Accelerometer calibration to 50 kHz with a quadrature laser interferometer

被引:0
|
作者
Sill, RD [1 ]
机构
[1] Endevco Corp, San Juan Capistrano, CA 92675 USA
关键词
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Absolute accelerometer calibration is performed over the extended range of 80 Hz to 50 000 Hz using a quadrature Michelson laser interferometer, a beryllium electromagnetic shaker and a unique signal conditioning algorithm. High-speed digitization and an advanced air bearing for the armature allows conventional fringe counting at the lower range of frequencies. A different technique is used at higher frequencies, where vibration displacements are much smaller than a wavelength. By effectively calibrating and numerically normalizing the output of the optical detectors, this system uses a digital tracking filter to pick out sub-wavelength components of motion. The beryllium armature provides adequate stiffness to allow a back-to-back arrangement, with the Unit Under Test on top, and the mirror and laser underneath.
引用
收藏
页码:767 / 773
页数:7
相关论文
共 50 条
  • [31] Calibration of elongation of fiber Bragg gratings by laser interferometer
    Mikel, Bretislav
    Cizek, Martin
    Holik, Milan
    Cip, Ondrej
    SIXTH INTERNATIONAL SYMPOSIUM ON PRECISION MECHANICAL MEASUREMENTS, 2013, 8916
  • [32] Design, alignment, and calibration of a focused laser differential interferometer
    Neet, Mallory C.
    Lawson, Joel M.
    Austin, Joanna M.
    APPLIED OPTICS, 2021, 60 (26) : 7903 - 7909
  • [33] APPLICATION OF A LASER INTERFEROMETER FOR THE CALIBRATION OF ANGULAR INDEXING.
    Fan, Kuang-Chao
    Chung-Kuo Chi Hsueh Kung Ch'eng Hsueh Pao/Journal of the Chinese Society of Mechanical Engineers, 1988, 9 (01): : 63 - 69
  • [34] ACCELEROMETER CALIBRATION USING A LASER-DOPPLER DISPLACEMENT METER
    TUSTIN, W
    LIN, F
    SOUND AND VIBRATION, 1995, 29 (04): : 12 - +
  • [35] Comparison of hydrophone calibration by reciprocity and heterodyne interferometer in the frequency range 500 kHz to 15 MHz
    Xing, Guangzhen
    Yang, Ping
    He, Longbiao
    MEASUREMENT TECHNOLOGY AND ITS APPLICATION, PTS 1 AND 2, 2013, 239-240 : 10 - 15
  • [36] PRECISION PHASE ANGLE CALIBRATION STANDARD FOR FREQUENCIES UP TO 50 KHZ.
    Turgel, Raymond S.
    IEEE Transactions on Instrumentation and Measurement, 1985, IM-34 (04): : 509 - 516
  • [38] Dynamic probe calibration up to 10 kHz using laser interferometry
    Eindhoven Univ of Technology, Eindhoven, Netherlands
    Meas J Int Meas Confed, 3 (107-111):
  • [39] Enhanced ellipse fitting in a two-detector homodyne quadrature laser interferometer
    Pozar, T.
    Mozina, J.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2011, 22 (08)
  • [40] Homodyne Quadrature Laser Interferometer Applied for the Studies of Optodynamic Wave Propagation in a Rod
    Pozar, Tomaz
    Mozina, Janez
    STROJNISKI VESTNIK-JOURNAL OF MECHANICAL ENGINEERING, 2009, 55 (10): : 575 - 580