Characterization of a GaAs photon-counting detector for mammography

被引:8
|
作者
Ghammraoui, Bahaa [1 ]
Gkoumas, Spyridon [2 ]
Glick, Stephen J. [1 ]
机构
[1] US FDA, Off Sci & Engn Labs, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
[2] DECTRIS Ltd, Baden, Switzerland
关键词
x-ray detectors; photon-counting detector; breast mammography; GaAs detector; SPECTRAL MAMMOGRAPHY; DIGITAL MAMMOGRAPHY; QUANTUM EFFICIENCY; RAY;
D O I
10.1117/1.JMI.8.3.033504
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate the potential of a prototype gallium arsenide (GaAs) photon-counting detector (PCD) for imaging of the breast. Approach: First, the contrast-to-noise ratio (CNR) using different aluminum/poly(methyl methacrylate) (PMMA) phantoms of different thicknesses were measured. Second, microcalcification detection accuracy using a receiver operating characteristic study with three observers reading an ensemble of images was measured. Finally, the feasibility of using a GaAs system with two energy bins for contrast-enhanced mammography was investigated. Results: For the first two studies, the GaAs detector was compared with a commercial mammography system. The CNR was estimated by imaging 18-, 36-, and 110-mu m-thick aluminum targets placed on top of 6 cm of PMMA plates and was found to be similar or better over a range of exposures. We observed a similar performance of detecting microcalcifications with the GaAs detector over a range of clinically applicable dose levels with a small increase at lower dose levels. The results also showed that contrast-enhanced spectral mammography using a GaAs PCD is feasible and beneficial. Conclusions: Results from this study suggest that performance with this new detector seems either slightly improved or equivalent to a commercial mammography system that used an energy-integrated detector. No attempt at optimizing exposure techniques for the GaAs detector was performed. Further research is needed to determine optimal acquisition parameters for the GaAs detector and to develop more sophisticated material decomposition algorithms that promise to provide improved quantitative estimates of iodine uptake. (C) 2021 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:14
相关论文
共 50 条
  • [21] Detection of microcalcifications and tumor tissue in mammography using a CdTe-series photon-counting detector
    Nakajima, Ai
    Ihori, Akiko
    Nishide, Hiroko
    Koyama, Shuji
    Yamakawa, Tsutomu
    Yamamoto, Shuichiro
    Okada, Masahiro
    Kodera, Yoshie
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [22] The technical development of photon-counting detector CT
    McCollough, Cynthia H.
    Rajendran, Kishore
    Leng, Shuai
    Yu, Lifeng
    Fletcher, Joel G.
    Stierstorfer, Karl
    Flohr, Thomas G.
    EUROPEAN RADIOLOGY, 2023, 33 (08) : 5321 - 5330
  • [23] Phasor imaging with a widefield photon-counting detector
    Colyer, Ryan A.
    Siegmund, Oswald H. W.
    Tremsin, Anton S.
    Vallerga, John V.
    Weiss, Shimon
    Michalet, Xavier
    JOURNAL OF BIOMEDICAL OPTICS, 2012, 17 (01)
  • [24] AJR Forum on Photon-Counting Detector CT
    Dane, Bari
    Amrhein, Timothy J.
    Kay, Fernando
    Schwartz, Fides R.
    Toia, Giuseppe V.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 223 (06)
  • [25] Musculoskeletal imaging with a prototype photon-counting detector
    M. Gruber
    P. Homolka
    M. Chmeissani
    M. Uffmann
    M. Pretterklieber
    F. Kainberger
    European Radiology, 2012, 22 : 205 - 210
  • [26] The Future Arrived Photon-Counting Detector CT
    Alkadhi, Hatem
    Runge, Val
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 439 - 440
  • [27] AUTOMATIC SPECTROSCOPIC PHOTON-COUNTING DETECTOR SYSTEM
    KUREPA, JM
    PANIC, BM
    LEVI, BJ
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1977, 32 (9-10) : 413 - 417
  • [28] Photon-Counting Spectral Phase-Contrast Mammography
    Fredenberg, E.
    Roessl, E.
    Koehler, T.
    van Stevendaal, U.
    Schulze-Wenck, I.
    Wieberneit, N.
    Stampanoni, M.
    Wang, Z.
    Kubik-Huch, R. A.
    Hauser, N.
    Lundqvist, M.
    Danielsson, M.
    Aslund, M.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [29] Clinical evaluation of a photon-counting tomosynthesis mammography system
    Maidment, Andrew D. A.
    Ullberg, Christer
    Francke, Tom
    Lindqvist, Lars
    Sokolov, Skiff
    Lindman, Karin
    Adelow, Leif
    Sunden, Per
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2006, 4046 : 144 - 151
  • [30] Advanced Neuroimaging With Photon-Counting Detector CT
    Abel, Frederik
    Schubert, Tilman
    Winklhofer, Sebastian
    INVESTIGATIVE RADIOLOGY, 2023, 58 (07) : 472 - 481