Noninvasive measurement of blood hemoglobin concentration by time-resolved transmission spectroscopy

被引:2
|
作者
Sun, M. X. [1 ,2 ]
Zhang, L. Y. [3 ]
Jiang, C. Y. [1 ,2 ]
Yuan, Y. [1 ,2 ]
Yang, J. C. [1 ,2 ]
Tian, L. [1 ,2 ]
Wu, J. P. [1 ,2 ]
Li, Y. X. [1 ,2 ]
机构
[1] Chinese Acad Med Sci, Inst Biomed Engn, Laser Med Lab, Tianjin 300192, Peoples R China
[2] Peking Union Med Coll, Tianjin 300192, Peoples R China
[3] Mil Transportat Coll, Basic Dept, Tianjin 300161, Peoples R China
来源
OPTIK | 2015年 / 126卷 / 04期
关键词
Blood hemoglobin concentration; Time-resolved transmission spectroscopy; Noninvasive measurement; Laplace transform; LIGHT-SCATTERING; REFLECTANCE;
D O I
10.1016/j.ijleo.2014.07.148
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
An optical spectroscopic method is investigated theoretically for noninvasive and continuous monitoring of blood hemoglobin levels. The method may provide accurate measurement of hemoglobin concentration by detection and analysis of dynamic dual wavelength time-resolved transmission under a condition of artificial blood flow kinetics in a human finger. This condition can assist light scattering and leads to a better signal-to-noise ratio necessary to accurate measurement of transmission. The modified parametric slopes were derived from the Laplace transformed data of the time-resolved transmittance. Continuous hemoglobin concentration was measured in the range from 6 to 16 g/dl and the results show that the sensitivity of hemoglobin sensing is enhanced when we use selected values of the Laplace transform parameter p=5 x 10(10) s(-1) comparing with the continuous wave situation. (C) 2014 Elsevier GmbH. All rights reserved.
引用
收藏
页码:460 / 463
页数:4
相关论文
共 50 条
  • [1] NONINVASIVE IN-VIVO MEASUREMENT OF BLOOD SPECTRUM BY TIME-RESOLVED NEAR-INFRARED SPECTROSCOPY
    NAHM, W
    GEHRING, H
    SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) : 174 - 179
  • [2] INFLUENCE OF BLOOD-VESSELS ON THE MEASUREMENT OF HEMOGLOBIN OXYGENATION AS DETERMINED BY TIME-RESOLVED REFLECTANCE SPECTROSCOPY
    LIU, H
    CHANCE, B
    HIELSCHER, AH
    JACQUES, SL
    TITTEL, FK
    MEDICAL PHYSICS, 1995, 22 (08) : 1209 - 1217
  • [3] Measurement of porcine luteinizing hormone concentration in blood by time-resolved fluoroimmunoassay
    Noguchi, Michiko
    Yoshioka, Koji
    Kaneko, Hiroyuki
    Iwamura, Shokichi
    Takahashi, Toru
    Suzuki, Chie
    Arai, Sachiko
    Wada, Yasunori
    Itoh, Seigo
    JOURNAL OF VETERINARY MEDICAL SCIENCE, 2007, 69 (12): : 1291 - 1294
  • [4] Non-invasive measurement of blood glucose level by time-resolved transmission spectroscopy: A feasibility study
    Sun, Meixiu
    Chen, Nanguang
    OPTICS COMMUNICATIONS, 2012, 285 (06) : 1608 - 1612
  • [5] Time-resolved terahertz transmission spectroscopy of dielectrics
    Kuzel, P
    Petzelt, J
    FERROELECTRICS, 2000, 239 (1-4) : 949 - 956
  • [6] Transmission Spectroscopy in Noninvasive Blood Glucose Measurement
    Pozhar K.V.
    Biomedical Engineering, 2018, 52 (4) : 247 - 250
  • [7] MEASUREMENT MODELS FOR TIME-RESOLVED SPECTROSCOPY - A COMMENT
    HOFMANN, HF
    MAHLER, G
    QUANTUM AND SEMICLASSICAL OPTICS, 1995, 7 (04): : 489 - 497
  • [8] Noninvasive Measurement of Hemoglobin Concentration
    Opp, A.
    Kraitl, J.
    Gehring, H.
    ANESTHESIA AND ANALGESIA, 2007, 105 : S112 - S112
  • [9] TIME-RESOLVED SPECTROSCOPY OF HEMOGLOBIN AND MYOGLOBIN IN RESTING AND ISCHEMIC MUSCLE
    CHANCE, B
    NIOKA, S
    KENT, J
    MCCULLY, K
    FOUNTAIN, M
    GREENFELD, R
    HOLTOM, G
    ANALYTICAL BIOCHEMISTRY, 1988, 174 (02) : 698 - 707
  • [10] Simultaneous measurement of cerebral hemoglobin oxygen saturation and blood volume in asphyxiated neonates by near-infrared time-resolved spectroscopy
    Nakamura, Shinji
    Koyano, Kosuke
    Jinnai, Wataru
    Hamano, Satoshi
    Yasuda, Saneyuki
    Konishi, Yukihiko
    Kuboi, Toru
    Kanenishi, Kenji
    Nishida, Tomoko
    Kusaka, Takashi