Fine particle mass from the Diskus inhaler and Turbuhaler inhaler in children with asthma

被引:77
|
作者
Bisgaard, H [1 ]
Klug, B
Sumby, BS
Burnell, PKP
机构
[1] Rigshosp, Natl Univ Hosp, Dept Paediat, Pulm Serv, DK-2100 Copenhagen, Denmark
[2] Glaxo Wellcome Res & Dev Ltd, Ware, Herts, England
关键词
aerosol; children; Diskus; dry powder inhaler; inspiratory flow; Turbuhaler;
D O I
10.1183/09031936.98.11051111
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
The study aimed to investigate dose consistency and particle distribution from the dry powder inhalers Diskus and Turbuhaler. Full profiles of inhalation pressure versus time were recorded in 18 4 gr old and 18 8 yr old asthmatic children through Diskus and Turbuhaler inhalers, These data mere used in an inhalation profile simulator to assess drug delivery from both a Diskus inhaler and a Turbuhaler inhaler, and in particular to assess the proportion of drug emitted in the coarse (>4.7 mu m) and fine (<4.7 mu m) particle size range from each type of inhaler. The inhalation profile more accurately represents the changes in flow rate over time through the device than the constant Bow rate usually applied with an impactor alone. The aerosol cloud was released before the peak inspiratory effort had been achieved and accordingly the early part and not the peak of the inspiratory performance is a determinant of the quality of the aerosol. The mean (SD) amount of drug in large particles (>4.7 mu m), fine particles (<4.7 mu m) and very fine particles (<2.1 mu m) in percentage of label claim from the Fluticasone Diskus was 72 (5), 15 (2) and 2 (1) from the 4 Sr old children and 71 (3), 18 (2) and 2 (1) from the 8 yr old children, respectively. Similar particle fractions from the Budesonide Turbuhaler were 35 (9), 21 (10) and 7 (5) from 4 yr old children and 30 (7), 32 (9) and 12 (6) from 8 yr old children. In conclusion, the Diskus inhaler provides an improved dose consistency through the varying age groups and inspiratory flow performances when compared to the Turbuhaler in terms of the proportion of the dose emitted at each particle size, This improvement is at the expense of a low fine particle mass and a high proportion of coarse particles from the Diskus as compared with the Turbuhaler.
引用
收藏
页码:1111 / 1115
页数:5
相关论文
共 50 条
  • [21] Inhaler Technique in Children With Asthma: A Systematic Review
    Gillette, Chris
    Rockich-Winston, Nicole
    Kuhn, JoBeth A.
    Flesher, Susan
    Shepherd, Meagan
    ACADEMIC PEDIATRICS, 2016, 16 (07) : 605 - 615
  • [22] Efficacy and Safety of Salmeterol/Fluticasone Propionate Combination Delivered by the Diskus™ or Pressurised Metered-Dose Inhaler in Children with Asthma
    Teresa Bracamonte
    Uwe Schauer
    Andrzej Emeryk
    Alex Godwood
    Sima Balsara
    Clinical Drug Investigation, 2005, 25 : 1 - 11
  • [23] ASSESSMENT OF THE ABILITY OF YOUNG-CHILDREN TO USE A POWDER INHALER DEVICE (TURBUHALER)
    GOREN, A
    NOVISKI, N
    AVITAL, A
    MAAYAN, C
    STAHL, E
    GODFREY, S
    SPRINGER, C
    PEDIATRIC PULMONOLOGY, 1994, 18 (02) : 77 - 80
  • [24] Efficacy and safety of salmeterol/fluticasone propionate combination delivered by the Diskus™or pressurised metered-dose inhaler in children with asthma
    Bracamonte, T
    Schauer, U
    Emeryk, A
    Godwood, A
    Balsara, S
    CLINICAL DRUG INVESTIGATION, 2005, 25 (01) : 1 - 11
  • [25] Peak inspiratory flow rates generated through the Novolizer® and the Turbuhaler® dry powder inhaler devices by children with stable asthma
    von Berg, Andrea
    Kremer, Hans-Joachim
    Ellers-Lenz, Barbara
    Conrad, Frank
    Erb, Katharina
    Maus, Joachim
    Hermann, Robert
    JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2007, 20 (01): : 50 - 58
  • [26] A nurse-led intervention study: Promoting compliance with Diskus Inhaler use in asthma patients
    Mac Hale, Elaine
    Costello, Richard W.
    Cowman, Seamus
    NURSING OPEN, 2014, 1 (01): : 42 - 52
  • [27] TERBUTALINE BY METERED-DOSE INHALER - CONVENTIONAL INHALER VERSUS TUBE SPACER FOR CHILDREN WITH ASTHMA
    BECKER, AB
    SIMONS, FER
    BENOIT, TC
    GILLESPIE, CA
    ANNALS OF ALLERGY, 1985, 55 (05): : 724 - 728
  • [28] Effect of Flow Rate on Emitted Fine Particle Mass (FPM) from a Dry Powder Inhaler (DPI) and a Metered Dose Inhaler (MDI) with Spacer Delivering Salbutamol.
    Nagel, Mark
    Suggett, Jason
    Ali, Rubina
    Doyle, Cathy
    EUROPEAN RESPIRATORY JOURNAL, 2024, 64
  • [29] Albuterol via Turbuhaler versus albuterol via pressurized metered-dose inhaler in asthma
    Chapman, KR
    Friberg, K
    Balter, MS
    Hyland, RH
    Alexander, M
    Abboud, RT
    Peters, S
    Jennings, BH
    ANNALS OF ALLERGY ASTHMA & IMMUNOLOGY, 1997, 78 (01) : 59 - 63
  • [30] Impact of Education on Inhaler Technique in Hospitalized Children with Asthma
    Wei, H.
    Press, V. G.
    Siros, M.
    Volerman, A.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2020, 201