Biomechanical analysis of plate osteosynthesis systems for proximal humerus fractures

被引:31
|
作者
Lever, James P. [2 ]
Aksenov, Sergei A. [2 ]
Zdero, Rad [1 ]
Ahn, Henry [2 ]
Mckee, Michael D. [2 ]
Schemitsch, Emil H. [1 ,2 ]
机构
[1] St Michaels Hosp, Martin Orthopaed Biomech Lab, Toronto, ON M5B 1W8, Canada
[2] Univ Toronto, Dept Surg, Toronto, ON, Canada
关键词
biomechanics; proximal humerus; fracture; plate fixation;
D O I
10.1097/BOT.0b013e31815c89ce
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To biomechanically assess five proximal, humeral, fracture-plate-fixation systems. Methods: Surgical neck fractures, with and without cortical contact, were created in 25 fresh-frozen cadaveric humeri. Five methods of plate fixation were used for repair: construct A [an eight-hole, low contact dynamic compression (LCDC) plate contoured into a blade shape, supported by one, 70-mm-long, 4.5-mm-diameter cortical screw acting as a truss], construct B (a 10-hole LCDC plate arrangement identical to construct A, but using one, 70-mm-long, 3.5-mm-diameter cortical screw as a truss), construct C [a five-hole dynamic compression (DC) blade plate with one, 6.5-mm-diameter cancellous screw], construct D (a five-hole T-plate supported by three, 6.5-mm-diameter cancellous screws), and construct E (a five-hole cloverleaf plate supported by five, 4-mm-diameter cancellous screws). Plates were posterior to the bicipital groove, 10 mm distal to the greater tuberosity tip, on the lateral aspect of the humeral shaft. Screw fixation was done using standard AO compression plating techniques. Stiffness of constructs was measured in bending and axial compression. Locked plates were not assessed. Results: For cortical contact [abduction of 20 degrees (P = 0.02), flexion of 20 degrees (P = 0.02), flexion of 90 degrees (P = 0.005)] and no cortical contact [flexion of 90 degrees (P = 0.0001)], construct A was significantly stiffer than other constructs. For no cortical contact in abduction of 90 degrees (P = 0.05), construct A was significantly stiffer than other constructs. Conclusions: Construct A was significantly stiffer than other constructs.
引用
收藏
页码:23 / 29
页数:7
相关论文
共 50 条
  • [41] Biomechanical Assessment of Three Osteosynthesis Constructs by Periprosthetic Humerus Fractures
    Harb, Afif
    Welke, Bastian
    Liodakis, Emmanouil
    Razaeian, Sam
    Zhang, Dafang
    Krettek, Christian
    Hurschler, Christof
    Hawi, Nael
    ADVANCES IN ORTHOPEDICS, 2020, 2020
  • [42] Biomechanical study of different plate configurations for distal humerus osteosynthesis
    Bogataj, M.
    Kosel, F.
    Norris, R.
    Krkovic, M.
    Brojan, M.
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2015, 53 (05) : 381 - 392
  • [43] Biomechanical study of different plate configurations for distal humerus osteosynthesis
    M. Bogataj
    F. Kosel
    R. Norris
    M. Krkovic
    M. Brojan
    Medical & Biological Engineering & Computing, 2015, 53 : 381 - 392
  • [44] Proximal humerus fractures: a comparative biomechanical analysis of intra and extramedullary implants
    B. Füchtmeier
    R. May
    R. Hente
    M. Maghsudi
    M. Völk
    J. Hammer
    M. Nerlich
    L. Prantl
    Archives of Orthopaedic and Trauma Surgery, 2007, 127 : 441 - 447
  • [45] Proximal humerus fractures:: a comparative biomechanical analysis of intra and extramedullary implants
    Fuechtmeier, B.
    May, R.
    Hente, R.
    Maghsudi, M.
    Voelk, M.
    Hammer, J.
    Nerlich, M.
    Prantl, L.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2007, 127 (06) : 441 - 447
  • [46] Biomechanical analysis of a novel locking plate with smooth pegs versus a conventional locking plate with threaded screws for proximal humerus fractures
    Yamamoto, Nobuyuki
    Hongo, Michio
    Berglund, Lawrence J.
    Sperling, John W.
    Cofield, Robert H.
    An, Kai-Nan
    Steinmann, Scott P.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2013, 22 (04) : 445 - 450
  • [47] Osteosynthesis by locking plate for proximal tibial fractures
    Favreau, Henri
    Tamir, Mekki
    Adam, Philippe
    Ollivier, Matthieu
    Bonnomet, Francois
    Ehlinger, Matthieu
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2024, 55
  • [48] Double Plate Osteosynthesis of proximal Ulna Fractures
    Leschinger, T.
    Hackl, M.
    Mueller, L. P.
    Wegmann, K.
    UNFALLCHIRURG, 2018, 121 (11): : 923 - 929
  • [49] To Detect the Outcome of Proximal Humerus Fractures Treated With a Locking Proximal Humerus Plate
    Ahmed, Nasir
    Khan, Tarique Ali
    Anwar, Muhammad Absar
    Ali, Ahmed
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2015, 9 (01): : 281 - 285
  • [50] NEW OSTEOSYNTHESIS, SPIRAL PIN NAILING, FOR FRACTURES OF THE PROXIMAL HUMERUS
    YANO, S
    ISCHIDA, F
    HIROHATA, K
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1984, 122 (04): : 527 - 527