Instrumental and randomised clinical comparison between laryngeal mask airway Proseal and Supreme in pediatric patients

被引:0
|
作者
Santambrogio, Luisa [1 ]
Righi, Sabrina [1 ]
Pinciroli, Rosa L. [1 ]
Piro, Eugenia [2 ]
D'alessio, Antonio [2 ]
Minuto, Aldo [1 ]
机构
[1] ASST Ovestmilanese Civil Hosp Legnano, Anesthesia & Intens Care Dept, Via Papa Giovanni Paolo 2, I-20025 Milan, Italy
[2] ASST Ovestmilanese Civil Hosp Legnano, Pediat Surg Dept, Via Papa Giovanni Paolo 2, I-20025 Milan, Italy
关键词
Laryngeal mask airway; Child; Mechanical ventilation; General anesthesia; Airway management; Supraglottic device; OROPHARYNGEAL LEAK PRESSURE; LMA SUPREME(TM); COHORT EVALUATION; CHILDREN; DEVICES; ANESTHESIA; MANAGEMENT; INSERTION;
D O I
10.1016/j.tacc.2019.11.001
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Laryngeal mask airway (LMA) has proven safe and effective for airway management in children during general anesthesia; mostly the second generation LMAs, none of which resulted clearly better despite previous studies. Aim: Proseal-LMA (PLMA) and Supreme-LMA (SLMA) are compared for structural characteristics and oropharyngeal leak pressure as primary aims, and mechanical ventilation, insertion, intra- and postoperative complications as secondary aims. Trial design: prospective randomised parallel clinical trial. Methods: instrumental evaluation: measurements of length, cross section, loading test on free air were performed to assess the resistance during ventilation. Then, 39 children received PLMA and 39 received SLMA for airway management during elective surgery: intraoperative and postoperative data were analyzed. Results: transversal section area was 35,71 mm(2) in PLMA and 59,61 in SLMA, whilst length was 133,67 and 123,07 mm; the loading test on free air confirmed lower resistance of SLMA. Oropharyngeal leak pressure resulted significant lower in SLMA. No significant difference in other clinical parameters and complications. Conclusions: instrumental measurements highlight that SLMA offers the advantage of less resistance to the airflow, allowing to keep lower oropharyngeal leak pressure during mechanical ventilation. Despite the differences reported, we observed only a few minor complications in both groups. (C) 2019 Elsevier Ltd. All rights reserved.
引用
收藏
页码:14 / 21
页数:8
相关论文
共 50 条
  • [21] Comparisons of clinical performance of guardian laryngeal mask with laryngeal mask airway ProSeal
    Pajiyar, Ajay Kumar
    Wen, Zhiting
    Wang, Haiyun
    Ma, Lin
    Miao, Lumin
    Wang, Guolin
    BMC ANESTHESIOLOGY, 2015, 15
  • [22] A randomized crossover comparison of the size 2 1/2 laryngeal mask airway ProSeal™ versus Laryngeal Mask Airway-Classic™ in pediatric patients
    Goldmann, K
    Jakob, C
    ANESTHESIA AND ANALGESIA, 2005, 100 (06): : 1605 - 1610
  • [23] A randomised trial comparing the laryngeal mask airway Supreme™ with the laryngeal mask airway Unique™ in children
    Jagannathan, N.
    Sohn, L. E.
    Sawardekar, A.
    Chang, E.
    Langen, K. E.
    Anderson, K.
    ANAESTHESIA, 2012, 67 (02) : 139 - 144
  • [24] Resistive load of laryngeal mask airway and proseal laryngeal mask airway in mechanically ventilated patients
    Natalini, G
    Rosano, A
    Lanza, G
    Martinelli, E
    Pletti, C
    Bernardini, A
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2003, 47 (06) : 761 - 764
  • [25] A comparison of the laryngeal tube s and the LMA ProSeal laryngeal mask airway
    Cook, T. M.
    ANAESTHESIA, 2007, 62 (12) : 1297 - 1298
  • [26] Comparison of classic, and ProSeal laryngeal mask airways (LMA) in pediatric patients
    Gungor, Gurcan
    Sutas Bozkurt, Pervin
    Yuceyayar, Lale
    IZMIR DR BEHCET UZ COCUK HASTANESI DERGISI, 2016, 6 (01): : 46 - 52
  • [27] The Proseal™ laryngeal mask airway in children:: a comparison between two insertion techniques
    Teoh, Chuan Yeong
    Lim, Felicia S. K.
    PEDIATRIC ANESTHESIA, 2008, 18 (02) : 119 - 124
  • [28] Comparison of proseal laryngeal mask and endotracheal tube for airway safety in pediatric strabismus surgery
    Kim, Yong H.
    SAUDI MEDICAL JOURNAL, 2012, 33 (08) : 914 - 914
  • [29] Comparison of proseal laryngeal mask and endotracheal tube for airway safety in pediatric strabismus surgery
    Gul, Rauf
    Goksu, Sitki
    Ugur, Berna K.
    Sahin, Levent
    Koruk, Senem
    Okumus, Seydi
    Erbagci, Ibrahim
    SAUDI MEDICAL JOURNAL, 2012, 33 (04) : 388 - 394
  • [30] Airway protection with the ProSeal laryngeal mask airway
    Brimacombe, J
    Keller, C