Feasibility of dual-chamber (DDD) pacing via a single-pass (VDD) pacing lead employing a floating atrial ring (dipole): Case series, future considerations, and refinements

被引:1
|
作者
Kassotis, J
Voigt, L
Mongwa, M
Reddy, CVR
机构
[1] New York Presbyterian Hosp Syst, New York Methodist Hosp, Clin Electrophysiol Serv, Div Cardiol, Brooklyn, NY 11215 USA
[2] New York Presbyterian Hosp Syst, New York Methodist Hosp, Dept Med, Brooklyn, NY 11215 USA
关键词
D O I
10.1177/000331970505600313
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
The objective of this study was to assess the feasibility of DDD pacing from a standard single-pass VDD pacemaker system. Over the past 2 decades significant advances have been made in the development of single-pass VDD pacing systems. These have been shown in long-term prospective studies to effectively preserve atrioventricular (AV) synchrony in patients with AV block and normal sinus node function. What remains problematic is the development of a single-pass pacing system capable of DDD pacing. Such a lead configuration would be useful in those patients with peripheral venous anomalies and in younger patients with congenital anomalies, which may require lead revisions in the future. In addition, with the increased use of resynchronization (biventricular pacing) therapy, the availability of a reliable single-pass lead will minimize operative time, enhance patient safety, and minimize the amount of hardware within the heart. The feasibility of DDD pacing via a Medtronic Capsure VDD-2 (Model #5038) pacing lead was evaluated. Twenty patients who presented with AV block and normal sinus node function were recruited for this study. Atrial pacing thresholds and sensitivities were assessed intraoperatively in the supine position with various respiratory maneuvers. Five patients who agreed to participate in long-term follow-up received a dual-chamber generator and were evaluated periodically over a 12-month period. Mean atrial sensitivity was 2.35 +/- 0.83 mV at the time of implantation. Effective atrial stimulation was possible in all patients at the time of implantation (mean stimulation threshold 3.08 +/- 1.04 V @ 0.5 ms [bipolar], 3.34 +/- 0.95 V @ 0.5 ms [unipolar]). Five of the 20 patients received a Kappa KDR701 generator, and atrial electrical properties were followed up over a 1-year period. There was no significant change in atrial pacing threshold or incidence of phrenic nerve stimulation over the 1-year follow-up. A standard single-pass VDD pacing lead system was capable of DDD pacing intraoperatively and during long-term follow-up. Despite higher than usual thresholds via the atrial dipole, pacemaker telemetry revealed < 10% use of atrial pacing dipole over a 12-month period, which would minimally deplete the pacemaker's battery. In addition, the telemetry confirmed appropriate sensing and pacing of the atrial dipole throughout the study period. At this time such systems can serve as back-up DDD pacing systems with further refinements required to optimize atrial thresholds in all patients.
引用
收藏
页码:323 / 329
页数:7
相关论文
共 7 条
  • [1] Encapsulation of the atrial dipole of chronically implanted VDD leads; a means of single-pass lead DDD pacing?
    Papouchado, M
    Barnes, E
    Crick, JCP
    EUROPACE '97 - THE OFFICIAL MEETING OF THE WORKING GROUPS ON CARDIAC PACING AND ARRHYTHMIAS OF THE EUROPEAN SOCIETY OF CARDIOLOGY, 1997, : 877 - 880
  • [2] Radiographic assessment of atrial dipole position in single pass lead VDD and DDD pacing
    Bongiorni, MG
    Moracchini, PV
    Nava, A
    Caprioli, V
    Gascón, D
    Morra, A
    Di Gregorio, F
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11): : 2240 - 2245
  • [3] Feasibility of Atrial Sensing Via a Free-Floating Single-Pass Defibrillation Lead for Dual-Chamber Defibrillators
    Andreas Schuchert
    Michael Niehaus
    Ludwig Binner
    Roberto Belke
    Thomas Meinertz
    Journal of Interventional Cardiac Electrophysiology, 2003, 8 : 209 - 214
  • [4] Feasibility of atrial sensing via a free-floating single-pass defibrillation lead for dual-chamber defibrillators
    Schuchert, A
    Niehaus, M
    Binner, L
    Belke, R
    Meinertz, T
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2003, 8 (03) : 209 - 214
  • [5] Radiographic assessment of atrial dipole position in single pass lead VDD and DDD pacing (vol 21, pg 2240, 1998)
    Bongiorni, MG
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (02): : VIII - VIII
  • [6] Single lead DDD pacing with atrial stimulation via floating ring electrode using overlapping biphasic impulse stimulation
    Tse, HF
    Lau, CP
    VI ASIAN-PACIFIC SYMPOSIUM ON CARDIAC PACING AND ELECTROPHYSIOLOGY, 1997, : 117 - 120
  • [7] Is single lead DDD pacing possible? Acute and long term results of atrial stimulation via floating ring electrode using overlapping biphasic stimulation
    HungFat, T
    ChuPak, L
    SimKeung, L
    Zoe, L
    CIRCULATION, 1996, 94 (08) : 797 - 797