Strategies for the safe magnetic resonance imaging of pacemaker-dependent patients

被引:55
|
作者
Gimbel, JR
Bailey, SM
Tchou, PJ
Ruggieri, PM
Wilkoff, BL
机构
[1] Pk W Hosp, Knoxville, TN USA
[2] Cleveland Clin Fdn, Cleveland, OH 44195 USA
来源
关键词
magnetic resonance imaging; pacemaker; electromagnetic interference; safety;
D O I
10.1111/j.1540-8159.2005.00230.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To determine if strategies used to safely scan nonpacemaker dependent patients could be applied to facilitate safe MRI of pacemaker-dependent patients. Interventions: Ten pacemaker-dependent patients underwent a total of 11 MRI scans of the head and neck. Screening, reprogramming VOO or DOO at 60 ppm, and monitoring strategies were used to facilitate MRI. A transmit-receive coil was used and MRI pulse sequences were modified to limit the whole-body specific absorption rate (SAR). Results: All scans proceeded uneventfully. No difficulties in post-MRI telemetry or interrogation were seen and no post-MRI programming changes were noted. No patient experienced arrhythmia or symptoms during or immediately after MRI Battery status remained unchanged, No patient experienced post-MRI change in sensing thresholds. Three patients showed no change in the atrial or ventricular pacing thresholds when the pre-MRI values were compared to the immediate post-MRI values and the 3-month follow-up values. All other patients showed a rise or fall of 0.5 V in their chamber threshold values when the pre-MRI, post-MRI, and 3-month follow-up values were compared. More patients showed a fall in their pacing thresholds than a rise post-MRI. Conclusion: While clearly a higher risk group, like nonpacemaker-dependent patients, MRI might be performed in pacemaker-dependent patients if appropriate pacemaker reprogramming, patient monitoring, and MRI scanning techniques are implemented.
引用
收藏
页码:1041 / 1046
页数:6
相关论文
共 50 条
  • [41] Safety and Outcomes of Magnetic Resonance Imaging in Patients With Abandoned Pacemaker and Defibrillator Leads
    Higgins, John V.
    Gard, Joseph J.
    Espinosa, Raul E.
    Sheldon, Seth H.
    Asirvatham, Samuel I.
    Acker, Nancy
    Dalzell, Connie
    Cha, Yong-Mei
    Felmlee, Joel P.
    Wood, Christopher P.
    Watson, Robert E.
    Friedman, Paul A.
    CIRCULATION, 2013, 128 (22)
  • [42] ASSESSMENT OF DIAGNOSTIC NEED FOR MAGNETIC RESONANCE IMAGING IN MEDICARE PATIENTS WITH PACEMAKER IMPLANTS
    Eggleston, A.
    Lim, M.
    Veath, B. K.
    Mura, T.
    VALUE IN HEALTH, 2012, 15 (07) : A611 - A611
  • [43] MRI Mode Programming for Safe Magnetic Resonance Imaging in Patients With a Magnetic Resonance Conditional Cardiac Device
    Nakai, Toshiko
    Kurokawa, Sayaka
    Ikeya, Yukitoshi
    Iso, Kazuki
    Takahashi, Keiko
    Sasaki, Naoko
    Ashino, Sonoko
    Okubo, Kimie
    Okumura, Yasuo
    Kunimoto, Satoshi
    Watanabe, Ichiro
    Hirayama, Atsushi
    INTERNATIONAL HEART JOURNAL, 2016, 57 (02) : 173 - 176
  • [44] Preserved diagnostic utility of dobutamine stress echocardiography in pacemaker-dependent patients with absolute chronotropic incompetence
    Oral, H
    Armstrong, WF
    Bach, DS
    AMERICAN HEART JOURNAL, 1999, 138 (02) : 364 - 368
  • [45] Magnetic resonance imaging is safe in patients with left bundle branch pacing
    Bernardes-Souza, Breno
    Bhatt, Raj V.
    Bayoneta, Aileen Joy S.
    Seo, Jungsik
    Finn, John P.
    Mori, Shumpei
    Do, Duc H.
    Boyle, Noel G.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2023, 46 (07): : 717 - 720
  • [46] Magnetic Resonance Imaging in a Patient with a Dual Chamber Pacemaker
    Millar, Lynne Martina
    Robinson, Andrew George
    O'Flaherty, Maurice Thomas
    Eames, Niall
    Johnston, Nicola
    Heyburn, Gary
    CASE REPORTS IN MEDICINE, 2010, 2010
  • [47] Safety Aspects of Magnetic Resonance Imaging for Pacemaker Holders
    Bernardi, P.
    Cavagnaro, M.
    Pisa, S.
    Piuzzi, E.
    ICEAA: 2009 INTERNATIONAL CONFERENCE ON ELECTROMAGNETICS IN ADVANCED APPLICATIONS, VOLS 1 AND 2, 2009, : 869 - 872
  • [48] Pacemaker complication during magnetic resonance imaging - Reply
    Coman, JA
    Martin, ET
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (01) : 162 - 162
  • [49] NUCLEAR MAGNETIC-RESONANCE IMAGING IN A DDR PACEMAKER
    WELTER, DA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04): : 621 - 621
  • [50] NUCLEAR-MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A PACEMAKER
    INBAR, S
    LARSON, J
    BURT, T
    MAFEE, M
    EZRI, MD
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 1993, 305 (03): : 174 - 175