New liquid nitrogen cryogenic modulator for comprehensive two-dimensional gas chromatography

被引:76
|
作者
Harynuk, J [1 ]
Górecki, T [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
关键词
gas chromatography; comprehensive two-dimensional; cryogenic modulators; modulators;
D O I
10.1016/j.chroma.2003.08.097
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new liquid nitrogen (LN2)jet-based thermal modulator for performing comprehensive two-dimensional (2D) gas chromatographic (GC x GC) separations has been designed and constructed. Temperature measurements of the trapping zone, a segment of uncoated fused silica capillary, show that it can be cooled to - 196degreesC in about 300 ms. A film of liquid nitrogen develops on the outside of the trapping capillary even when the oven temperature is in excess of 200degreesC. Compounds as volatile as propane can be trapped by the modulator and held for periods of at least 1 min without breakthrough. The peak widths for n-alkanes are on the order of 80 ms at half height after passing through an 80 cm second dimension column. Repeated analysis of gasoline demonstrated excellent run-to-run reproducibility of the system. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:53 / 63
页数:11
相关论文
共 50 条
  • [21] Comprehensive two-dimensional gas chromatography
    Hinshaw, JV
    LC GC EUROPE, 2004, 17 (02) : 86 - +
  • [22] Comprehensive two-dimensional gas chromatography
    Hinshaw, JV
    LC GC NORTH AMERICA, 2004, 22 (01) : 32 - +
  • [23] Comprehensive two-dimensional liquid chromatography
    Shellie, Robert A.
    Haddad, Paul R.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 386 (03) : 405 - 415
  • [24] Evaluation of a Microfabricated Thermal Modulator for Comprehensive Two-Dimensional Microscale Gas Chromatography
    Kim, Sung-Jin
    Serrano, Gustavo
    Wise, Kensall D.
    Kurabayashi, Katsuo
    Zellers, Edward T.
    ANALYTICAL CHEMISTRY, 2011, 83 (14) : 5556 - 5562
  • [25] Comprehensive two-dimensional gas chromatography
    Ryan, D
    Marriott, P
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 376 (03) : 295 - 297
  • [26] Comprehensive two-dimensional gas chromatography
    Danielle Ryan
    Philip Marriott
    Analytical and Bioanalytical Chemistry, 2003, 376 : 295 - 297
  • [27] Thermal modulator array for analyte modulation and comprehensive two-dimensional gas chromatography
    Burger, BV
    Snyman, T
    Burger, WJG
    van Rooyen, WF
    JOURNAL OF SEPARATION SCIENCE, 2003, 26 (1-2) : 123 - 128
  • [28] Single-jet, single-stage cryogenic modulator for comprehensive two-dimensional gas chromatography (GC x GC)
    Adahchour, M
    Beens, J
    Brinkman, UAT
    ANALYST, 2003, 128 (03) : 213 - 216
  • [29] Comprehensive two-dimensional gas chromatography using liquid nitrogen modulation: set-up and applications
    Pursch, M
    Eckerle, P
    Biel, J
    Streck, R
    Cortes, H
    Sun, K
    Winniford, B
    JOURNAL OF CHROMATOGRAPHY A, 2003, 1019 (1-2) : 43 - 51
  • [30] Chemometric analysis of comprehensive two-dimensional gas chromatography data using cryogenic modulation
    Xie, LL
    Marriott, PJ
    Adams, M
    ANALYTICA CHIMICA ACTA, 2003, 500 (1-2) : 211 - 222