Influence of deep learning reconstruction on task-based model observer performance in CT: an anthropomorphic head phantom study

被引:1
|
作者
Hernandez-Giron, Irene [1 ]
Kaasalainen, Touko [2 ,3 ]
Makela, Teemu [2 ,3 ]
Peltonen, Juha [2 ,3 ]
Kortesniemi, Mika [2 ,3 ]
机构
[1] Leiden Univ Med Ctr LUMC, Radiol Dept, Div Image Proc, Leiden, Netherlands
[2] Univ Helsinki, HUS Diagnost Ctr, Helsinki, Finland
[3] Helsinki Univ Hosp, Helsinki, Finland
关键词
Model observer; CT; deep learning reconstruction; anthropomorphic phantom; detectability; IMAGE QUALITY;
D O I
10.1117/12.2612649
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
Deep learning-based (DL) reconstruction has been introduced in CT, with two major manufacturers offering such methods in the clinic, which are trained mostly on patient data (or a combination of patient and phantom data). Our purpose was to investigate the influence of DL-based reconstruction on object detectability compared to the current standard of iterative reconstruction in CT head routine protocols, combining a model observer analyzing the detectability of lesion-like objects (brain, bone and lung tissue equivalent, 5 mm diameter, 25mm length) in a commercial anthropomorphic head phantom. The phantom was scanned 10 times in two CT systems (same manufacturer, different model) with routine head protocol and images reconstructed with FBP, iterative (IR) and deep-learning (DL) based methods. As input for the model observer, ROIs were subtracted centered on the locations of the cylinders and for each of them four background locations were selected nearby. The locations of the ROIs in the phantom were analogous for both scanners' data. The non-prewhitening matched filter with an eye filter (NPWE) model observer was applied (Burgess eye filter, peak at 4 cy/deg, 50 cm eye-monitor distance). In visual inspection, the phantom brain background ROIs showed differences in noise texture between the reconstruction methods, with a more uniform distribution for DL-based methods in both CT systems. The average d' and range was, for system 1: [lung-FBP: -124.9 (-178.2, -99.1); IR: -126.7 (-188.2; -102.9); DL:-136.2 (-181.9, - 119.3)]; [bone-FBP: 206.7 (166.7, 269.7); IR: 215.4 (175.8, 278.1); DL: 268.3 (215.3, 339.5)]; soft tissue-FBP: -14.6 (-19.6, -9.8); IR: -15.5 (-20.7; -10.2); DL:-18.8 (-24.6, -10.6)]. The NPWE model obtained consistent higher d' values in the DL-based reconstructed images compared to iterative and FBP for the three materials for both systems.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] The Discriminative Power and Stability of Radiomics Features With Computed Tomography Variations Task-Based Analysis in an Anthropomorphic 3D-Printed CT Phantom
    Jimenez-Del-Toro, Oscar
    Aberle, Christoph
    Bach, Michael
    Schaer, Roger
    Obmann, Markus M.
    Flouris, Kyriakos
    Konukoglu, Ender
    Stieltjes, Bram
    Mueller, Henning
    Depeursinge, Adrien
    INVESTIGATIVE RADIOLOGY, 2021, 56 (12) : 820 - 825
  • [22] Correlation between a deep-learning-based model observer and human observer for a realistic lung nodule localization task in chest CT
    Gong, Hao
    Walther, Andrew
    Hu, Qiyuan
    Koo, Chi Wan
    Takahashi, Edwin A.
    Levin, David L.
    Johnson, Tucker F.
    Hora, Megan J.
    Leng, Shuai
    Fletcher, J. G.
    McCollough, Cynthia H.
    Yu, Lifeng
    MEDICAL IMAGING 2019: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2019, 10952
  • [23] Super resolution deep learning reconstruction for coronary CT angiography: A structured phantom study
    Higaki, Toru
    Tatsugami, Fuminari
    Ohana, Mickael
    Nakamura, Yuko
    Kawashita, Ikuo
    Awai, Kazuo
    EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2024, 12
  • [24] Detectability of Hypoattenuating Liver Lesions with Deep Learning CT Reconstruction: A Phantom and Patient Study
    Cao, Jinjin
    Mroueh, Nayla
    Mercaldo, Nathaniel
    Lennartz, Simon
    Kongboonvijit, Sasiprang
    Rao, Shravya Srinivas
    Pisuchpen, Nisanard
    Baliyan, Vinit
    Pierce, Theodore T.
    Anderson, Mark A.
    Sertic, Madeleine
    Shenoy-Bhangle, Anuradha S.
    Kambadakone, Avinash R.
    RADIOLOGY, 2024, 313 (01)
  • [25] Performance evaluation of deep learning image reconstruction algorithm for dual-energy spectral CT imaging: A phantom study
    Li, Haoyan
    Li, Zhentao
    Gao, Shuaiyi
    Hu, Jiaqi
    Yang, Zhihao
    Peng, Yun
    Sun, Jihang
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2024, 32 (03) : 513 - 528
  • [26] Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms
    Richard, Samuel
    Husarik, Daniela B.
    Yadava, Girijesh
    Murphy, Simon N.
    Samei, Ehsan
    MEDICAL PHYSICS, 2012, 39 (07) : 4115 - 4122
  • [27] Performance of clinically available deep learning image reconstruction in computed tomography: a phantom study
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Takata, Tadanori
    Mitsui, Wataru
    Ueta, Hiroshi
    Yoneda, Norihide
    Kobayashi, Satoshi
    JOURNAL OF MEDICAL IMAGING, 2020, 7 (06)
  • [28] Image quality improvement with deep learning-based reconstruction on abdominal ultrahigh-resolution CT: A phantom study
    Shirasaka, Takashi
    Kojima, Tsukasa
    Funama, Yoshinori
    Sakai, Yuki
    Kondo, Masatoshi
    Mikayama, Ryoji
    Hamasaki, Hiroshi
    Kato, Toyoyuki
    Ushijima, Yasuhiro
    Asayama, Yoshiki
    Nishie, Akihiro
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (07): : 286 - 296
  • [29] A systematic task-based image quality assessment of photon-counting and energy integrating CT as a function of reconstruction kernel and phantom size
    Bhattarai, Mridul
    Bache, Steve
    Abadi, Ehsan
    Samei, Ehsan
    MEDICAL PHYSICS, 2024, 51 (02) : 1047 - 1060
  • [30] Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data
    J. Greffier
    S. Si-Mohamed
    D. Dabli
    H. de Forges
    A. Hamard
    P. Douek
    J. P. Beregi
    J. Frandon
    European Radiology, 2021, 31 : 5324 - 5334