Evaluation of Four Qualitative Third-Generation HIV Antibody Assays and the Fourth-Generation Abbott HIV Ag/Ab Combo Test

被引:3
|
作者
Wians, Frank H. [1 ]
Moore, Holly A. [1 ]
Briscoe, Donna [1 ]
Anderson, Kathy M. [1 ]
Hicks, Pamela S. [2 ]
Smith, Debbie L. [2 ]
Clark, Tammy A. [3 ]
Preton, Maggie M.
Gammons, Beckie
Ray, Cynthia S.
Bond, Christine
West, Joshua T.
机构
[1] Baylor Univ, Med Ctr, Dept Pathol, Dallas, TX USA
[2] Texas Hlth Arlington Mem Hosp, Arlington, TX USA
[3] Texas Hlth Preshyterian Hosp, Baylor All Saims, Dallas, TX USA
来源
LABMEDICINE | 2011年 / 42卷 / 09期
关键词
human immunodeficiency virus; fourth-generation assay; comparative method; test method; agreement; P24; ANTIGEN; VIRUS; DIAGNOSIS; RISK;
D O I
10.1309/LM38NCGQ0GWJIMZV
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective: Compare the agreement between four third-generation rapid or automated human immunodeficiency virus (HIV) tests vs the fourth-generation Abbott ARCHITECT HIV Ag/Ab Combo assay (Combo Test) (Abbott Laboratories, Abbott Park, IL) in 7 clinical laboratories located throughout the Dallas-Fort Worth, TX, metroplex area. Methods: We tested a total of 220 specimens from the general population of patients being screened for HIV antibody status, including Western blot (WB) and HIV-1 RNA viral load by NAAT in patients with initially positive HIV screening tests. We performed a method precision study of the Combo Test and evaluated the agreement between the 4 comparative methods (CMs) against the Combo Test method (TM) using a 2 x 2 contingency table to determine values for positive percent agreement (PPA), negative percent agreement (NPA), and overall percent agreement (OPA) according to the guideline (EP12-A2) for the evaluation of qualitative assays published by the Clinical Laboratory Standards Institute (CLSI). Results: The Combo Test yielded consistent (reproducible) results when the HIV antibody (or p24 antigen) concentration was 20% away or more from C-50. In addition, values for PPA, NPA, and OPA for CM vs TM result's were all 100%, or not significantly different from 100%. Conclusion: The high detection rate of the Combo Test, when coupled with its high diagnostic sensitivity and specificity, rapid TAT (<30 minutes) of test results using a fully automated instrument (ARCHITECT), and excellent agreement with HIV-1/2 antibody results by several commercially available HIV-1/2 screening tests and with WB and NAAT results, indicate that the Abbott HIV-1/2 Ab/Ag Test is a useful screening test in the identification of HIV-infected individuals.
引用
收藏
页码:523 / 535
页数:13
相关论文
共 50 条
  • [1] HIV-1 virion lysis following centrifugation improves the sensitivity of the Fourth-Generation HIV Ag/Ab combo assay
    Wandera, Allan
    Ssekatawa, Kenneth
    Kato, Charles Drago
    Kwizera, Eliah
    Mujinya, Pastori
    Siida, Robert
    BMC RESEARCH NOTES, 2024, 17 (01)
  • [2] Performance evaluation of a new fourth-generation HIV Ag/Ab combination electrochemiluminescence immunoassay - evaluation of a new HIV assay
    Wang, Tingting
    Li, Dongdong
    Yan, Kening
    Yuan, Yu
    Yang, Tingfu
    Du, Xiaoqing
    Yan, Xuedan
    Tao, Chuanmin
    Wang, Lanlan
    INTERNATIONAL JOURNAL OF STD & AIDS, 2014, 25 (04) : 267 - 272
  • [3] Performance evaluation of a new automated fourth-generation HIV Ag/Ab combination chemiluminescence immunoassay
    Zhao, Yanhua
    Gou, Yu
    Li, Dongdong
    Wang, Tingting
    Huang, Xiyue
    Shi, Mingqiao
    Tao, Chuanmin
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (05) : E115 - E117
  • [4] Performance of the fourth-generation Bio-Rad GS HIV Combo Ag/Ab enzyme immunoassay for diagnosis of HIV infection in Southern Africa
    Piwowar-Manning, Estelle
    Fogel, Jessica M.
    Richardson, Paul
    Wolf, Shauna
    Clarke, William
    Marzinke, Mark A.
    Fiamma, Agnes
    Donnell, Deborah
    Kulich, Michal
    Mbwambo, Jessie K. K.
    Richter, Linda
    Gray, Glenda
    Sweat, Michael
    Coates, Thomas J.
    Eshleman, Susan H.
    JOURNAL OF CLINICAL VIROLOGY, 2015, 62 : 75 - 79
  • [5] Evaluation of the performance of the Abbott ARCHITECT HIV Ag/Ab Combo Assay
    Chavez, Pollyanna
    Wesolowski, Laura
    Patel, Pragna
    Delaney, Kevin
    Owen, S. Michele
    JOURNAL OF CLINICAL VIROLOGY, 2011, 52 : S51 - S55
  • [6] Field evaluation of the Abbott ARCHITECT HIV Ag/Ab Combo immunoassay
    Kfutwah, Anfumbom
    Lemee, Veronique
    Ngono, Helene Valerie
    De Oliveira, Fabienne
    Njouom, Richard
    Plantier, Jean-Christophe
    JOURNAL OF CLINICAL VIROLOGY, 2013, 58 : E70 - E75
  • [7] MULTICENTER EVALUATION OF THREE NOVEL 4TH GENERATION HIV AG/AB COMBO ASSAYS: ABBOTT ARCHITECT, ROCHE HIV COMBI AND SIEMENS ADVIA CENTAUR
    Serhir, B.
    Vincelette, J.
    Frost, E.
    Bergevin, M.
    Beliveau, C.
    Phaneuf, D.
    Sanfacon, R.
    Poirier, A.
    Doualla-Bell, F.
    Tremblay, C.
    SEXUALLY TRANSMITTED INFECTIONS, 2013, 89 : A99 - A99
  • [8] Clinical Experience of Fourth Generation HIV Ag/Ab Combo Test as a Screening Tool in a Low Prevalence Area
    Lee, E.
    Lee, J.
    Park, Y.
    Seok, Y.
    Kim, H.
    Hong, D.
    Choi, J.
    Kim, H.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2010, 12 (06): : 890 - 890
  • [9] Identification of early HIV infections using the fourth generation Abbott ARCHITECT HIV Ag/Ab Combo chemiluminescent microparticle immunoassay (CIA) in San Diego County
    Manlutac, Anna Liza M.
    Giesick, Jill S.
    McVay, Patricia A.
    JOURNAL OF CLINICAL VIROLOGY, 2013, 58 : E44 - E47
  • [10] Performance evaluation of a new fourth-generation HIV combination antigen–antibody assay
    A. Mühlbacher
    H. Schennach
    J. van Helden
    T. Hebell
    G. Pantaleo
    P. Bürgisser
    C. Cellerai
    P. Permpikul
    M. I. Rodriguez
    A. Eiras
    F. Alborino
    P. Cunningham
    M. Axelsson
    S. Andersson
    O. Wetlitzky
    C. Kaiser
    P. Möller
    G. de Sousa
    Medical Microbiology and Immunology, 2013, 202 : 77 - 86