Original vs. non-original abutments for screw-retained single implant crowns: An in vitro evaluation of internal fit, mechanical behaviour and screw loosening

被引:38
|
作者
Alonso-Perez, Raquel [1 ]
Bartolome, Jose F. [2 ]
Ferreiroa, Alberto [1 ]
Salido, Maria P. [1 ]
Pradies, Guillermo [1 ]
机构
[1] Univ Complutense Madrid, Dept Buccofacial Prosthesis, Fac Odontol, Madrid, Spain
[2] CSIC, ICMM, Madrid, Spain
关键词
biomaterials; material sciences; prosthodontics; FRACTURE-RESISTANCE; MARGINAL ACCURACY; ROTATIONAL MISFIT; MICROBIAL LEAKAGE; TORQUE VALUES; INTERFACE; MICROGAP; SURVIVAL; FATIGUE; CONNECTIONS;
D O I
10.1111/clr.13390
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives To assess the internal accuracy, mechanical behaviour under static load and screw loosening before and after cyclic loading of implant-supported crowns restored with original components or with two compatible non-original brands. Materials and Methods Sixty-three dental implants were divided into three groups (n = 21 each): Group 1 used original components, and groups 2 and 3 used non-original components. Internal accuracy was measured throughout a cross section of the sample groups (n = 8) using scanning electron microscopy (SEM) to evaluate the fit for implant-abutments, implant-crowns and crown-abutments. To evaluate load-bearing capacity, eight samples from each group were loaded until fracture according to ISO Norm 14801. The removal torque value (RTV) was evaluated in the screws connecting the crown with the abutment and the abutment with the implant before and after cyclic loading (2 x 10(6) cycles) in five samples from each group. The data were analysed using the Kruskal-Wallis test (p < 0.05). Results Statistically significant differences were found among the mean crown-abutment gaps measured in Group 1 (9.3 mu m), Group 2 (45.4 mu m) and Group 3 (44.9 mu m). Higher values for mean load-bearing capacity were found in Group 1 (1,098 N) than in groups 2 and 3 (1,057 and 973 N, respectively); however, these differences were not statistically significant. Original abutment-implant screws exhibited lower percentages of torque reduction than the non-originals. Conclusions Enhanced fit is expected when original components are used. The original abutments exhibited lower percentages of torque reduction after cyclic loading than non-originals.
引用
收藏
页码:1230 / 1238
页数:9
相关论文
共 19 条
  • [1] Screw loosening of original and non-original abutments in implant dentistry: an in vitro study
    Pournasiri, I
    Farid, F.
    Jafari, H. Zaker
    Simdar, N.
    Maleki, D.
    JOURNAL OF OSSEOINTEGRATION, 2022, 14 (03) : 155 - 158
  • [2] Evaluation of Prosthodontic Complications in Screw-Retained and Cement-Retained Crowns in Dental Implants: An Original Research
    Rajput, Mansi
    Kohli, Arshdeep S.
    Abdul, Hina N.
    Sandhu, Sunpreet K.
    Thakkar, Radhika
    Karre, Shivani
    JOURNAL OF PHARMACY AND BIOALLIED SCIENCES, 2024, 16 : S609 - S612
  • [3] Effect of abutment finish lines on the mechanical behavior and marginal fit of screw-retained implant crowns: An in vitro study
    Garcia-Gonzalez, Marta
    Gonzalez-Gonzalez, Ignacio
    Garcia-Garcia, Ismael
    Blason-Gonzalez, Sergio
    Lamela-Rey, Maria Jesus
    Fernandez-Canteli, Alfonso
    Alvarez-Arenal, Angel
    JOURNAL OF PROSTHETIC DENTISTRY, 2022, 127 (02): : 318.e1 - 318.e10
  • [4] Marginal fit of cemented and screw-retained crowns incorporated on the Straumann (ITI)® Dental Implant System: an in vitro study
    Tosches, N. A.
    Braegger, U.
    Lang, N. P.
    CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (01) : 79 - 86
  • [5] Fracture Strength of Implant-Supported Ceramic Crowns with Customized Zirconia Abutments: Screw Retained vs. Cement Retained
    Lima Verde Nogueira, Lorenna Bastos
    Vilarinho Soares Moura, Carmem Dolores
    Francischone, Carlos Eduardo
    Valente, Valdimar Silva
    Mendes Alencar, Suya Moura
    Moura, Walter Leal
    Soares Martins, Gregorio Antonio
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2016, 25 (01): : 49 - 53
  • [6] Digital vs Conventional Workflow for Screw-Retained Single-Implant Crowns: A Comparison of Key Considerations
    Di Fiore, Adolfo
    Vigolo, Paolo
    Graiff, Lorenzo
    Stellini, Edoardo
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2018, 31 (06) : 577 - 579
  • [7] Accuracy of Original vs. Non-Original Abutments Using Various Connection Geometries for Single Unit Restorations: A Systematic Review
    Rizvi, Naureen
    Alyahya, Yasir
    Rizvi, Ali
    Narvekar, Unnati
    Petridis, Haralampos
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2022, 31 (07): : E21 - E52
  • [8] Intrasubject comparison of digital vs. conventional workflow for screw-retained single-implant crowns: Prosthodontic and patient-centered outcomes
    Delize, Vincent
    Bouhy, Alice
    Lambert, France
    Lamy, Marc
    CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (09) : 892 - 902
  • [9] Evaluation of debonding force of screw-retained lithium disilicate implant-supported crowns cemented to abutments of different designs and surface treatments
    Khalifa, Ahmed Abdel Latif
    Metwally, Nayrouz Adel
    Khamis, Mohamed Moataz
    JOURNAL OF PROSTHETIC DENTISTRY, 2025, 133 (01): : 215 - 220
  • [10] Peri-implant and esthetic outcomes of cemented and screw-retained crowns using zirconia abutments in single implant-supported restorations-A systematic review and meta-analysis
    Dini, Caroline
    Borges, Guilherme Almeida
    Costa, Raphael Cavalcante
    Magno, Marcela Barauna
    Maia, Lucianne Cople
    Ricardo Barao, Valentim Adelino
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (10) : 1143 - 1158