Measurement of dose perturbation around shielded ovoids in high-dose-rate brachytherapy

被引:9
|
作者
Hira, Mukhraj [1 ]
Podgorsak, Matthew B. [1 ]
Jaggernauth, Wainwright [1 ]
Malhotra, Harish K. [1 ]
机构
[1] Roswell Pk Canc Inst, Dept Radiat Med, Buffalo, NY 14623 USA
关键词
High-dose-rate brachytherapy; Shielded Fletcher-Suit-Delclos applicator; MONTE-CARLO CALCULATIONS; FLETCHER-SUIT-DELCLOS; RADIOCHROMIC FILM; DOSIMETRY; RECOMMENDATIONS; DISTRIBUTIONS; APPLICATORS; BLADDER; RECTUM;
D O I
10.1016/j.brachy.2010.08.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: To analyze the effect of tungsten shields present in a Fletcher-Suit-Delclos ovoid by comparing the dose distribution computed by a treatment planning system (TPS) to the delivered dose distribution measured by radiochromic film dosimetry. METHODS AND MATERIALS: Gafchromic/EBT films were carefully wrapped around the caps (diameter 20-25 mm) of shielded as well as unshielded ovoids, including their anterior and posterior ends. The ovoids were irradiated to a dose of 300 cGy using a high dose rate remote afterloading unit. The films were scanned using Vidar VXR-16 Scanner. The dose distribution in the planes above, below, and on the sides of the ovoid were compared with the dose distribution computed by TPS, which does not account for the presence of shields. RESULTS: The dose distributions obtained about the unshielded ovoid from film dosimetry was in order of what is computed by TPS (90% measurements +/- 5%, maximum 8%). The dose reduction in the anterior part of the shielded ovoid affects maximally the dose to the bladder where a reduction up to 20% was noted. The reduction of dose in the posterior part of the ovoid, which is designed to shield rectum was as high as 23%. Where the shields are not present, insignificant difference in the measured and computed dose values was noticed. CONCLUSIONS: The TPS may substantially overestimate the dose to the bladder and rectum, including regions that lie in the shadow of the solid angle subtended by the shields if it does not account for the presence of tungsten shields. (C) 2011 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:232 / 241
页数:10
相关论文
共 50 条
  • [31] Biologic treatment planning for high-dose-rate brachytherapy
    Manning, MA
    Zwicker, RD
    Arthur, DW
    Arnfield, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03): : 839 - 845
  • [32] High-dose-rate brachytherapy - High-dose, high-tech, and high results
    Petereit, DG
    Fowler, JF
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1159 - 1161
  • [33] Anthropomorphic phantom to investigate the bladder dose in gynecological high-dose-rate brachytherapy
    Silva, R. M. V.
    Belinato, W.
    Macedo, L. E.
    Souza, D. N.
    BRACHYTHERAPY, 2015, 14 (05) : 633 - 641
  • [34] Monotherapeutic high-dose-rate brachytherapy for prostate cancer: A dose reduction trial
    Yoshioka, Yasuo
    Konishi, Koji
    Suzuki, Osamu
    Nakai, Yasutomo
    Isohashi, Fumiaki
    Seo, Yuji
    Otani, Yuki
    Koizumi, Masahiko
    Yoshida, Ken
    Yamazaki, Hideya
    Nonomura, Norio
    Ogawa, Kazuhiko
    RADIOTHERAPY AND ONCOLOGY, 2014, 110 (01) : 114 - 119
  • [35] Reporting small bowel dose in cervix cancer high-dose-rate brachytherapy
    Liao, Yixiang
    Dandekar, Virag
    Chu, James C. H.
    Turian, Julius
    Bernard, Damian
    Kiel, Krystyna
    MEDICAL DOSIMETRY, 2016, 41 (01) : 28 - 33
  • [36] Comparison of dose and catheter optimization algorithms in prostate high-dose-rate brachytherapy
    Poulin, Eric
    Varfalvy, Nicolas
    Aubin, Sylviane
    Beaulieu, Luc
    BRACHYTHERAPY, 2016, 15 (01) : 102 - 111
  • [37] Parameterised rectal dose and associations with latetoxicity in high-dose-rate prostate brachytherapy
    Hamlett, L.
    Wood, D. A.
    Aitkenhead, A.
    Hodgson, C.
    Wylie, J.
    Choudhury, A.
    Logue, J.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S88 - S88
  • [38] IMPACT OF SURFACE CURVATURE ON DOSE DELIVERY IN INTRAOPERATIVE HIGH-DOSE-RATE BRACHYTHERAPY
    Oh, Moonseong
    Wang, Zhou
    Malhotra, Harish K.
    Jaggernauth, Wainwright
    Podgorsak, Matthew B.
    MEDICAL DOSIMETRY, 2009, 34 (01) : 63 - 74
  • [39] The Impact of Tandem-and-Ovoids Displacements on High-dose-rate (HDR) Plans
    Anderson, J.
    Zhang, W.
    Kim, Y.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S434 - S435
  • [40] Assessment of Absolute Dose Rate in Tungsten-Shielded Applicators for High-Dose-Rate Brachytherapy: Comparing Results Between Advanced Dose Calculation Methodologies and Physical Measurements
    Tien, C. J.
    Chen, Z.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E501 - E501