Investigation of Imaging and Spectroscopy Performances of a 1" x 1" LaBr3:Ce Scintillator Readout by Silicon Drift Detectors for Nuclear Physics Measurements

被引:0
|
作者
Busca, P. [1 ,2 ]
Butt, A. D. [1 ,2 ]
Fiorini, C. [1 ,2 ]
Marone, A. [1 ,2 ]
Occhipinti, M. [1 ,2 ]
Peloso, R. [1 ,2 ]
Quaglia, R. [1 ,2 ]
Camera, F. [2 ,3 ]
Giaz, A. [2 ]
Million, B. [2 ]
Giacomini, G. [4 ]
Piemonte, C. [4 ]
机构
[1] Politecn Milan, Dipartimento Elettron Informaz & Bioingn, I-20133 Milan, Italy
[2] INFN, Sez Milano, I-20133 Milan, Italy
[3] Univ Milan, Dipartimento Fis, I-20133 Milan, Italy
[4] FBK, I-38123 Trento, Italy
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this work we propose a gamma detector based on Silicon Drift Detectors to readout a 1" x 1" LaBr3:Ce scintillator for gamma-ray spectroscopy and imaging application. Position sensitivity could be useful in nuclear physics research to limit the worsening effect of Doppler broadening in gamma-ray spectroscopy at relativistic velocities. The main difficulty in determining the position of the gamma-ray interaction in the crystal is associated to the thickness/diameter ratio of the crystal (in our case 1:1) and the use of reflectors on all lateral and top sides of the crystal. This last choice enhances energy resolution, but makes imaging capability more challenging because light is spread over all photodetectors. Preliminary results show that the camera is able to detect shifts in the measured signals, when a Cs-137 (662 keV) source is moved with steps of 5 mm. A modified version of the centroid method is finally implemented to evaluate the imaging capability of the system.
引用
收藏
页数:4
相关论文
共 27 条
  • [21] GAMMA-RAY SIMULATED SPECTRUM DECONVOLUTION OF A LaBr3 1-x 1-in. SCINTILLATOR FOR NONDESTRUCTIVE ATR FUEL BURNUP ON-SITE PREDICTIONS
    Navarro, Jorge
    Ring, Terry A.
    Nigg, David W.
    NUCLEAR TECHNOLOGY, 2015, 190 (02) : 183 - 192
  • [22] Portable system for imaging of α, β, and X-ray sources with silicon pixel detectors and Medipix1 readout
    Bertolucci, E
    Boerkamp, T
    Maiorino, M
    Mettivier, G
    Montesi, MC
    Russo, P
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (04) : 1845 - 1850
  • [23] Detectors for high-resolution gamma-ray imaging based on a single CsI (T1) scintillator coupled to an array of silicon drift detectors
    Fiorini, C
    Longoni, A
    Perotti, F
    Labanti, C
    Rossi, E
    Lechner, P
    Strüder, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 346 - 350
  • [24] X-ray photoemission spectroscopy investigation of Ce1-xEuxCrO3 nano-powders
    Taheri, M.
    Konuma, M.
    Razavi, F. S.
    SURFACE AND INTERFACE ANALYSIS, 2017, 49 (02) : 122 - 126
  • [25] 1 cm2 and 3 cm2 Multi-Linear Silicon Drift Detectors for 2D X-Ray Spectroscopic Imaging and Compton Scattering
    Castoldi, Andrea
    Guazzoni, Chiara
    Hartmann, Robert
    Strueder, Lothar
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 23 - +
  • [26] NUCLEAR REACTION MEASUREMENTS WITH H-1, H-2, HE-3 AND HE-4 INCIDENT ON SILICON DETECTORS
    KING, TR
    KRAUSHAAR, JJ
    RISTINEN, RA
    SMYTHE, R
    STUPIN, DM
    NUCLEAR INSTRUMENTS & METHODS, 1970, 88 (01): : 17 - +
  • [27] Discovery of a silicon-based ferrimagnetic wheel structure in VxSi12- (x=1-3) clusters: photoelectron spectroscopy and density functional theory investigation
    Huang, Xiaoming
    Xu, Hong-Guang
    Lu, Shengjie
    Su, Yan
    King, R. B.
    Zhao, Jijun
    Zheng, Weijun
    NANOSCALE, 2014, 6 (24) : 14617 - 14621