Task-driven image acquisition and reconstruction in cone-beam CT

被引:25
|
作者
Gang, Grace J. [1 ]
Stayman, J. Webster [1 ]
Ehtiati, Tina [2 ]
Siewerdsen, Jeffrey H. [1 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Siemens Healthcare AX Div, Forcheim, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2015年 / 60卷 / 08期
关键词
task-driven imaging; detectability index; cone-beam CT; tube current modulation; reconstruction kernel; orbital tilt; image quality; HUMAN-OBSERVER PERFORMANCE; CASCADED SYSTEMS-ANALYSIS; COMPUTED-TOMOGRAPHY; NOISE POWER; OPTIMIZATION; SIGNAL; ARM; DETECTABILITY; QUALITY; MODEL;
D O I
10.1088/0031-9155/60/8/3129
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This work introduces a task-driven imaging framework that incorporates a mathematical definition of the imaging task, a model of the imaging system, and a patient-specific anatomical model to prospectively design image acquisition and reconstruction techniques to optimize task performance. The framework is applied to joint optimization of tube current modulation, view-dependent reconstruction kernel, and orbital tilt in cone-beam CT. The system model considers a cone-beam CT system incorporating a flat-panel detector and 3D filtered backprojection and accurately describes the spatially varying noise and resolution over a wide range of imaging parameters in the presence of a realistic anatomical model. Task-based detectability index (d') is incorporated as the objective function in a task-driven optimization of image acquisition and reconstruction techniques. The orbital tilt was optimized through an exhaustive search across tilt angles ranging +/- 30 degrees. For each tilt angle, the view-dependent tube current and reconstruction kernel (i.e. the modulation profiles) that maximized detectability were identified via an alternating optimization. The task-driven approach was compared with conventional unmodulated and automatic exposure control (AEC) strategies for a variety of imaging tasks and anthropomorphic phantoms. The task-driven strategy outperformed the unmodulated and AEC cases for all tasks. For example, d' for a sphere detection task in a head phantom was improved by 30% compared to the unmodulated case by using smoother kernels for noisy views and distributing mAs across less noisy views (at fixed total mAs) in a manner that was beneficial to task performance. Similarly for detection of a line-pair pattern, the task-driven approach increased d' by 80% compared to no modulation by means of view-dependent mA and kernel selection that yields modulation transfer function and noise-power spectrum optimal to the task. Optimization of orbital tilt identified the tilt angle that reduced quantum noise in the region of the stimulus by avoiding highly attenuating anatomical structures. The task-driven imaging framework offers a potentially valuable paradigm for prospective definition of acquisition and reconstruction protocols that improve task performance without increase in dose.
引用
收藏
页码:3129 / 3150
页数:22
相关论文
共 50 条
  • [31] Reconstruction of a Cone-Beam CT Image Via Forward Iterative Projection Matching
    Brock, R.
    Docef, A.
    Murphy, M.
    MEDICAL PHYSICS, 2010, 37 (06) : 3112 - +
  • [32] Reconstruction of a cone-beam CT image via forward iterative projection matching
    Brock, R. Scott
    Docef, Alen
    Murphy, Martin J.
    MEDICAL PHYSICS, 2010, 37 (12) : 6212 - 6220
  • [33] An Image Reconstruction Strategy for Truncated Projections of Planar Object in Cone-Beam CT
    Zhang, Feng
    Yan, Bin
    Li, Lei
    Xi, Xiaoqi
    Jiang, Hua
    Chen, Siyu
    Cui, Jinxian
    Cong, Dan
    Tian, Ying
    2015 12TH INTERNATIONAL CONFERENCE ON FUZZY SYSTEMS AND KNOWLEDGE DISCOVERY (FSKD), 2015, : 1113 - 1117
  • [34] Image reconstruction in cone-beam CT with a spherical detector using the BPF algorithm
    Zuo, Nianming
    Zou, Yu
    Jiang, Tianzi
    Pan, Xiaochuan
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [35] Image Reconstruction Using OS-EM Method in Cone-beam CT
    Dong, Baoyu
    2008 IEEE INTERNATIONAL SYMPOSIUM ON KNOWLEDGE ACQUISITION AND MODELING WORKSHOP PROCEEDINGS, VOLS 1 AND 2, 2008, : 108 - 111
  • [36] Task-driven source-detector trajectories in cone-beam computed tomography: II. Application to neuroradiology
    Capostagno, Sarah
    Stayman, J. Webster
    Jacobson, Matthew
    Ehtiati, Tina
    Weiss, Clifford R.
    Siewerdsen, Jeffrey H.
    JOURNAL OF MEDICAL IMAGING, 2019, 6 (02)
  • [37] An improved exact FBP algorithm for image reconstruction in cone-beam helical CT
    Ma, Jianhua
    Chen, Wufan
    2006 INTERNATIONAL CONFERENCE ON COMPUTATIONAL INTELLIGENCE AND SECURITY, PTS 1 AND 2, PROCEEDINGS, 2006, : 1635 - 1640
  • [38] Exact image reconstruction in reverse helical cone-beam CT for radiation therapy
    Cho, S.
    Xia, D.
    Pelizzari, C.
    Pan, X.
    MEDICAL PHYSICS, 2007, 34 (06) : 2633 - 2634
  • [39] Statistical Cone-Beam CT Image Reconstruction using the Cell Broadband Engine
    Knaup, Michael
    Kalender, Willi A.
    Kachelriess, Marc
    2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6, 2006, : 2837 - 2840
  • [40] Image reconstruction with a shift-variant filtration in circular cone-beam CT
    Yu, LF
    Pan, XC
    Pelizzari, CA
    INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 2004, 14 (05) : 213 - 221