Feasibility study of portable multi-energy computed tomography with photon-counting detector for preclinical and clinical applications

被引:5
|
作者
Lee, Chang-Lae [1 ]
Hong, Key Jo [1 ]
Kim, Namwoo [1 ]
Han, Kwanhee [1 ]
Kim, Dongkyu [2 ]
Jung, Hoe-Su [2 ]
Lee, Sangmin [1 ]
Park, Junyoung [1 ]
Lee, Kyoung-Yong [1 ]
Lee, Jee Eun [3 ]
Choi, Yuna [1 ]
Cho, Minkook [1 ]
机构
[1] Samsung Elect, Hlth & Med Equipment Business Unit, PCD R&D Grp, Suwon, Gyeonggido, South Korea
[2] Daegu Gyeongbuk Med Innovat Fdn, Lab Anim Ctr, Daegu, South Korea
[3] Bupyeong Serim Hosp, Dept Family Med, Incheon, South Korea
关键词
CEREBROSPINAL-FLUID; CT; STROKE; PERFUSION; MATTER; WHITE; MRI;
D O I
10.1038/s41598-021-02210-5
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
In this study, preclinical experiments were performed with an in-house developed prototypal photon-counting detector computed tomography (PCD CT) system. The performance of the system was compared with the conventional energy-integrating detector (EID)-based CT, concerning the basic image quality biomarkers and the respective capacities for material separation. The pre- and the post-contrast axial images of a canine brain captured by the PCD CT and EID CT systems were found to be visually similar. Multi-energy images were acquired using the PCD CT system, and machine learning-based material decomposition was performed to segment the white and gray matters for the first time in soft tissue segmentation. Furthermore, to accommodate clinical applications that require high resolution acquisitions, a small, native, high-resolution (HR) detector was implemented on the PCD CT system, and its performance was evaluated based on animal experiments. The HR acquisition mode improved the spatial resolution and delineation of the fine structures in the canine's nasal turbinates compared to the standard mode. Clinical applications that rely on high-spatial resolution expectedly will also benefit from this resolution-enhancing function. The results demonstrate the potential impact on the brain tissue segmentation, improved detection of the liver tumors, and capacity to reconstruct high-resolution images both preclinically and clinically.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Dedicated Breast Computed Tomography With a Photon-Counting Detector Initial Results of Clinical In Vivo Imaging
    Berger, Nicole
    Marcon, Magda
    Saltybaeva, Natalia
    Kalender, Willi A.
    Alkadhi, Hatem
    Frauenfelder, Thomas
    Boss, Andreas
    INVESTIGATIVE RADIOLOGY, 2019, 54 (07) : 409 - 418
  • [22] Review of Technical Advancements and Clinical Applications of Photon-counting Computed Tomography in Imaging of the Thorax
    Farhadi, Faraz
    Rajagopal, Jayasai R.
    Nikpanah, Moozhan
    Sahbaee, Pooyan
    Malayeri, Ashkan A.
    Pritchard, William F.
    Samei, Ehsan
    Jones, Elizabeth C.
    Chen, Marcus Y.
    JOURNAL OF THORACIC IMAGING, 2021, 36 (02) : 84 - 94
  • [23] Performance of an improved readout IC for multi-energy photon-counting x-ray detector arrays
    Clajus, Martin
    Cajipe, Victoria B.
    Hayakawa, Satoshi
    Tuemer, Tuemay O.
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS X, 2008, 7079
  • [24] Detection of Microcalcifications in Spiral Breast Computed Tomography with Photon-Counting Detector Is Feasible: A Specimen Study
    Wetzl, Matthias
    Wenkel, Evelyn
    Balbach, Eva
    Dethlefsen, Ebba
    Hartmann, Arndt
    Emons, Julius
    Kuhl, Christiane
    Beckmann, Matthias W.
    Uder, Michael
    Ohlmeyer, Sabine
    DIAGNOSTICS, 2021, 11 (05)
  • [25] Compton scatter tomography with photon-counting detector: a preliminary study
    Wang, Kai
    Cheng, Haitao
    Chen, Xi
    Mou, Xuanqin
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [26] Photon-counting detector computed tomography Paradigm shift in cardiac CT imaging
    Kravchenko, Dmitrij
    Layer, Yannik Christian
    Vecsey-Nagy, Milan
    Tremamunno, Giuseppe
    Varga-Szemes, Akos
    Schlett, Christopher L.
    Bamberg, Fabian
    Emrich, Tilman
    Hagar, Muhammad Taha
    RADIOLOGIE, 2024, 64 (12): : 928 - 934
  • [27] Development of a Multi-Energy Photon-Counting Readout ASIC for CZT Detectors
    Deng, Zhi
    Zhu, Xuezhou
    Chen, Yu
    Liu, Yinong
    Xing, Yuxiang
    Feng, Chuqing
    Wang, Guangqi
    Liu, Hui
    Wu, Sen
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [28] Quantitative material decomposition using spectral computed tomography with an energy-resolved photon-counting detector
    Lee, Seungwan
    Choi, Yu-Na
    Kim, Hee-Joung
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (18): : 5457 - 5482
  • [29] A review of applications of photon-counting computed tomography in head and neck imaging
    Dogra, Siddhant
    Shekhrajka, Nitesh
    Moonis, Gul
    BRITISH JOURNAL OF RADIOLOGY, 2024,
  • [30] Effects of Detector Sampling on Noise Reduction in Clinical Photon-Counting Whole-Body Computed Tomography
    Klein, Laura
    Dorn, Sabrina
    Amato, Carlo
    Heinze, Sarah
    Uhrig, Monika
    Schlemmer, Heinz-Peter
    Kachelriess, Marc
    Sawall, Stefan
    INVESTIGATIVE RADIOLOGY, 2020, 55 (02) : 111 - 119