Field deployable sensor;
Ion mobility spectrometry;
Atmospheric pressure ionization;
Effect of humidity;
MEMBRANE INLET;
PRESSURE;
TEMPERATURE;
CHROMATOGRAPHY;
RESOLUTION;
ANILINES;
D O I:
10.1016/j.snb.2015.04.102
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
Vaporized water molecules are unavoidable present in the ambient atmosphere and can considerably affect the analytical signals obtained with sensor techniques during real-time field measurements. Also the introduction of humidity into stand-alone ion mobility spectrometers cannot completely be avoided and the water vapor is transported via the carrier gas into spectrometer. For this case, we systematically investigated the way in which ion mobility measurements in positive mode are influenced. The drift times obtained with humid carrier gas are the same as those detected under dry conditions. They are constant over a range of water concentrations up to 2000 ppm. The nature of ions formed during the atmospheric chemical ionization processes is therefore independent on the humidity. However, the relative abundance of product ions can strongly be affected depending on the properties of substances investigated. While the signal intensity obtained for amines is comparatively unaffected by humidity, a significant decrease is observable for toluenes, chlorinated compounds and ketones. (C) 2015 Elsevier B.V. All rights reserved.
机构:
NIST, Boulder, CO 80305 USANIST, Boulder, CO 80305 USA
Kirk, Ansgar T.
Zimmermann, Stefan
论文数: 0引用数: 0
h-index: 0
机构:
Univ Calif Berkeley, Berkeley Sensor & Actuator Ctr, Berkeley, CA 94720 USA
Dragerwerk AG & Co KGaA, Lubeck, Germany
Dragerwerk AG & Co KGaA, Abt Chem & Biochem Sensoren, Lubeck, GermanyNIST, Boulder, CO 80305 USA