theo M.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (10) : 3473 - 3479
  • [10] Validation of a high-sensitivity assay for zatebradine in dried blood spots of human blood at pg/ml concentrations using HILIC-MS/MS
    Michi, Marco
    Ferrari, Luca
    BIOANALYSIS, 2010, 2 (11) : 1863 - 1871