Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients

被引:14
|
作者
Asawaworarit, Rosepon [1 ]
Satitpitakul, Vannarut [1 ]
Taweekitikul, Parichart [1 ]
Pongpirui, Krit [2 ]
机构
[1] Chulalongkorn Univ, King Chulalongkorn Mem Hosp, Excellence Ctr Cornea & Limbal Stem Cell Transpla, Fac Med,Dept Ophthalmol, Bangkok, Thailand
[2] Chulalongkorn Univ, Fac Med, Bangkok, Thailand
来源
PLOS ONE | 2022年 / 17卷 / 05期
关键词
REPEATABILITY; ASTIGMATISM; EYES; RELIABILITY; KERATOMETRY;
D O I
10.1371/journal.pone.0268856
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients. Methods This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated. Results This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and keratoconic patients (ICC = 0.852) with mean differences of more than 1.0 diopter is clinically significant. Wider 95% limit of agreement were found in keratoconic eyes. Most of other parameters showed good to excellent agreement except WTW which showed poor to moderate agreement in both groups. IOL power showed clinically significant differences in patients with keratoconus. Conclusions The agreement of TCP measured by three devices was excellent in normal and keratoconic patients. However, TCP cannot be used interchangeably between devices.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Agreement Between Swept-source Optical Coherence Tomography and Rotating Scheimpflug Camera in Measurement of Corneal Parameters in Normal and Keratoconic Eyes
    Ghassemi, Hamed
    Zarei-Ghanavati, Mehran
    Khastavan, Mina
    Atighehchian, Mehrnaz
    Khorasani, Abbas Azimi
    Latifi, Golshan
    JOURNAL OF OPHTHALMIC & VISION RESEARCH, 2023, 18 (04) : 386 - 395
  • [2] Corneal astigmatism agreement between a swept-source ocular coherence tomography and Scheimpflug-Placido based optical biometers
    Mclintock, Cameron
    Uprety, Samir
    Mckelvie, James
    INTERNATIONAL OPHTHALMOLOGY, 2025, 45 (01)
  • [3] The evaluation of corneal tomography using swept-source optical coherence tomography and Scheimpflug imaging in healthy and ectasia eyes
    Herber, Robert
    Pillunat, Lutz E.
    Raiskup, Frederik
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2023, 64 (08)
  • [4] Scheimpflug Camera and Swept-Source Optical Coherence Tomography in Pachymetry Evaluation of Diabetic Patients
    Krysik, Katarzyna
    Dobrowolski, Dariusz
    Stanienda-Sokol, Karolina
    Wylegala, Edward A.
    Lyssek-Boron, Anita
    JOURNAL OF OPHTHALMOLOGY, 2019, 2019
  • [5] Comparison of Scheimpflug and swept-source anterior segment optical coherence tomography in normal and keratoconus eyes
    Ghoreishi, Seyed Mohammad
    Mortazavi, Seyed Ali Akbar
    Abtahi, Zahra-Alsadat
    Abtahi, Mohammad-Ali
    Sonbolestan, Seyed Ali
    Abtahi, Seyed-Hossein
    Mohammadinia, Mohadeseh
    Isfahani, Kimia Nekouie
    INTERNATIONAL OPHTHALMOLOGY, 2017, 37 (04) : 965 - 971
  • [6] Comparison of Scheimpflug and swept-source anterior segment optical coherence tomography in normal and keratoconus eyes
    Seyed Mohammad Ghoreishi
    Seyed Ali Akbar Mortazavi
    Zahra-Alsadat Abtahi
    Mohammad-Ali Abtahi
    Seyed Ali Sonbolestan
    Seyed-Hossein Abtahi
    Mohadeseh Mohammadinia
    Kimia Nekouie Isfahani
    International Ophthalmology, 2017, 37 : 965 - 971
  • [7] Agreement in anterior segment measurements between swept-source optical coherence and dual Scheimpflug tomography devices in keratoconus eyes
    Lee, Yunjin
    Oh, Joo Youn
    Choi, Hyuk Jin
    Kim, Mee Kum
    Yoon, Chang Ho
    SCIENTIFIC REPORTS, 2024, 14 (01):
  • [8] Corneal thickness and elevation measurements using swept-source optical coherence tomography and slit scanning topography in normal and keratoconic eyes
    Jhanji, Vishal
    Yang, Bingzhi
    Yu, Marco
    Ye, Cong
    Leung, Christopher K. S.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 41 (08): : 735 - 745
  • [9] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Herber, Robert
    Lenk, Janine
    Pillunat, Lutz E.
    Raiskup, Frederik
    EYE AND VISION, 2022, 9 (01)
  • [10] Comparison of corneal tomography using a novel swept-source optical coherence tomographer and rotating Scheimpflug system in normal and keratoconus eyes: repeatability and agreement analysis
    Robert Herber
    Janine Lenk
    Lutz E. Pillunat
    Frederik Raiskup
    Eye and Vision, 9