Microneedle Insertion into Visco-Hyperelastic Model for Skin for Healthcare Application

被引:3
|
作者
Widianto, Davira P. [1 ]
Stewart, Benjamin G. [1 ]
Mena-Lapaix, Juan L. [2 ]
Shafer, Richard H. [3 ]
Burns, Andrew [4 ]
Prausnitz, Mark R. [2 ]
Alizadeh, Azar [4 ]
Sitaraman, Suresh K. [1 ]
机构
[1] Georgia Inst Technol, George W Woodruff Sch Mech Engn, Atlanta, GA 30332 USA
[2] Georgia Inst Technol, Sch Chem & Biomol Engn, Atlanta, GA 30332 USA
[3] Georgia Inst Technol, Inst Elect & Nanotechnol, Flexible Wearable Elect Adv Res Program, Atlanta, GA 30332 USA
[4] GE Res, Niskayuna, NY USA
关键词
microneedle; cohesive zone modeling; finite-element analysis; nonlinear constitutive behavior; PARAMETERS; BEHAVIOR; FAILURE;
D O I
10.1109/ECTC32696.2021.00236
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Recently, microneedle patches have been explored for extracting interstitial fluid with the goal of extracting temporally relevant, clinical-grade information for human health monitoring. As compared to traditional hypodermic needles, the sub-millimeter scale of microneedles allows for the creation of micropores providing access into human skin interstitial fluid while minimizing interactions with blood vessels and nerves, leading to painless insertion with little to no bleeding. An essential sub-component is the actuator, responsible for driving the microneedle into the skin with a precise force and velocity to ensure reliable insertion. Reliability, in this case, consists of two criteria: the ability of the microneedle to 1) penetrate the skin, and 2) withstand penetration forces without mechanical failure. Evaluation of these criteria requires a thorough understanding of the non-linear, time-dependent interactions between the microneedle and the skin during insertion, including rupture and tearing of the skin on the micron scale, and the resultant stresses on the microneedle. To this end, a comprehensive finite-element model was developed to simulate the microneedle insertion process. This analysis yielded a prediction of complete microneedle insertion without failure of the microneedle and an estimated insertion force of 0.055 N per microneedle, well within the capability of the actuator system considered. This insertion force was validated using experimental data obtained through microneedle insertion in whole skin samples. The model was then used to perform several parametric studies, yielding valuable insights for possible future design improvements.
引用
收藏
页码:1481 / 1489
页数:9
相关论文
共 50 条
  • [21] Visco-Hyperelastic Model for Soft Rubber-like Materials
    Mohammed, Mohd Afandi P.
    SAINS MALAYSIANA, 2014, 43 (03): : 451 - 457
  • [22] Rheological model for stress-softened visco-hyperelastic materials
    Tauheed, Firozut
    Sarangi, Somnath
    INTERNATIONAL JOURNAL OF DAMAGE MECHANICS, 2015, 24 (05) : 728 - 749
  • [23] A visco-hyperelastic constitutive description of elastomeric foam
    Yang, LM
    Shim, VPW
    INTERNATIONAL JOURNAL OF IMPACT ENGINEERING, 2004, 30 (8-9) : 1099 - 1110
  • [24] A visco-hyperelastic formulation for the rheology of immiscible blends
    Yao, Donggang
    JOURNAL OF RHEOLOGY, 2012, 56 (04) : 767 - 795
  • [25] Study on the Visco-Hyperelastic Behavior of the Zona Pellucida
    Boccaccio, A.
    Lamberti, L.
    Papi, M.
    Douet, C.
    Goudet, G.
    De Spirito, M.
    Pappalettere, C.
    ADVANCEMENT OF OPTICAL METHODS IN EXPERIMENTAL MECHANICS, VOL 3, 2015, : 53 - 62
  • [26] Microstructural buckling in soft visco-hyperelastic laminates
    Lapina, Tatiana
    Xiang, Yuhai
    Yao, Qi
    Chen, Dean
    Li, Ji
    Steinmann, Paul
    Rudykh, Stephan
    INTERNATIONAL JOURNAL OF SOLIDS AND STRUCTURES, 2025, 311
  • [27] A Visco-Hyperelastic Constitutive Model for Multilayer Polymer Membranes and its Application in Packaging Air Cushion
    Liu, Yingfeng
    Rao, Qiong
    Chen, Ming
    Peng, Xiongqi
    Shi, Shaoqing
    INTERNATIONAL JOURNAL OF APPLIED MECHANICS, 2016, 8 (05)
  • [28] Fast porous visco-hyperelastic soft tissue model for surgery simulation: Application to liver surgery
    Marchesseau, Stephanie
    Heimann, Tobias
    Chatelin, Simon
    Willinger, Remy
    Delingette, Herve
    PROGRESS IN BIOPHYSICS & MOLECULAR BIOLOGY, 2010, 103 (2-3): : 185 - 196
  • [29] A structurally frame-indifferent model for anisotropic visco-hyperelastic materials
    Ciambella, J.
    Nardinocchi, P.
    JOURNAL OF THE MECHANICS AND PHYSICS OF SOLIDS, 2021, 147
  • [30] A visco-hyperelastic model with damage for the knee ligaments under dynamic constraints
    Arnoux, P.J.
    Chabrand, P.
    Jean, M.
    Bonnoit, J.
    Computer Methods in Biomechanics and Biomedical Engineering, 2002, 5 (02) : 167 - 174