Ionization chamber dosimetry of small photon fields:: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams

被引:0
|
作者
Sánchez-Doblado, F
Andreo, P
Capote, R
Leal, A
Perucha, M
Arráns, R
Núñez, L
Mainegra, E
Lagares, JI
Carrasco, E
机构
[1] Hosp Univ Virgen Macarena, E-41009 Seville, Spain
[2] Univ Seville, Dipartimento Fis Med & Biofis, Seville, Spain
[3] Univ Stockholm, Karolinska Inst, Div Med Radiat Phys, SE-17176 Stockholm, Sweden
[4] Clin Puerta Hierro, Madrid, Spain
[5] Natl Res Council Canada, Ottawa, ON, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2003年 / 48卷 / 14期
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Absolute dosimetry with ionization chambers of the narrow photon fields used in stereotactic techniques and IMRT beamlets is constrained by lack of electron equilibrium in the radiation field. It is questionable that stopping-power ratio in dosimetry protocols, obtained for broad photon beams and quasi-electron equilibrium conditions, can be used in the dosimetry of narrow fields while keeping the uncertainty at the game level as for the broad beams used in accelerator calibrations. Monte Carlo simulations have been performed for two 6 MV clinical accelerators (Elekta SL-18 and Siemens Mevatron Primus), equipped with radiosurgery applicators and MLC. Narrow circular and Z-shaped on-axis and off-axis fields, as well as broad IMRT configured beams, have been simulated together with reference 10 x 10 cm(2) beams. Phase-space data have been used to generate 3D dose distributions which have been compared satisfactorily with experimental profiles (ion chamber, diodes and film). Photon and electron spectra at various depths in water have been calculated, followed by Spencer-Attix (Delta = 10 keV) stopping-power ratio calculations which have been compared to those used in the IAEA TRS-398 code of practice. For water/air and PMMA/air stopping-power ratios, agreements within 0.1% have been obtained for the 10 x 10 cm2 fields. For radiosurgery applicators and narrow MLC beams, the calculated s(w,air) values agree with the reference within +/-0.3%, well within the estimated standard uncertainty of the reference stopping-power ratios (0.5%). Ionization chamber dosimetry of narrow beams at the photon qualities used in this work (6 MV) can therefore be based on stopping-power ratios data in dosimetry protocols. For a modulated 6 MV broad beam used in clinical IMRT, s(w,air) agrees within 0.1% with the value for 10 x 10 cm(2) confirming that at low energies IMRT absolute dosimetry can also be based on data for open reference fields. At higher energies (24 MV) the difference in s(w,air) was up to 1.1%, indicating that the use of protocol data for narrow beams in such cases is less accurate than at low energies, and detailed calculations of the dosimetry parameters involved should be performed if similar accuracy to that of 6 MV is sought.
引用
收藏
页码:2081 / 2099
页数:19
相关论文
共 50 条
  • [21] Dosimetry of small beams used in radiosurgery: A comparison between different detectors and Monte Carlo simulation
    Francescon, P
    Cavedon, C
    Cora, S
    Stancanello, J
    Scalchi, P
    Avanzo, M
    MEDICAL PHYSICS, 2005, 32 (06) : 2004 - 2004
  • [22] An Experimental and Monte Carlo Study of Output Factors for Small Radiosurgery Beams
    Yang, Y.
    Kim, J.
    Li, F.
    Li, X.
    Li, T.
    Yue, N.
    Heron, D.
    Huq, M.
    MEDICAL PHYSICS, 2008, 35 (06) : 2785 - +
  • [23] Surface dosimetry for oblique tangential photon beams: A Monte Carlo simulation study
    Chow, James C. L.
    MEDICAL PHYSICS, 2008, 35 (01) : 70 - 76
  • [24] An EGSnrc Monte Carlo study of the microionization chamber for reference dosimetry of narrow irregular IMRT beamlets
    Capote, R
    Sánchez-Doblaclo, F
    Leal, A
    Lagares, JI
    Arráns, R
    Hartmann, GH
    MEDICAL PHYSICS, 2004, 31 (09) : 2416 - 2422
  • [25] Experimental and Monte-Carlo characterization of the novel compact ionization chamber PTW 31023 for reference and relative dosimetry in high energy photon beams
    Buesing, Isabel
    Brant, Andre
    Lange, Tobias
    Delfs, Bjoern
    Poppinga, Daniela
    Kranzer, Rafael
    Looe, Hui Khee
    Poppe, Bjoern
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (04): : 303 - 313
  • [26] Monte carlo study on 6 MV photon beams of a CyberKnife® stereotactic radiosurgery system
    Dong Xiaoqing
    Luo Wenyun
    Yue Kun
    Wang Chuanshan
    Zhou Shiqing
    Pan Fuxing
    Wang Enmin
    Wang Chaozhuang
    NUCLEAR SCIENCE AND TECHNIQUES, 2010, 21 (01) : 16 - 19
  • [27] Monte carlo study on 6 MV photon beams of a CyberKnife stereotactic radiosurgery system
    DONG Xiaoqing1 LUO Wenyun1 YUE Kun1 WANG Chuanshan1 ZHOU Shiqing1 PAN Fuxing1 WANG Enmin2 WANG Chaozhuang2 1 Shanghai Applied Radiation Institute
    Nuclear Science and Techniques, 2010, 21 (01) : 16 - 19
  • [28] Monte Carlo Calculated Stopping-Power of EBT3 and EBT-XD Film for Pristine Proton Beams
    Shi, C.
    Chen, C.
    Mah, D.
    Chan, M.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [29] Monte Carlo based investigation of ionization chamber wall perturbation factors for high energy photon beams
    Wulff, J.
    Ubrich, F.
    Heverhagen, J.
    Zink, K.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S113 - S113
  • [30] DOSIMETRY OF A SMALL AIR CAVITY FOR CLINICAL ELECTRON BEAMS: A MONTE CARLO STUDY
    Chow, James C. L.
    Grigorov, Grigor N.
    MEDICAL DOSIMETRY, 2010, 35 (02) : 92 - 100