共 50 条
Feasibility of online seizure detection with continuous EEG in the intensive care unit
被引:10
|作者:
Ponten, S. C.
[1
,2
]
Ronner, H. E.
[1
]
Strijers, R. L. M.
[1
]
Visser, M. C.
[2
]
Peerdeman, S. M.
[3
]
Vandertop, W. P.
[3
]
Beishuizen, A.
[4
]
Girbes, A. R. J.
[4
]
Stam, C. J.
[1
]
机构:
[1] Vrije Univ Amsterdam Med Ctr, Dept Clin Neurophysiol, NL-1007 MB Amsterdam, Netherlands
[2] Vrije Univ Amsterdam Med Ctr, Dept Neurol, NL-1007 MB Amsterdam, Netherlands
[3] Vrije Univ Amsterdam Med Ctr, Dept Neurosurg, NL-1007 MB Amsterdam, Netherlands
[4] Vrije Univ Amsterdam Med Ctr, Dept Intens Care, NL-1007 MB Amsterdam, Netherlands
来源:
关键词:
Coma;
Continuous EEG;
Epilepsy monitoring;
Intensive care unit;
NONCONVULSIVE STATUS EPILEPTICUS;
CRITICALLY-ILL PATIENTS;
TRAUMATIC BRAIN-INJURY;
SYNCHRONIZATION LIKELIHOOD;
COMATOSE;
ICU;
ADULT;
D O I:
10.1016/j.seizure.2010.09.007
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
Introduction: Continuous EEG (cEEG) is of great interest in view of the reported high prevalence of non-convulsive seizures on intensive care units (ICUs). Here, we describe our experiences applying a seizure warning system using cEEG monitoring. Methods: Fifty comatose ICU patients were included prospectively and monitored. Twenty-eight patients had post-anoxic encephalopathy (PAE) and 22 had focal brain lesions. A measure of neuronal interactions, synchronization likelihood, was calculated online over 10 s EEG epochs and instances when the synchronization likelihood exceeded a threshold where marked as seizures. Results: Five patients developed seizures. Our method detected seizures in three patients, in the other patients seizures were missed because of they were non-convulsive and had a focal character. The average false positive rate was 0.676/h. Discussion: This is our first attempt to implement online seizure detection in the ICU. Despite problems with artifacts and that we missed focally oriented seizures, we succeeded in monitoring patients online. Given the relatively high occurrence of seizures, online seizure detection with cEEG merits further development for use in ICUs. (C) 2010 British Epilepsy Association. Published by Elsevier Ltd. All rights reserved.