Measurement of contact angles at room temperature in high magnetic field

被引:10
|
作者
Li, Chuanjun [1 ]
Cao, Yang
Guo, Rui
He, Shengya
Xuan, Weidong
Li, Xi
Zhong, Yunbo
Ren, Zhongming [1 ]
机构
[1] Shanghai Univ, State Key Lab Adv Special Steel, Shanghai 200072, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2017年 / 88卷 / 11期
基金
中国国家自然科学基金;
关键词
SURFACE-TENSION; DROP SIZE; WATER; REDUCTION; SHAPE;
D O I
10.1063/1.5011361
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The contact angle (CA) goniometer adaptable to a superconducting magnet was developed based on the sessile drop method. The goniometer mainly consisted of the sampling system, the supporting system, and the image acquisition system. Some improvements were taken to avoid the effects of the magnetic field (MF) on the CA measurement. As an example, the CAs of water on two substrates of silica and polymethyl methacrylate (PMMA) were measured using the goniometer. The results with and without a MF showed a good repeatability and reliability. Additionally, the MF was found to reduce the CA of water, which probably stemmed from the change of the surface tension in the MF. The CA goniometer will become an important tool which is used to study the wettability of liquids on a solid in the MF. Published by AIP Publishing.
引用
收藏
页数:4
相关论文
共 50 条
  • [31] MEASUREMENT OF CONTACT ANGLES IN PRESENCE OF MASS TRANSFER
    PONTER, AB
    THORNLEY, PG
    NATURE, 1965, 207 (5003) : 1288 - &
  • [32] MEASUREMENT OF CONTACT PARAMETERS OF FLAT ON FLAT CONTACT SURFACES AT HIGH TEMPERATURE
    Botto, D.
    Lavella, M.
    Gola, M. M.
    PROCEEDINGS OF THE ASME TURBO EXPO 2012, VOL 7, PTS A AND B, 2012, : 1325 - 1332
  • [33] A thin film magnetic field sensor of sub-pT resolution and magnetocardiogram (MCG) measurement at room temperature
    Yabukami, S.
    Kato, K.
    Ohtomo, Y.
    Ozawa, T.
    Arai, K. I.
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2009, 321 (07) : 675 - 678
  • [34] Development of pressure cell for specific heat measurement at low temperature and high Magnetic field
    Kawae, T.
    Yaita, K.
    Yoshida, Y.
    Inagaki, Y.
    Ohashi, M.
    Oomi, G.
    Matsubayashi, K.
    Matsumoto, T.
    Uwatoko, Y.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (02):
  • [35] Non-contact high temperature strain measurement
    Wang, Shurong
    Li, Zhigang
    Guangxue Xuebao/Acta Optica Sinica, 1998, 18 (12): : 1760 - 1764
  • [36] Measurement of contact angle and work of adhesion at high temperature
    Eustathopoulos, N
    Sobczak, N
    Passerone, A
    Nogi, K
    JOURNAL OF MATERIALS SCIENCE, 2005, 40 (9-10) : 2271 - 2280
  • [37] Measurement of contact angle and work of adhesion at high temperature
    N. Eustathopoulos
    N. Sobczak
    A. Passerone
    K. Nogi
    Journal of Materials Science, 2005, 40 : 2271 - 2280
  • [38] TEMPERATURE-DEPENDENCE OF CONTACT ANGLES ON ELASTOMERS
    BUDZIAK, CJ
    VARGHABUTLER, EI
    NEUMANN, AW
    JOURNAL OF APPLIED POLYMER SCIENCE, 1991, 42 (07) : 1959 - 1964
  • [39] TILTING-PLATE APPARATUS FOR MEASUREMENT OF CONTACT ANGLES IN CORROSIVE LIQUIDS AS A FUNCTION OF TEMPERATURE AND PRESSURE
    MURACA, RF
    LONGWELL, AP
    BURNS, EA
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1963, 34 (05): : 479 - &