Accelerometer calibration to 50 kHz with a quadrature laser interferometer

被引:0
|
作者
Sill, RD [1 ]
机构
[1] Endevco Corp, San Juan Capistrano, CA 92675 USA
关键词
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Absolute accelerometer calibration is performed over the extended range of 80 Hz to 50 000 Hz using a quadrature Michelson laser interferometer, a beryllium electromagnetic shaker and a unique signal conditioning algorithm. High-speed digitization and an advanced air bearing for the armature allows conventional fringe counting at the lower range of frequencies. A different technique is used at higher frequencies, where vibration displacements are much smaller than a wavelength. By effectively calibrating and numerically normalizing the output of the optical detectors, this system uses a digital tracking filter to pick out sub-wavelength components of motion. The beryllium armature provides adequate stiffness to allow a back-to-back arrangement, with the Unit Under Test on top, and the mirror and laser underneath.
引用
收藏
页码:767 / 773
页数:7
相关论文
共 50 条
  • [1] Accelerometer sensitivity calibration by phase differences of a single channel laser interferometer: theory
    Chen, S
    SIXTH INTERNATIONAL CONFERENCE ON VIBRATION MEASUREMENTS BY LASER TECHNIQUES: ADVANCES AND APPLICATIONS, 2004, 5503 : 417 - 422
  • [2] Shock measurements with a homodyne interferometer for accelerometer calibration
    Rojas-Ramírez, S
    Elías-Juárez, A
    Schlaak, HJ
    von Martens, HJ
    FOURTH INTERNATIONAL CONFERENCE ON VIBRATION MEASUREMENTS BY LASER TECHNIQUES: ADVANCES AND APPLICATIONS, 2000, 4072 : 146 - 157
  • [3] Measurement of exciting pulse for calibration of high-g accelerometer using lateral laser interferometer
    Fan, Jinbiao
    Li, Xi
    Xu, Peng
    Zu, Jing
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2015, 44 (02): : 497 - 502
  • [4] A 50 m laser interferometer for automatic calibration of surveying tapes using wireless communication
    Kim, Jae Wan
    Jeong, Don Young
    Kim, Myung Soon
    Kang, Chu-Shik
    Kim, Jong-Ahn
    Eom, Tae Bong
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2008, 19 (01)
  • [5] LASER INTERFEROMETER CALIBRATION SYSTEM FOR EXTENSOMETERS
    HADE, G
    CONNER, M
    KUO, JT
    BULLETIN OF THE SEISMOLOGICAL SOCIETY OF AMERICA, 1968, 58 (05) : 1379 - &
  • [6] A 5-KHZ TO 50-KHZ SYNTHETIC CYLINDRICAL NEARFIELD CALIBRATION ARRAY
    VANBUREN, AL
    JEVNAGER, MD
    TIMS, AC
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1985, 77 (05): : 1927 - 1932
  • [7] DYNAMIC FORCE CALIBRATION BY LASER INTERFEROMETER
    Feng, Meng
    Min, Zhang Zhi
    Yue, Zhang
    Zhangwei
    SEVENTH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2011, 8321
  • [8] A laser interferometer accelerometer for vibration sensitive cryogenic experiments
    Bajpai, R.
    Tomaru, T.
    Yamamoto, K.
    Ushiba, T.
    Kimura, N.
    Suzuki, T.
    Yamada, T.
    Honda, T.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2022, 33 (08)
  • [9] HETERODYNE LASER INTERFEROMETER WITH ELECTRICAL FEEDBACK TO MAINTAIN QUADRATURE
    HUBER, DR
    SETTERSTROM, RD
    KNIGHT, LV
    BULLETIN OF THE AMERICAN PHYSICAL SOCIETY, 1977, 22 (09): : 1151 - 1151
  • [10] Dual-probe homodyne quadrature laser interferometer
    Pozar, Tomaz
    Mozina, Janez
    APPLIED OPTICS, 2012, 51 (18) : 4021 - 4027