Improving plan quality for prostate volumetric-modulated arc therapy

被引:7
|
作者
Wright, Katrina [1 ]
Ferrari-Anderson, Janet [1 ]
Barry, Tamara [1 ]
Bernard, Anne [2 ]
Brown, Elizabeth [1 ]
Lehman, Margot [1 ,3 ]
Pryor, David [1 ,3 ,4 ]
机构
[1] Princess Alexandra Hosp, Radiat Oncol Dept, 199 Ipswich Rd, Brisbane, Qld 4102, Australia
[2] Univ Queensland, QFAB Bioinformat, Inst Mol Biosci, Brisbane, Qld, Australia
[3] Univ Queensland, Sch Med, Brisbane, Qld, Australia
[4] Queensland Univ Technol, Australian Prostate Canc Res Ctr Queensland, Inst Hlth & Biomed Innovat, Fac Hlth, Brisbane, Qld, Australia
关键词
Prostate; VMAT; Optimization; Quality assurance; ENERGY; VMAT;
D O I
10.1016/j.meddos.2017.07.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
We critically evaluated the quality and consistency of volumetric-modulated arc therapy (VMAT) prostate planning at a single institution to quantify objective measures for plan quality and establish clear guidelines for plan evaluation and quality assurance. A retrospective analysis was conducted on 34 plans generated on the Pinnacle(3) version 9.4 and 9.8 treatment planning system to deliver 78 Gy in 39 fractions to the prostate only using VMAT. Data were collected on contoured structure volumes, overlaps and expansions, planning target volume (PTV) and organs at risk volumes and relationship, dose volume histogram, plan conformity, plan homogeneity, low dose wash, and beam parameters. Standard descriptive statistics were used to describe the data. Despite a standardized planning protocol, we found variability was present in all steps of the planning process. Deviations from protocol contours by radiation oncologists and radiation therapists occurred in 12% and 50% of cases, respectively, and the number of optimization parameters ranged from 12 to 27 (median 17). This contributed to conflicts within the optimization process reflected by the mean composite objective value of 0.07 (range 0.01 to 0.44). Methods used to control low-intermediate dose wash were inconsistent. At the PTV rectum interface, the dose-gradient distance from the 74.1 Gy to 40 Gy isodose ranged from 0.6 cm to 2.0 cm (median 1.0 cm). Increasing collimator angle was associated with a decrease in monitor units and a single full 6 MV arc was sufficient for the majority of plans. A significant relationship was found between clinical target volume-rectum distance and rectal tolerances achieved. A linear relationship was determined between the PTV volume and volume of 40 Gy isodose. Objective values and composite objective values were useful in determining plan quality. Anatomic geometry and overlap of structures has a measurable impact on the plan quality achieved for prostate patients being treated with VMAT. By evaluating multiple planning variables, we have been able to determine important factors influencing plan quality and develop predictive models for quality metrics that have been incorporated into our new protocol and will be tested and refined in future studies. Crown Copyright (C) 2017 Published by Elsevier Inc. on behalf of American Association of Medical Dosimetrists. All rights reserved.
引用
收藏
页码:348 / 356
页数:9
相关论文
共 50 条
  • [11] Volumetric-modulated arc therapy planning using multicriteria optimization for localized prostate cancer
    Ghandour, Sarah
    Matzinger, Oscar
    Pachoud, Marc
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 258 - 269
  • [12] Dosimetric comparison of hybrid volumetric-modulated arc therapy, volumetric-modulated arc therapy, and intensity-modulated radiation therapy for left-sided early breast cancer
    Lin, Jia-Fu
    Yeh, Dah-Cherng
    Yeh, Hui-Ling
    Chang, Chen-Fa
    Lin, Jin-Ching
    MEDICAL DOSIMETRY, 2015, 40 (03) : 262 - 267
  • [13] Impact of machines on plan quality: volumetric modulated arc therapy and intensity modulated radiation therapy
    Clemente, S.
    Cozzolino, M.
    Oliviero, C.
    Fiorentino, A.
    Chiumento, C.
    Fusco, V.
    CLINICAL & TRANSLATIONAL ONCOLOGY, 2014, 16 (02): : 141 - 146
  • [14] Impact of machines on plan quality: volumetric modulated arc therapy and intensity modulated radiation therapy
    S. Clemente
    M. Cozzolino
    C. Oliviero
    A. Fiorentino
    C. Chiumento
    V. Fusco
    Clinical and Translational Oncology, 2014, 16 : 141 - 146
  • [15] Dosimetric Effect of Jaw Tracking in Volumetric-Modulated Arc Therapy
    Thongsawad, Sangutid
    Khamfongkhruea, Chirasak
    Tannanonta, Chirapha
    JOURNAL OF MEDICAL PHYSICS, 2018, 43 (01) : 52 - 57
  • [16] Volumetric-modulated arc therapy for adjuvant treatment of endometrial cancer
    Cima, Simona
    Cima, S.
    Valli, M. C.
    Richetti, A.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2019, 195 (12) : 1142 - 1143
  • [17] Use of volumetric-modulated arc therapy for treatment of Hodgkin lymphoma
    Lee, Young K.
    Bedford, James L.
    Taj, Mary
    Saran, Frank H.
    MEDICAL DOSIMETRY, 2013, 38 (04) : 372 - 375
  • [18] Analysis of a volumetric-modulated arc therapy (VMAT) single phase prostate template as a class solution
    Hoffmann, Matthew
    Pacey, Jacqueline
    Goodworth, Josie
    Laszcyzk, Andrea
    Ford, Richard
    Chick, Brendon
    Greenham, Stuart
    Westhuyzen, Justin
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2019, 24 (01) : 92 - 96
  • [19] A generalized inverse planning tool for volumetric-modulated arc therapy
    Cao, Daliang
    Afghan, Muhammad K. N.
    Ye, Jinsong
    Chen, Fan
    Shepard, David M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (21): : 6725 - 6738
  • [20] Assessment of the robustness of volumetric-modulated arc therapy for lung radiotherapy
    Edmunds, K.
    Bedford, J.
    BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1023):